Evaluation of a Novel Immunoassay for Lacosamide Therapeutic Drug Monitoring: Comparison With a Liquid Chromatography-Mass Spectrometry Assay

被引:4
作者
D'Urso, Annachiara [1 ]
Ricotta, Tiziano [2 ]
de Grazia, Ugo [1 ]
机构
[1] Fdn IRCCS Ist Neurol Carlo Besta, Milan, Italy
[2] Tema Ric Srl, Castenaso, Italy
关键词
lacosamide; therapeutic drug monitoring; liquid chromatography; mass spectrometry; immunoassay; PARTIAL-ONSET SEIZURES; ANTIEPILEPTIC DRUGS; EFFICACY; EPILEPSY; SAFETY;
D O I
10.1097/FTD.0000000000000450
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Monitoring serum levels of lacosamide, other than to establish individualized reference ranges may be helpful in several settings, including patients with liver and/or kidney failure or settings that may result in altered pharmacokinetic characteristics and to assess patients' compliance with therapy. In this study, the EurekaOne liquid chromatography-mass spectrometry (LC-MS/MS) method (in use method) and the ARK immunoassay method (new method) for lacosamide monitoring were compared. Methods: Lacosamide concentrations were determined in 39 patient samples using (1) antiepileptic drug LC-MS/MS kit by EurekaOne on a Thermo Fisher Scientific TSQuantum Access Max system and (2) the lacosamide immunoassay by ARK Diagnostic Inc. (research use only kit), on a Abbott Architect System. Results: Measured total imprecision of the new method is 6.29% at 6.59 mu mol/L, 8.82% at 30.20 mu mol/L, and 6.45% at 64.51 mu mol/L. Passing-Bablok regression analysis showed a nonsignificant intercept of -0.03015 [95% confidence interval (CI), -1.2243 to 0.8593] and a slope of 1.05 (95% CI, 0.9973-1.1166), showing that the method does not deviate from linearity and absence of proportional systematic error. Bland-Altman analysis showed a systematic bias of -3.296% (95% CI, -5.81 to -0.78) with 95% of the LC-MS/MS-ARK mean % of differences ranging from -18.5 to 11.9. Despite this bias, data of the combined imprecision of the 2 methods show that the new method is still acceptable within the maximum allowable error of 15%. Conclusions: The performance of the new ARK method on the Architect system is acceptable and may be used routinely to measure serum lacosamide concentration in the clinic although the nature of the bias has to be carefully addressed.
引用
收藏
页码:663 / 668
页数:6
相关论文
共 20 条
[11]   Adjunctive lacosamide for partial-onset seizures: Efficacy and safety results from a randomized controlled trial [J].
Halasz, Peter ;
Kalviainen, Reetta ;
Mazurkiewicz-Beldzinska, Maria ;
Rosenow, Felix ;
Doty, Pamela ;
Hebert, David ;
Sullivan, Timothy .
EPILEPSIA, 2009, 50 (03) :443-453
[12]   CLINICAL PERSPECTIVES ON LACOSAMIDE [J].
Halford, Jonathan J. ;
Lapointe, Marc .
EPILEPSY CURRENTS, 2009, 9 (01) :1-9
[13]  
Horstmann Rolf, 2002, Epilepsia, V43, P188
[14]  
Kestelyn C, 2011, BIOANALYSIS, V3, P2515, DOI [10.4155/BIO.11.261, 10.4155/bio.11.261]
[15]   Advances in anti-epileptic drug testing [J].
Krasowski, Matthew D. ;
McMillin, Gwendolyn A. .
CLINICA CHIMICA ACTA, 2014, 436 :224-236
[16]  
Landmark CJ, 2010, EXPERT REV NEUROTHER, V10, P119, DOI [10.1586/ern.09.136, 10.1586/ERN.09.136]
[17]  
Nickel B, 2008, J CLIN PHARMACOL, V48, P1129
[18]   Lacosamide neurotoxicity associated with concomitant use of sodium channel-blocking antiepileptic drugs: A pharmacodynamic interaction? [J].
Novy, Jan ;
Patsalos, Philip N. ;
Sander, Josemir W. ;
Sisodiya, Sanjay M. .
EPILEPSY & BEHAVIOR, 2011, 20 (01) :20-23
[19]   A sensitive and rapid method for quantification of lacosamide and desmethyl lacosamide by LC-MS/MS [J].
Payto, Drew ;
Foldvary-Schaefer, Nancy ;
So, Norman ;
Bruton, Monica ;
Wang, Sihe .
BIOANALYSIS, 2014, 6 (23) :3161-3168
[20]   Therapeutic Drug Monitoring of Lacosamide in Norway: Focus on Pharmacokinetic Variability, Efficacy and Tolerability [J].
Svendsen, Torleiv ;
Brodtkorb, Eylert ;
Baftiu, Arton ;
Burns, Margrete Larsen ;
Johannessen, Svein I. ;
Landmark, Cecilie Johannessen .
NEUROCHEMICAL RESEARCH, 2017, 42 (07) :2077-2083