Pitfalls and complications with locking plate for proximal humerus fracture

被引:198
作者
Clavert, Philippe [1 ]
Adam, Philippe [1 ]
Bevort, Adrien [1 ]
Bonnomet, Francois [1 ]
Kempf, Jean-Francois [1 ]
机构
[1] CHU Strasbourg, CCOM, Shoulder & Elbow Serv, F-67400 Illkirch Graffenstaden, France
关键词
Shoulder; proximal humerus fracture; complication; locking plate; OPEN REDUCTION; INTERNAL-FIXATION; LOCKED PLATE; TRAUMA CARE; HEMIARTHROPLASTY; MULTICENTER; SHOULDER;
D O I
10.1016/j.jse.2009.09.005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: The aim of this study was to identify specific complications of locking plate fixation of proximal humerus fractures. Patients and Methods: Seventy-threee adult patients with a displaced 3- (24%) or 4-part (76%) fracture of the proximal humerus were treated over a period of 2 years under the supervision of a trauma surgeon. Fourty-four patients came back for a clinical and radiographic examinations at least 18 months after the trauma; the others were evaluated at,6 weeks and 3 and 6 months. Results: Out of the 73 patients (64.4% females, mean age of 65), 11 patients needed a second surgery and 18 were lost for follow-up after 6 months. Mean final constant score was 62.3 points. The incidence of secondary displacement was 8.2%. Nonunion rate was 5.5%, affecting the constant score (P=.018). 16.4% of the patients developed a partial necrosis of the humeral head at the latest follow-up, which influenced on the constant score (P=.029). Quality of the reduction of the greater tuberosity influenced final results (P=.037). Screw cutout rate was 13.7%, with an influence to the constant score (P=.001). A too high plate positioning influenced the constant score (P=.002). Conclusion: Locked screw-plates provide more secure fixation of fractures, especially in weak bone. Complications rate remains high. Two complications are to be distinguished: 1) technical complications in plate positioning, length of the screws or secondary screw cutout strongly influence the final clinical result; and 2) specific complications related to this technology such as pseudarthrosis or plate fracture.
引用
收藏
页码:489 / 494
页数:6
相关论文
共 50 条
  • [41] Locking plate fixation for proximal humerus fractures-when do I use a fibular strut?
    Stone, Geoffrey P.
    Christmas, Kaitlyn N.
    Mighell, Mark A.
    ANNALS OF JOINT, 2020, 5
  • [42] Principles of Locking Plate Fixation of Proximal Humerus Fractures
    Omid, Reza
    Trasolini, Nicholas A.
    Stone, Michael A.
    Namdari, Surena
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2021, 29 (11) : E523 - E535
  • [43] Assessment of Functional Outcome and Postoperative Complications in Proximal Humerus Fracture Patients Managed With Proximal Humerus Internal Locking System (PHILOS) Plating
    Agrawal, Udit
    Vaibhav, B. K.
    Rao, Harsh Kirthi
    Praseeth, K. R.
    Narayandas, Durga K.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (06)
  • [44] Endosteal Strut Augment Reduces Complications Associated With Proximal Humeral Locking Plates
    Neviaser, Andrew S.
    Hettrich, Carolyn M.
    Beamer, Brandon S.
    Dines, Joshua S.
    Lorich, Dean G.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (12) : 3300 - 3306
  • [45] Die winkelstabile Plattenosteosynthese bei Frakturen am proximalen Humerus mit der LPHP (Locking Proximal Humerus Plate)Internal Fixation of Proximal Humerus Fractures Using the Locking Proximal Humerus Plate
    Michael Plecko
    Aurel Kraus
    Operative Orthopädie und Traumatologie, 2005, 17 (1) : 25 - 50
  • [46] Factors associated with failure of locking plate fixation in proximal humerus fractures
    Cosic, Filip
    Kirzner, Nathan
    Edwards, Elton
    Page, Richard
    Kimmel, Lara
    Gabbe, Belinda
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2025, 56 (02):
  • [47] Anterior fracture dislocation of the proximal humerus: Management and treatment results
    Schirren, Mirjam
    Siebenbuerger, Georg
    Fleischhacker, Evi
    Biermann, Niklas
    Helfen, Tobias
    Boecker, Wolfgang
    Ockert, Ben
    OBERE EXTREMITAET-SCHULTER-ELLENBOGEN-HAND-UPPER EXTREMITY-SHOULDER ELBOW HAND, 2019, 14 (02): : 103 - 109
  • [48] Dual plate for comminuted proximal humerus fractures
    Choi, Sungwook
    Seo, Kyu-Bum
    Kwon, Yong Suk
    Kang, Hyunseong
    Cho, Changhee
    Rho, Joseph Y.
    ACTA ORTHOPAEDICA BELGICA, 2019, 85 (04): : 429 - 436
  • [49] Complex fractures of the proximal humerus in the elderly-outcome and complications after locking plate fixation
    Schliemann B.
    Siemoneit J.
    Theisen Ch.
    Kösters C.
    Weimann A.
    Raschke M.J.
    MUSCULOSKELETAL SURGERY, 2012, 96 (Suppl 1) : S3 - S11
  • [50] Functional outcomes of locking plate fixation in complex proximal humerus fractures
    Shaikh, Saeed Ahmed
    Arain, Muhammad Arif
    Rehman, Muhammad Mansoor
    Ahmed, Nadeem
    Samejo, Muhammad Qasim Ali
    RAWAL MEDICAL JOURNAL, 2020, 45 (03): : 633 - 636