Convection-enhanced high-flux hemodialysis

被引:16
作者
Lee, Kyungsoo
Jeong, Jae Hoon
Mun, Cho Hae
Lee, Sa Ram
Yoo, Kyu Jae
Park, Yong Woo
Won, Yong Soon
Min, Byoung Goo
机构
[1] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Seoul 110744, South Korea
[2] Seoul Natl Univ, Interdisciplinary Program Biomed Engn, Seoul 110744, South Korea
[3] Seoul Natl Univ, Inst Med Engn, Med Res Ctr, Seoul 110744, South Korea
[4] Korea Artificial Organ Ctr, Seoul, South Korea
[5] Seoul Natl Univ Hosp, Inst Canc Res, Seoul 110744, South Korea
[6] Soonchunhyang Univ, Dept Thorac & Cardiovasc Surg, Puchon, South Korea
关键词
hemodialysis; hemodiafiltration; internal filtration; convection; backfiltration; hemodialyzer;
D O I
10.1111/j.1525-1594.2007.00436.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Internal filtration contributes to convective clearance in high-flux hemodialysis but its contribution is limited by low pressure gradients. Therefore, a modification using a conventional dialyzer was conceived to enhance internal filtration and backfiltration (BF) rates. The modified dialyzer includes two longitudinal independent regions for blood flow, which were created by redesigning dialyzer caps. Blood pressures remained higher than dialysate pressures in one region and lower in the other region, allowing continuous internal filtration and BF in these respective regions. Modified and conventional dialyzers were compared in terms of pressure gradients and solute clearances. Thus, our experiments involved two groups: the modified dialyzer group and the conventional dialyzer group. A renal failure model was established using a dog weighing 25-30 kg by renal artery and vein ligation. With the exception of the dialyzers, experimental conditions were identical in the two groups. The pressure gradients between blood and dialysate were much higher for the modified dialyzer than for the conventional dialyzer. No significant differences were observed with respect to small solute clearances between the two groups, but mid-range solute clearances were significantly higher in the modified group. More optimization is required before the devised unit can be used clinically. However, the devised unit offers a straightforward means of regulating internal filtration and BF rates.
引用
收藏
页码:653 / 658
页数:7
相关论文
共 17 条
  • [1] COLLINS GR, 2004, Patent No. 6719907
  • [2] Dellanna F, 1996, NEPHROL DIAL TRANSPL, V11, P83
  • [3] FALKVALL T, 1998, Patent No. 5730712
  • [4] Ghezzi P M, 1992, ASAIO J, V38, P61, DOI 10.1097/00002480-199201000-00015
  • [5] Effect of dialysis membranes and middle molecule removal on chronic hemodialysis patient survival
    Leypoldt, JK
    Cheung, AK
    Carroll, CE
    Stannard, DC
    Pereira, BJG
    Agodoa, LY
    Port, FK
    [J]. AMERICAN JOURNAL OF KIDNEY DISEASES, 1999, 33 (02) : 349 - 355
  • [6] Evaluation of hygiene and safety controls for on-line paired hemodiafiltration (PHF)
    Mandolfo, S
    Corsi, A
    Wratten, ML
    Sereni, L
    Imbasciati, E
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (02) : 160 - 165
  • [7] Push/pull hemodiafiltration: Technical aspects and clinical effectiveness
    Miwa, M
    Shinzato, T
    [J]. ARTIFICIAL ORGANS, 1999, 23 (12) : 1123 - 1126
  • [8] ULTRAFILTRATION AND BACKFILTRATION DURING HEMODIALYSIS
    OFSTHUN, NJ
    LEYPOLDT, JK
    [J]. ARTIFICIAL ORGANS, 1995, 19 (11) : 1143 - 1161
  • [9] Convective-control led double high flux hemodiafiltration: A novel blood purification modality
    Pisitkun, T
    Eiam-Ong, S
    Tiranathanagul, K
    Sakunsrijinda, C
    Manotham, K
    Hanvivatvong, O
    Suntaranuson, P
    Praditpornsilpa, K
    Chusil, S
    Tungsanga, K
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2004, 27 (03) : 195 - 204
  • [10] Enhancement of convective transport by internal filtration in a modified experimental hemodialyzer
    Ronco, C
    Orlandini, G
    Brendolan, A
    Lupi, A
    La Greca, G
    [J]. KIDNEY INTERNATIONAL, 1998, 54 (03) : 979 - 985