摘要
Objective: To evaluate the functional results of otosclerosis Surgery using diode laser. Study Design: Retrospective cohort analysis. Patients: One hundred seven patients operated on for otosclerosis with a diode laser (119 ears, all primary cases) and 141 patients operated on with a conventional technique (141 ears, all primary cases). Revision cases using the diode laser were also described. Methods: Preoperative tomographic computed scan findings and intraoperative observations were collected. Pure-tone and vocal audiometry was performed preoperatively and postoperatively (at 3 mo and 1 yr). Results: In the laser group, the air-bone gap was 29 +/- 0.8 dB (n = 112) preoperatively and 9 +/- 0.6 dB (n = 58) at 1 year. Air conduction was improved by 22 +/- 1.7 dB at 1 year (n = 58). In the conventional group, the air-bone gap was 32 +/- 0.9 dB (n = 127) preoperatively and 10 +/- 0.6 dB (n = 127) at 1 year. Air conduction was improved by 25 +/- 1.1 dB (n = 127) at 1 year. No difference of hearing gain was observed between the 2 groups at 1 year. A decreased rate of footplate fracture was observed with the diode laser (3.6%) compared with the conventional technique (21.3%). Conclusion: Diode laser is a reliable and safe device for otoselerosis surgery. The functional results were similar to those reported in other series.
引用
收藏
页码:441 / 446
页数:6
相关论文
共 24 条
  • [1] [Anonymous], 1995, Otolaryngol Head Neck Surg, V113, P186
  • [2] Clinical results after stapedotomy: A comparison between the erbium: Yttrium-aluminum-gamet laser and the conventional technique
    Arnoldner, Christoph
    Schwab, Burkard
    Lenarz, Thomas
    [J]. OTOLOGY & NEUROTOLOGY, 2006, 27 (04) : 458 - 465
  • [3] Stapedectomy training with the carbon dioxide laser
    Beatty, TW
    Haberkamp, TJ
    Khafagy, YW
    Bresemann, JA
    [J]. LARYNGOSCOPE, 1997, 107 (11) : 1441 - 1444
  • [4] BUCHMANN CA, COMPARISON ARGON CO2, V21, P227
  • [5] Hearing outcome following laser-assisted stapes surgery
    Garin, P
    Van Prooyen-Keyser, S
    Jamart, J
    [J]. JOURNAL OF OTOLARYNGOLOGY, 2002, 31 (01) : 31 - 34
  • [6] Häusler R, 1999, ACTA OTO-LARYNGOL, V119, P207
  • [7] CO2 laser stapedotomy with the "one-shot" technique -: clinical results
    Jovanovic, S
    Schönfeld, U
    Scherer, H
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2004, 131 (05) : 750 - 757
  • [8] Retrospective case study of carbon dioxide laser, stapedotomy with lens-based and mirror-based micromanipulators
    Lescanne, E
    Moriniere, S
    Gohler, C
    Manceau, A
    Beutter, P
    Robier, A
    [J]. JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2003, 117 (04) : 256 - 260
  • [9] Lesinski S G, 1989, Laryngoscope, V99, P1
  • [10] Lippert BM, 2001, LASER SURG MED, V28, P11, DOI 10.1002/1096-9101(2001)28:1<11::AID-LSM1010>3.3.CO