Numerical Evaluation of Image Homogeneity, Signal-to-Noise Ratio, and Specific Absorption Rate for Human Brain Imaging at 1.5, 3, 7, 10.5, and 14T in an 8-Channel Transmit/Receive Array

被引:49
作者
Cao, Zhipeng [1 ]
Park, Joshua [2 ]
Cho, Zang-Hee [2 ]
Collins, Christopher M. [3 ]
机构
[1] Penn State Univ, Dept Bioengn, Hershey, PA USA
[2] Gachon Univ Med & Sci, Neurosci Res Inst, Inchon, South Korea
[3] NYU, Dept Radiol, Coll Med, New York, NY 10016 USA
基金
美国国家卫生研究院;
关键词
high-field MRI; FDTD; Bloch simulation; signal-to-noise ratio; RF heating; SAR; STATIC MAGNETIC-FIELD; MRI; TISSUE; PULSE; POWER; HEAD; COIL; RECONSTRUCTION; OPTIMIZATION; SENSITIVITY;
D O I
10.1002/jmri.24689
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo predict signal-to-noise ratio (SNR) trends and absorbed energy in magnetic resonance imaging (MRI) of the brain up to 14T. Materials and MethodsA human head in an eight-channel transmit/receive coil was simulated with Maxwell and Bloch equations to determine excitation homogeneity with radiofrequency (RF) shimming, image homogeneity, SNR, and absorbed energy in MRI from 1.5 to 14T considering realistic field distributions and relaxation properties. ResultsRF shimming alone achieved a standard deviation in excitation flip angle less than 10 degrees in mid-brain up to 14T, but produced a small region with low excitation on a lower slice. Current reconstruction methods may produce shading artifacts at 14T. SNR increases with a greater-than-linear rate for gradient recalled echo (GRE) sequences having short (2 msec) echo time (TE) and long relaxation time (TR) (approximate to 2.3-fold increase from 7T to 14T), but a less-than-linear rate if TE is 10 msec (approximate to 1.6-fold increase from 7T to 14T). Depending on the sequence, SNR per square root of imaging time may produce a less-than-linear increase with B-0. Whole-head absorbed energy shows a less-than-quadratic increase with B-0 (1.7-fold increase from 7T to 14T). ConclusionNumerical simulations indicate that with proper preparation and precautions, imaging of the human brain at up to 14T could be performed safely, with advantages in SNR. J. Magn. Reson. Imaging 2015;41:1432-1439. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1432 / 1439
页数:8
相关论文
共 36 条
[1]  
[Anonymous], P 10 ANN M ISMRM HON
[2]   Bloch-Based MRI System Simulator Considering Realistic Electromagnetic Fields for Calculation of Signal, Noise, and Specific Absorption Rate [J].
Cao, Zhipeng ;
Oh, Sukhoon ;
Sica, Christopher T. ;
McGarrity, John M. ;
Horan, Timothy ;
Luo, Wei ;
Collins, Christopher M. .
MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (01) :237-247
[3]   RADIOFREQUENCY FIELD PROPAGATION IN CONDUCTIVE NMR SAMPLES [J].
CARLSON, JW .
JOURNAL OF MAGNETIC RESONANCE, 1988, 78 (03) :563-573
[4]   An Approach to Rapid Calculation of Temperature Change in Tissue Using Spatial Filters to Approximate Effects of Thermal Conduction [J].
Carluccio, Giuseppe ;
Erricolo, Danilo ;
Oh, Sukhoon ;
Collins, Christopher M. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (06) :1735-1741
[5]  
Cho Z-H, 2009, S EXTR HIGH FIELD MR
[6]   kT-points: Short three-dimensional tailored RF pulses for flip-angle homogenization over an extended volume [J].
Cloos, M. A. ;
Boulant, N. ;
Luong, M. ;
Ferrand, G. ;
Giacomini, E. ;
Le Bihan, D. ;
Amadon, A. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (01) :72-80
[7]   Array-optimized composite pulse for excellent whole-brain homogeneity in high-field MRI [J].
Collins, Christopher M. ;
Wang, Zhangwei ;
Mao, Weihua ;
Fang, Jieming ;
Liu, Wanzhan ;
Smith, Michael B. .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (03) :470-474
[8]   Calculation of Radiofrequency Electromagnetic Fields and Their Effects in MRI of Human Subjects [J].
Collins, Christopher M. ;
Wang, Zhangwei .
MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (05) :1470-1482
[9]   Numerical calculations of the static magnetic field in three-dimensional multi-tissue models of the human head [J].
Collins, CM ;
Yang, B ;
Yang, QX ;
Smith, MB .
MAGNETIC RESONANCE IMAGING, 2002, 20 (05) :413-424
[10]   Signal-to-noise ratio and absorbed power as functions of main magnetic field strength, and definition of "90°" RF pulse for the head in the birdcage coil [J].
Collins, CM ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (04) :684-691