Validation of non-invasive arterial pressure monitoring during carotid endarterectomy

被引:32
|
作者
Heusdens, J. F. [1 ]
Lof, S. [2 ]
Pennekamp, C. W. A. [2 ]
Specken-Welleweerd, J. C. [2 ]
de Borst, G. J. [2 ]
van Klei, W. A. [1 ]
van Wolfswinkel, L. [1 ]
Immink, R. V. [1 ,3 ]
机构
[1] Univ Med Ctr Utrecht, Dept Anaesthesiol, Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Vasc Surg, Utrecht, Netherlands
[3] Acad Med Ctr Amsterdam, Dept Anaesthesiol, H 1-158,POB 22660, NL-1100 DD Amsterdam, Netherlands
关键词
carotid endarterectomy; measurement techniques; arterial pressure; Nexfin; non-invasive continuous arterial pressure measurement; RANDOMIZED CONTROLLED-TRIALS; BLOOD-PRESSURE; CARDIAC-SURGERY; RADIAL ARTERY; STENOSIS; NEXFIN; PHENYLEPHRINE; METAANALYSIS; ANESTHESIA; MANAGEMENT;
D O I
10.1093/bja/aew268
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Patients undergoing carotid endarterectomy require strict arterial blood pressure (BP) control to maintain adequate cerebral perfusion. In this study we tested whether non-invasive beat-to-beat Nexfin finger BP (BPfin) can replace invasive beat-to-beat radial artery BP (BPrad) in this setting. Methods: In 25 consecutive patients (median age 71 yr) scheduled for carotid endarterectomy and receiving general anaesthesia, BPfin and BPrad were monitored simultaneously and ipsilaterally during the 30-min period surrounding carotid artery cross-clamping. Validation was guided by the standard set by the Association for the Advancement of Medical Instrumentation (AAMI), which considers a BP monitor adequate when bias (precision) is <5 (8) mm Hg, respectively. Results: BPfin, us BPrad bias (precision) was -3.3 (10.8), 6.1 (5.7) and 3.5 (5.2) mm Hg for systolic, diastolic, and mean BP, respectively. One subject was excluded due to a poor quality BP curve. In another subject, mean BPfin overestimated mean BPrad by 13.5 mm Hg. Conclusion: Mean BPfin could be considered as an alternative for mean BPrad during a carotid endarterectomy, based on the AAMI criteria. In 23 of 24 patients, the use of mean BPfin would not lead to decisions to adjust mean BPrad values outside the predefined BP threshold.
引用
收藏
页码:316 / 323
页数:8
相关论文
共 50 条
  • [1] NON-INVASIVE PERIOPERATIVE MONITORING OF CAROTID ENDARTERECTOMY
    GARRETT, WV
    SLAYMAKER, EE
    BARNES, RW
    JOURNAL OF SURGICAL RESEARCH, 1979, 26 (03) : 255 - 261
  • [2] Non-invasive arterial pressure monitoring revisited
    Frederic Michard
    Daniel I. Sessler
    Bernd Saugel
    Intensive Care Medicine, 2018, 44 : 2213 - 2215
  • [3] Non-invasive arterial pressure monitoring revisited
    Michard, Frederic
    Sessler, Daniel I.
    Saugel, Bernd
    INTENSIVE CARE MEDICINE, 2018, 44 (12) : 2213 - 2215
  • [4] Non-invasive arterial pressure monitoring during liver transplantation.
    Findlay, JY
    Gali, B
    Keegan, MT
    Jankowski, CJ
    Burke, CM
    Plevak, DJ
    LIVER TRANSPLANTATION, 2004, 10 (06) : C2 - C2
  • [5] NON-INVASIVE EVALUATION AFTER CAROTID ENDARTERECTOMY
    DOORLY, TPG
    SHANIK, DG
    EUROPEAN SURGICAL RESEARCH, 1981, 13 (01) : 48 - 49
  • [6] Techniques for Non-Invasive Monitoring of Arterial Blood Pressure
    Meidert, Agnes S.
    Saugel, Bernd
    FRONTIERS IN MEDICINE, 2018, 4
  • [7] NON-INVASIVE DETECTION OF CAROTID STENOSIS FOLLOWING ENDARTERECTOMY
    CANTELMO, NL
    CUTLER, BS
    WHEELER, HB
    HERRMANN, JB
    CARDULLO, PA
    ARCHIVES OF SURGERY, 1981, 116 (08) : 1005 - 1008
  • [8] Postoperative monitoring of non-invasive absolute cerebral oxygen saturation after carotid endarterectomy
    Beran, M.
    De Deyne, C.
    Lauwers, G.
    Jans, F.
    Heylen, R.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2010, 27 (01) : 63 - 63
  • [9] Non-Invasive Cerebral Autoregulation Monitoring During Awake Carotid Endarterectomy Identifies Clinically Significant Brain Ischaemia
    Zipfel, Julian
    Bantle, Sebastian J.
    Magunia, Harry
    Schlensak, Christian
    Neunhoeffer, Felix
    Schuhmann, Martin U.
    Lescan, Mario
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2020, 60 (05) : 647 - 654
  • [10] The ClearSight System for Postoperative Arterial Blood Pressure Monitoring After Carotid Endarterectomy: A Validation Study
    Fassaert, Leonie M. M.
    Plate, Joost D. J.
    Westerink, Jan
    Immink, Rogier, V
    de Borst, Gert J.