Do small-aperture presbyopic corrections influence the visual field?

被引:18
作者
Atchison, David A. [1 ]
Blazaki, Stella [2 ]
Suheimat, Marwan [1 ]
Plainis, Sotiris [2 ,3 ]
Charman, W. Neil [3 ]
机构
[1] Queensland Univ Technol, Sch Optometry & Vis Sci, Inst Hlth & Biomed Innovat, Brisbane, Qld 4001, Australia
[2] Univ Crete, Inst Vis & Opt, Iraklion, Crete, Greece
[3] Univ Manchester, Fac Life Sci, Manchester, Lancs, England
关键词
depth-of-focus; KAMRA inlay; presbyopia; small-aperture optics; visual field; CORNEAL INLAY; INTEROCULAR DIFFERENCES; INTRACORNEAL INLAY; RETINAL IMAGES; HUMAN-EYE; FOCUS;
D O I
10.1111/opo.12265
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To explore the effect of small-aperture optics, designed to aid presbyopes by increasing ocular depth-of-focus, on measurements of the visual field. Methods: Simple theoretical and ray-tracing models were used to predict the impact of different designs of small-aperture contact lenses or corneal inlays on the proportion of light passing through natural pupils of various diameters as a function of the direction in the visual field. The left eyes of five healthy volunteers were tested using three afocal, hand-painted opaque soft contact lenses (www.davidthomas.com). Two were opaque over a 10 mm diameter but had central clear circular apertures of 1.5 and 3.0 mm in diameter. The third had an annular opaque zone with inner and outer diameters of 1.5 and 4.0 mm, approximately simulating the geometry of the KAMRA inlay (www.acufocus.com). A fourth, clear lens was used for comparison purposes. Visual fields along the horizontal meridian were evaluated up to 50 degrees eccentricity with static automated perimetry (Medmont M700, stimulus Goldmann-size III; www.medmont.com). Results: According to ray-tracing, the two lenses with the circular apertures were expected to reduce the relative transmittance of the pupil to zero at specific field angles (around 60 degrees for the conditions of the experimental measurements). In contrast, the annular stop had no effect on the absolute field but relative transmittance was reduced over the central area of the field, the exact effects depending upon the natural pupil diameter. Experimental results broadly agreed with these theoretical expectations. With the 1.5 and 3.0 mm pupils, only minor losses in sensitivity (around 2 dB) in comparison with the clear-lens case occurred across the central 10 degrees radius of field. Beyond this angle, sensitivity losses increased, to reach about 7 dB at the edge of the measured field (50 degrees). The field results with the annular stop showed at most only a slight loss in sensitivity (<= 3 dB) across the measured field. Conclusion: The present theoretical and experimental results support earlier clinical findings that KAMRA-type annular stops, unlike circular artificial pupils, have only minor effects on measurements of the visual field.
引用
收藏
页码:51 / 59
页数:9
相关论文
共 31 条
[11]  
FREEMAN E, 1952, Am J Optom Arch Am Acad Optom, V29, P347
[12]  
Gabriel P, 1988, CLIN EXP OPTOM, V71, P184
[13]   Retinal images viewed through a small aperture corneal inlay [J].
Inoue, Makoto ;
Bissen-Miyajima, Hiroko ;
Arai, Hiroyuki ;
Hirakata, Akito .
ACTA OPHTHALMOLOGICA, 2014, 92 (02) :E168-E169
[14]   VISUAL-FIELD CONSTRICTION CAUSED BY COLORED CONTACT-LENSES [J].
INSLER, MS ;
HENDRICKS, C ;
GEORGE, DM .
ARCHIVES OF OPHTHALMOLOGY, 1988, 106 (12) :1680-1682
[15]  
JOSEPHSON JE, 1987, AM J OPTOM PHYS OPT, V64, P38
[16]   SMALL-DIAMETER CORNEAL INLAY IN PRESBYOPIC OR PSEUDOPHAKIC PATIENTS [J].
KEATES, RH ;
MARTINES, E ;
TENNEN, DG ;
REICH, C .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1995, 21 (05) :519-521
[17]   Vignetting and Field of View with the KAMRA Corneal Inlay [J].
Langenbucher, Achim ;
Goebels, Susanne ;
Szentmary, Nora ;
Seitz, Berthold ;
Eppig, Timo .
BIOMED RESEARCH INTERNATIONAL, 2013, 2013
[18]   Corneal inlays for presbyopia correction [J].
Lindstrom, Richard L. ;
MacRae, Scott M. ;
Pepose, Jay S. ;
Hoopes, Phillip C., Sr. .
CURRENT OPINION IN OPHTHALMOLOGY, 2013, 24 (04) :281-287
[19]   Pupil shape as viewed along the horizontal visual field [J].
Mathur, Ankit ;
Gehrmann, Julia ;
Atchison, David A. .
JOURNAL OF VISION, 2013, 13 (06)
[20]  