Evaluation of direct reading photoionization detector performance under various operational parameters

被引:4
作者
Khoshakhlagh, Amir Hossein [1 ,2 ]
Golbabaei, Farideh [2 ]
Beygzadeh, Mojtaba [3 ]
Carrasco-Marin, Francisco [4 ]
Shahtaheri, Seyed Jamaleddin [2 ,5 ]
机构
[1] Kashan Univ Med Sci, Fac Hlth, Dept Occupat Hlth Engn, Kashan, Iran
[2] Univ Tehran Med Sci, Fac Publ Hlth, Dept Occupat Hlth Engn, Tehran, Iran
[3] Mat & Energy Res Ctr, Dept Energy, Tehran, Iran
[4] Univ Granada, Fac Sci, Carbon Mat Res Grp, Avda Fuente Nueva S-N, Granada, Spain
[5] Univ Tehran Med Sci, Inst Environm Res, Tehran, Iran
来源
ENVIRONMENTAL HEALTH ENGINEERING AND MANAGEMENT JOURNAL | 2021年 / 8卷 / 02期
基金
美国国家科学基金会;
关键词
Toluene; Temperatures; Humidity; Adsorption; Chromatography; Gas regression analysis; Research design; GAS-CHROMATOGRAPH; EXPOSURE; INDOOR; AIR; ADSORPTION; BTEX;
D O I
10.34172/EHEM.2021.15
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Background: A hand-held portable direct-reading monitor, including photoionization detector (PID) is renowned for its good sensitivity, considerable dynamic range, and nondestructive vapor detection ability in comparison to the tardy response of the PID in gas chromatography (GC), which its application has been restricted. In this study, the performance of a PID system (MultiRAE Lite) was evaluated as a replacement of GC in the measurement of toluene in a dynamic adsorption system. Methods: The test was done at different relative humidity levels (30%, 50%, and 80%), temperatures (21, 30, 40 degrees C), and toluene concentrations (20, 100, 200, and 400 ppm). Results: The PID achieved 48% of all measurements meeting the comparison criterion. The results showed that the performance of the PID could be altered by the variables. The best performance of the PID was at temperature of 21 degrees C, the relative humidity of 50%, and concentration of 200 ppm with the percentage of readings achieving the criterion of comparison to 58%, 54%, and 52%, respectively. The averages of the PID readings (mean +/- SD at 200 ppm = 207.9 +/- 8.7) were higher than the reference method measurements averages (mean +/- SD at 200 ppm = 203.5 1 +/- 5.8). The regression analysis of the toluene results from the PID and the reference method results indicated that the measurements were significantly correlated (r(2) = 0.93). Conclusion: According to the results, the device response is linear. Therefore, the findings are acceptable in adsorption studies. In this way, the measurement of the sample concentration should be performed using the same instrument before and after the reactor in order to calculate the adsorption efficiency.
引用
收藏
页码:123 / 128
页数:6
相关论文
共 17 条
[1]  
Azari MR, 2012, INT J OCCUP ENV MED, V3, P39
[2]   A comparative study on metal organic frameworks for indoor environment application: Adsorption evaluation [J].
Bahri, Mitra ;
Haghighat, Fariborz ;
Kazemian, Hossein ;
Rohani, Sohrab .
CHEMICAL ENGINEERING JOURNAL, 2017, 313 :711-723
[3]   AN EVALUATION OF THE RESPONSE OF SOME PORTABLE, DIRECT-READING 10.2 EV AND 11.8 EV PHOTOIONIZATION DETECTORS, AND A FLAME IONIZATION GAS-CHROMATOGRAPH FOR ORGANIC VAPORS IN HIGH HUMIDITY ATMOSPHERES [J].
BARSKY, JB ;
HEE, SSQ ;
CLARK, CS .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1985, 46 (01) :9-14
[4]   Measurement Capability of Field Portable Organic Vapor Monitoring Instruments Under Different Experimental Conditions [J].
Coffey, Christopher C. ;
Pearce, Terri A. ;
Lawrence, Robert B. ;
Hudnall, Judith B. ;
Slaven, James E. ;
Martin, Stephen B., Jr. .
JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2009, 6 (01) :1-8
[5]  
Coy JD, 2000, AM IND HYG ASSOC J, V61, P268, DOI 10.1080/15298660008984536
[6]  
Drummond I, 1997, AM IND HYG ASSOC J, V58, P820, DOI 10.1080/15428119791012333
[7]   Phytoremediation of BTEX from indoor air by Hyrcanian plants [J].
Fooladi, Mahta ;
Moogouei, Roxana ;
Jozi, Seyed Ali ;
Golbabaei, Farideh ;
Tajadod, Golnaz .
ENVIRONMENTAL HEALTH ENGINEERING AND MANAGEMENT JOURNAL, 2019, 6 (04) :233-240
[8]  
Jahangiri M, 2011, FRESEN ENVIRON BULL, V20, P1036
[9]   Effect of interferents on the performance of direct-reading organic vapor monitors [J].
LeBouf, Ryan F. ;
Coffey, Christopher C. .
JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 2015, 65 (03) :261-269
[10]  
McNair H.M., 2019, Basic Gas Chromatography