APPLICATIONS OF CFD TECHNIQUE IN THE DESIGN AND FLOW ANALYSIS OF IMPLANTABLE AXIAL FLOW BLOOD PUMP

被引:16
作者
Fan Hui-min [1 ]
Hong Fang-wen [2 ]
Zhang Guo-ping [2 ]
Ye Liang [1 ]
Liu Zhong-min [1 ]
机构
[1] Tongji Univ, Shanghai E Hosp, Shanghai 200120, Peoples R China
[2] China Ship Sci Res Ctr, Wuxi 214082, Peoples R China
基金
国家高技术研究发展计划(863计划);
关键词
axial flow blood pump; CFD; PIV; PERFORMANCE;
D O I
10.1016/S1001-6058(09)60084-6
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
Based on established numerical methods and hydrodynamic performance testing facilities, CFD technique are applied to improve the design of the implantable axial flow blood pump and the flow analysis. (1) Applying brushless machine magnet steel, reducing its thickness while increasing the length, the flow channel's cross-section is increased, with no space connection between the large and the small rotators, and with a cone transition segment from the bearing to the principal axis, the flow is made smoother. The rotating speed is lowered by 1000 rpm - 1200 rpm under the same flow rate and pressure head, and thus the hemolysis can be avoided. (2) Different outlet stator guiding vanes are selected for the same blood pump for analyses of hydrodynamic performances and flow fields. An excellent design not only can regulate the rotating flow field into an axial one, reduce the circulation loss and improve the pump efficiency, but also can avoid backflow, vortex and secondary flow at the pump outlet, and thus the thrombus can be prevented. (3) The calf live tests show that some residual clots exist at the inner wall of the outlet connection bends, which are analyzed and explained by CFD techniques and the corresponding improvements are proposed. All results are verified by hydrodynamic performance tests and PIV flow field tests, and consistent conclusions are obtained.
引用
收藏
页码:518 / 525
页数:8
相关论文
共 14 条
[1]   DESIGN OF IMPLANTABLE AXIAL-FLOW BLOOD PUMP AND NUMERICAL STUDIES ON ITS PERFORMANCE [J].
Fan Hui-min ;
Hong Fang-wen ;
Zhou Lian-di ;
Chen Yin-sheng ;
Ye Liang ;
Liu Zhong-min .
JOURNAL OF HYDRODYNAMICS, 2009, 21 (04) :445-452
[2]   SAR and temperature distributions in canonical head models exposed to near- and far-field electromagnetic radiation at different frequencies [J].
Lin, JC ;
Wang, ZW .
ELECTROMAGNETIC BIOLOGY AND MEDICINE, 2005, 24 (03) :405-421
[3]  
LIU ZM, 2008, Patent No. 20082200548022
[4]  
LIU ZM, 2007, Patent No. 200720077869
[5]  
LIU ZM, 2007, Patent No. 2007200778704
[6]   Computational fluid dynamics analysis of an intra-cardiac axial flow pump [J].
Mitoh, A ;
Yano, T ;
Sekine, K ;
Mitamura, Y ;
Okamoto, E ;
Kim, DW ;
Yozu, R ;
Kawada, S .
ARTIFICIAL ORGANS, 2003, 27 (01) :34-40
[7]  
Potapov Evgenij V, 2003, J Extra Corpor Technol, V35, P274
[8]   New technique for implantation of the inflow canula of Berlin Heart INCOR system [J].
Tschirkov, Alexander ;
Nikolov, Dimitar ;
Papantchev, Vassil .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2006, 30 (04) :678-679
[9]   Numerical and experimental analysis of an axial flow left ventricular assist device: The influence of the diffuser on overall pump performance [J].
Untaroiu, A ;
Throckmorton, AL ;
Patel, SM ;
Wood, HG ;
Allaire, PE ;
Olsen, DB .
ARTIFICIAL ORGANS, 2005, 29 (07) :581-591
[10]  
WU GH, 2004, BEIJING BIOMEDICAL E, V23, P215