The Biodurability of Covering Materials for Metallic Stents in a Bile Flow Phantom

被引:26
作者
Bang, Byoung Wook [1 ]
Jeong, Seok [1 ]
Lee, Don Haeng [1 ,2 ,3 ,4 ]
Lee, Jung Il [1 ]
Lee, Se Chul [5 ,6 ]
Kang, Sung-Gwon [5 ,6 ]
机构
[1] Inha Univ, Div Gastroenterol, Dept Internal Med, Sch Med, Sinheung Dong 400711, Incheon, South Korea
[2] Inha Univ, Ctr Adv Med Educ, Brain Korea Project 21, Sch Med, Sinheung Dong 400711, Incheon, South Korea
[3] Utah Inha DDS, Inchon, South Korea
[4] Adv Therapeut Researh Ctr, Inchon, South Korea
[5] Seoul Natl Univ, S&G Biotech Inc, Bundang Hosp, Songnam, South Korea
[6] Seoul Natl Univ, Dept Radiol, Bundang Hosp, Songnam, South Korea
关键词
Stents; Silicones; Polytetrafluoroethylene; Polyurethanes; Bile duct neoplasms; Common bile duct; MALIGNANT BILIARY OBSTRUCTION; RANDOMIZED-TRIAL; PALLIATIVE TREATMENT; FOLLOW-UP; STRICTURES; WALLSTENTS; JAUNDICE; BYPASS;
D O I
10.1007/s10620-011-1958-6
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Covered biliary metal stents have been introduced for the purpose of overcoming tumor ingrowth and treatment of benign biliary stricture. The aim of this study was to evaluate the biodurability of three commercially available biliary metal stent covering materials [e-PTFE (expanded polytetrafluoroethylene), silicone, and polyurethane] in a bile flow phantom. By operation of a peristaltic pump, human bile was circulated continuously in an experimental perfusion system containing covered metal stents. Each stent was removed, respectively, 1, 2, 4, and 6 months after bile exposure. We performed a gross inspection of the covered stents. The covering membrane was detached from the stent and observed by scanning electron microscopy (SEM). Finally, we measured tensile and tear strength of the membranes. Bile-staining of the membrane showed gradual progression after bile exposure; however, progress was the fastest in e-PTFE. SEM examination showed that the polyurethane surface was smooth, and the silicone surface was relatively smooth. However, e-PTFE had a rough and uneven surface. After bile exposure, there were no significant changes in polyurethane and silicone; however, biofilms and microcracks were observed in e-PTFE. In contrast to a gradual decrease of tensile/tear strength of polyurethane and silicone, those of e-PTFE showed a rapid reduction despite of the strongest baseline tensile and tear strength. e-PTFE tended to form biofilms more frequently than polyurethane and silicone during bile exposure. e-PTFE seemed to be less durable than silicone and polyurethane, however, as clinically applicable material because of strong absolute tensile/tear strengths.
引用
收藏
页码:1056 / 1063
页数:8
相关论文
共 24 条
[1]   RANDOMIZED TRIAL OF ENDOSCOPIC ENDOPROSTHESIS VERSUS OPERATIVE BYPASS IN MALIGNANT OBSTRUCTIVE-JAUNDICE [J].
ANDERSEN, JR ;
SORENSEN, SM ;
KRUSE, A ;
ROKKJAER, M ;
MATZEN, P .
GUT, 1989, 30 (08) :1132-1135
[2]   Palliation of malignant obstructive jaundice [J].
Baron, TH .
GASTROENTEROLOGY CLINICS OF NORTH AMERICA, 2006, 35 (01) :101-112
[3]   Partially covered self-expandable metallic stents for benign biliary strictures due to chronic pancreatitis [J].
Behm, B. ;
Brock, A. ;
Clarke, B. W. ;
Ellen, K. ;
Northup, P. G. ;
Dumonceau, J. M. ;
Kahaleh, M. .
ENDOSCOPY, 2009, 41 (06) :547-551
[4]   A survival analysis of patients with malignant biliary strictures treated by percutaneous metallic stenting [J].
Brountzos, Elias N. ;
Ptochis, Nikolaos ;
Panagiotou, Irene ;
Malagari, Katerina ;
Tzavara, Chara ;
Kelekis, Dimitrios .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 30 (01) :66-73
[5]   Management of biliary neoplastic obstruction with two different metallic stents implanted in one session [J].
Gandini, R ;
Fabiano, S ;
Pipitone, V ;
Spinelli, A ;
Reale, CA ;
Colangelo, V ;
Pampana, E ;
Romagnoli, A ;
Simonetti, G .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 28 (01) :48-52
[6]   Microbial biofilms associated with biliary stent clogging [J].
Guaglianone, Emilio ;
Cardines, Rita ;
Vuotto, Claudia ;
Di Rosa, Roberta ;
Babini, Valentina ;
Mastrantonio, Paola ;
Donelli, Gianfranco .
FEMS IMMUNOLOGY AND MEDICAL MICROBIOLOGY, 2010, 59 (03) :410-420
[7]  
Gwon DI, 2010, AM J ROENTGENOL, V195, P463
[8]   ePTFE/FEP-covered metallic stents for palliation of malignant biliary disease: Can tumor ingrowth be prevented? [J].
Hatzidakis, Adam ;
Krokidis, Miltiadis ;
Kalbakis, Kostantinos ;
Romanos, Jiannis ;
Petrakis, Ioannis ;
Gourtsoyiannis, Nicholas .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 30 (05) :950-958
[9]   Treatment of malignant biliary obstruction with polyurethane-covered wallstents [J].
Hausegger, KA ;
Thurnher, S ;
Bodendörfer, G ;
Zollikofer, CL ;
Uggowitzer, M ;
Kugler, C ;
Lammer, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (02) :403-408
[10]   INFLUENCE OF FIBRIL LENGTH UPON EPTFE GRAFT HEALING AND HOST MODIFICATION OF THE IMPLANT [J].
HIRABAYASHI, K ;
SAITOH, E ;
IJIMA, H ;
TAKENAWA, T ;
KODAMA, M ;
HORI, M .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (11) :1433-1447