Effect of dental implant cross-sectional design on cortical bone structure using finite element analysis

被引:15
作者
Abu-Hammad, Osama
Khraisat, Ameen
Dar-Odeh, Najla
El-Maaytah, Mohammed
机构
[1] Univ Jordan, Fac Dent, Dept Conservat Dent & Prosthodont, Amman, Jordan
[2] Univ Jordan, Fac Dent, Dept Oral Med Oral Surg & Periodont, Amman, Jordan
[3] UCL Hosp, Oral & Maxillofacial Surg Head & Neck Unit, London, England
[4] Eastman Dent Inst, Div Maxillofacial Diagnost Med & Surg Sci, London, England
关键词
implant design; implant thread; bone stress; strain; implant diameter; finite element analysis;
D O I
10.1111/j.1708-8208.2007.00048.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This finite element analysis investigation evaluated the effect of different implant cross-sectional designs on bone stress levels under different loading patterns. Materials and Methods: Finite element analysis program was used to construct four different three-dimensional models describing 4 x 10-mm implants in blocks of cortical and trabecular bone. A 5-mm-long abutment was modeled above each implant. The implant in model 1 was unthreaded, while in model 2 the implant was circularly threaded. The third implant in model 3 had the cross-sectional shape as a 16-sided star-shaped design. The implant in model 4 was constructed unthreaded, with a diameter of 4.5mm. Vertical and horizontal loads of 100 N each were applied on the top middle node of each implant assembly. All nodes at the bottom surface of the bone models were restrained. Results: By comparing models 1, 2, and 3, the lowest bone stress values under vertical and horizontal forces were observed around the unthreaded implant in model 1 (8.92 and 94.52MPa, respectively). The highest stress value under vertical loading was shown around the threaded implant in model 2 (10.07MPa), whereas the highest stress value under horizontal loading was observed around the star-shaped implant in model 3 (108.40MPa). Model 4, with a wider unthreaded design, had stress values under vertical and horizontal loading of 7.32 and 71.35MPa, respectively. Conclusions: It was concluded that the unthreaded implant design produced the least bone stress. An increase in implant diameter could produce marked reduction in stress value in the bone around the neck of the implant.
引用
收藏
页码:217 / 221
页数:5
相关论文
共 22 条
[1]  
Abu-Hammad OA, 2000, INT J ORAL MAX IMPL, V15, P559
[2]  
ABUHAMMAD O, UNPUB CLIN IMPLANT D
[3]  
Carlsson L, 1988, Int J Oral Maxillofac Implants, V3, P21
[4]   Application of finite element analysis in implant dentistry: A review of the literature [J].
Geng, JP ;
Tan, KBC ;
Liu, GR .
JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (06) :585-598
[5]  
Gentile MA, 2005, INT J ORAL MAX IMPL, V20, P930
[6]   Influence of titanium surface charge on fibroblast adhesion [J].
Hamdan, M ;
Blanco, L ;
Khraisat, A ;
Tresguerres, IF .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2006, 8 (01) :32-38
[7]   Stress and inflammation as a detrimental combination for peri-implant bone loss [J].
Heckmann, S. M. ;
Linke, J. J. ;
Graef, F. ;
Foitzik, Ch. ;
Wichmann, M. G. ;
Weber, H. -P. .
JOURNAL OF DENTAL RESEARCH, 2006, 85 (08) :711-716
[8]  
Khraisat Ameen, 2004, Clin Implant Dent Relat Res, V6, P222, DOI 10.1111/j.1708-8208.2004.tb00038.x
[9]  
Lai H, 1998, Chin J Dent Res, V1, P7
[10]   Biomechanical effect of abutment on stability of orthodontic mini-implant [J].
Motoyoshi, M ;
Yano, S ;
Tsuruoka, T ;
Shimizu, N .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) :480-485