Distal Embolization of Hydrophilic-Coating Material From Coronary Guidewires After Percutaneous Coronary Interventions

被引:49
作者
Grundeken, Maik J. [1 ]
Li, Xiaofei [2 ]
Kurpershoek, C. Eline [2 ]
Kramer, Miranda C. [1 ]
Vink, Aryan [3 ]
Piek, Jan J. [1 ]
Tijssen, Jan G. P. [1 ]
Koch, Karel T. [1 ]
Wykrzykowska, Joanna J. [1 ]
de Winter, Robbert J. [1 ]
van der Wal, Allard C. [2 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Heartctr, NL-1105 AZ Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Pathol, NL-1105 AZ Amsterdam, Netherlands
[3] Univ Med Ctr Utrecht, Dept Pathol, Utrecht, Netherlands
关键词
catheterization; coronary disease; embolism; myocardium; stents; thrombosis; CARDIAC-CATHETERIZATION; POLYMER EMBOLI; ARTERIES;
D O I
10.1161/CIRCINTERVENTIONS.114.001816
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Coronary guidewires are indispensable during percutaneous coronary interventions. Nowadays, most guidewires have hydrophilic coatings to improve their trackability, allowing easy lesion passage and facilitating balloon and stent positioning. Recent reports, however, have raised concerns about detachment and subsequent embolization of these hydrophilic coatings. Methods and Results-We have retrospectively reviewed the histological samples of the myocardium, obtained during autopsies in the period 2009 to 2013, from all patients who had a history of percutaneous coronary interventions (n=40). Foreign material was observed in the distal myocardium in 4 patients (10%). Furthermore, we have reviewed 205 thrombus specimens which were obtained during thrombus aspiration in the setting of primary percutaneous coronary interventions in the period 2005 to 2009. In 45% of the cases, foreign material was observed within the thrombus. Finally, we have examined the histopathologic appearance of hydrophilic-guidewire coating material ex vivo by embedding the coating in placenta specimen and cut and stain it in exactly the same manner as the myocardium and thrombus specimen. The histopathologic appearance of the hydrophilic coating ex vivo was identical to the foreign material found in vivo. Conclusions-Distal embolization of hydrophilic-coating material was observed in 10% of the patients who had a history of percutaneous coronary interventions. Hydrophilic-coating material was found in 45% of coronary thrombus specimen obtained during thrombus aspiration. These findings suggest that detachment and distal embolization of hydrophilic-coating material from coronary guidewires occur more often than the sparse literature on this topic suggests.
引用
收藏
页数:7
相关论文
共 14 条
[1]   Embolization of Hydrophilic Catheter Coating to the Lungs Report of a Case Mimicking Granulomatous Vasculitis [J].
Allan, Robert W. ;
Alnuaimat, Hassan ;
Edwards, William D. ;
Tazelaar, Henry D. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2009, 132 (05) :794-797
[2]  
Barnwell SL, 1997, AM J NEURORADIOL, V18, P1886
[3]   Autopsy demonstration of intramyocardial polymer gel emboli associated with a giant-cell reaction following cardiac catheterization: a case report [J].
El-Najjar, Vicky ;
Robinson, Morton .
CARDIOVASCULAR PATHOLOGY, 2012, 21 (01) :59-61
[4]   Complications of endovascular polymers associated with vascular introducer sheaths and metallic coils in 3 patients, with literature review [J].
Fealey, Michael E. ;
Edwards, William D. ;
Giannini, Caterina ;
Piepgras, David G. ;
Cloft, Harry ;
Rihal, Charanjit S. .
AMERICAN JOURNAL OF SURGICAL PATHOLOGY, 2008, 32 (09) :1310-1316
[5]   TRANS-LUMINAL DILATATION OF CORONARY-ARTERY STENOSIS [J].
GRUNTZIG, A .
LANCET, 1978, 1 (8058) :263-263
[6]   Presence of Older Thrombus Is an Independent Predictor of Long-Term Mortality in Patients With ST-Elevation Myocardial Infarction Treated With Thrombus Aspiration During Primary Percutaneous Coronary Intervention [J].
Kramer, Miranda C. A. ;
van der Wal, Allard C. ;
Koch, Karel T. ;
Ploegmakers, Johanna P. H. M. ;
van der Schaaf, Rene J. ;
Henriques, Jose P. S. ;
Baan, Jan, Jr. ;
Rittersma, Saskia Z. H. ;
Vis, Marije M. ;
Piek, Jan J. ;
Tijssen, Jan G. P. ;
de Winter, Robbert J. .
CIRCULATION, 2008, 118 (18) :1810-1816
[7]   Hydrophilic polymer emboli: an under-recognized iatrogenic cause of ischemia and infarct [J].
Mehta, Rupal I. ;
Mehta, Rashi I. ;
Solis, Orestes E. ;
Jahan, Reza ;
Salamon, Noriko ;
Tobis, Jonathan M. ;
Yong, William H. ;
Vinters, Harry V. ;
Fishbein, Michael C. .
MODERN PATHOLOGY, 2010, 23 (07) :921-930
[8]   Intravascular polymer material after coil embolization of a giant cerebral aneurysm [J].
Mehta, Rupal I. ;
Mehta, Rashi I. ;
Fishbein, Michael C. ;
Solis, Orestes E. ;
Jahan, Reza ;
Salamon, Noriko ;
Vinters, Harry V. ;
Yong, William H. .
HUMAN PATHOLOGY, 2009, 40 (12) :1803-1807
[9]  
Muenst S, 2012, PATHOLOGE, V33, P345, DOI 10.1007/s00292-012-1588-6
[10]   Plaque instability frequently occurs days or weeks before occlusive coronary thrombosis - A pathological thrombectomy study in primary percutaneous coronary intervention [J].
Rittersma, SZH ;
van der Wal, AC ;
Koch, KT ;
Piek, JJ ;
Henriques, JPS ;
Mulder, KJ ;
Ploegmakers, JPHM ;
Meesterman, M ;
de Winter, RJ .
CIRCULATION, 2005, 111 (09) :1160-1165