Calibration of the TonoLab Tonometer in Mice with Spontaneous or Experimental Glaucoma

被引:48
作者
Pease, Mary E. [1 ]
Cone, Frances E. [1 ]
Gelman, Scott [1 ]
Son, Janice L. [1 ]
Quigley, Harry A. [1 ]
机构
[1] Johns Hopkins Univ, Wilmer Ophthalmol Inst, Glaucoma Serv, Sch Med, Baltimore, MD 21218 USA
关键词
MOUSE INTRAOCULAR-PRESSURE; NONINVASIVE MEASUREMENT; OCULAR HYPERTENSION; REBOUND TONOMETRY; TONOPEN; MODEL; RATS; GOLDMANN;
D O I
10.1167/iovs.10-5556
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To measure the accuracy of TonoLab (TioLat, Helsinki, Finland) tonometry in mice with spontaneous or induced experimental glaucoma. METHODS. Chronic intraocular pressure (IOP) elevation was induced in one eye of 32 mice by injection of polystyrene beads and viscoelastic material. Three to 6 weeks later, the eyes were cannulated and manometrically set to 10, 20, 30, 40, or 50 mm Hg. The mice were 8-week and 8-month-old C57BL/6, 8-week-old DBA/2J, and 8-week-old CD1. The TonoLab calibration was also tested on five aged DBA/2J mice with spontaneous glaucoma. The relation of the TonoLab reading to manometric IOP was evaluated in multivariate linear regression models with axial length, IOP history, and mouse strain as independent variables. RESULTS. The slope of the relationship between TonoLab and manometric IOP in all the mice was 0.998, with an intercept of 2.3 mm Hg (adjusted R in univariate regression = 0.86). Neither the mice with bead-induced glaucoma nor those with spontaneous glaucoma (older DBA/2J mice) differed significantly from the control animals in having an excellent correlation between TonoLab and manometer IOP. Longer and wider mouse eyes had slightly higher tonometrically measured IOP, whether glaucomatous or control (multivariate regression, adjusted R-2 = 0.90, P < 0.0001). There was no difference in tonometric accuracy among the three mouse strains: CD1, C57BL/6, and DBA/2J, nor between 8-week and 8-month-old C57BL/6 mice (multivariate regression, P = 0.32). CONCLUSIONS. The TonoLab accurately reflects IOP in both normal mice and in eyes of mice with experimental or spontaneous glaucoma, with no detectable effect of age. (Invest Ophthalmol Vis Sci. 2011;52:858-864) DOI:10.1167/iovs.10-5556
引用
收藏
页码:858 / 864
页数:7
相关论文
共 26 条
[1]   Experimental mouse ocular hypertension: Establishment of the model [J].
Aihara, M ;
Lindsey, JD ;
Weinreb, RN .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (10) :4314-4320
[2]   Noninvasive Intraocular pressure measurements in mice by pneumotonometry [J].
Avila, MY ;
Múnera, A ;
Guzmán, A ;
Do, CW ;
Wang, Z ;
Stone, RA ;
Civan, MM .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (09) :3274-3280
[3]  
Avila MY, 2001, INVEST OPHTH VIS SCI, V42, P1841
[4]  
Cohan BE, 2001, INVEST OPHTH VIS SCI, V42, P2560
[5]   Differential susceptibility to experimental glaucoma among 3 mouse strains using bead and viscoelastic injection [J].
Cone, Frances E. ;
Gelman, Scott E. ;
Son, Janice L. ;
Pease, Mary E. ;
Quigley, Harry A. .
EXPERIMENTAL EYE RESEARCH, 2010, 91 (03) :415-424
[6]   Method for the noninvasive measurement of intraocular pressure in mice [J].
Danias, J ;
Kontiola, AI ;
Filippopoulos, T ;
Mittag, T .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (03) :1138-1141
[7]   Predictability and limitations of non-invasive murine tonometry: Comparison of two devices [J].
Filippopoulos, Theodoros ;
Matsubara, Akihisa ;
Danias, John ;
Huang, Wei ;
Dobberfuhl, Adam ;
Ren, Lizhen ;
Mittag, Thomas ;
Miller, Joan W. ;
Grosskreutz, Cynthia L. .
EXPERIMENTAL EYE RESEARCH, 2006, 83 (01) :194-201
[8]   Laser-induced mouse model of chronic ocular hypertension [J].
Grozdanic, SD ;
Betts, DM ;
Sakaguchi, DS ;
Allbaugh, RA ;
Kwon, YH ;
Kardon, RH .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (10) :4337-4346
[9]   Mouse genetic models: an ideal system for understanding glaucomatous neurodegeneration and neuroprotection [J].
Howell, Gareth R. ;
Libby, Richard T. ;
John, Simon W. M. .
GLAUCOMA: AN OPEN WINDOW TO NEURODEGENERATION AND NEUROPROTECTION, 2008, 173 :303-321
[10]   Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness [J].
Iliev, M. E. ;
Goldblum, D. ;
Katsoulis, K. ;
Amstutz, C. ;
Frueh, B. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2006, 90 (07) :833-835