Automated Tube Voltage Selection in Thoracoabdominal Computed Tomography at High Pitch Using a Third-Generation Dual-Source Scanner Image Quality and Radiation Dose Performance

被引:46
|
作者
Lurz, Markus [1 ]
Lell, Michael M. [1 ,2 ]
Wuest, Wolfgang [1 ]
Eller, Achim [1 ]
Scharf, Michael [1 ]
Uder, Michael [1 ,2 ]
May, Matthias Stefan [1 ,2 ]
机构
[1] Univ Erlangen Nurnberg, Dept Radiol, D-91054 Erlangen, Germany
[2] Imaging Sci Inst, Erlangen, Germany
关键词
tube voltage; dual-source CT; high pitch; chest; abdomen; image quality; CORONARY CT ANGIOGRAPHY; CURRENT MODULATION; ITERATIVE RECONSTRUCTIONS; KILOVOLT SELECTION; ABDOMINAL CT; REDUCTION; CHEST; STANDARD; PHANTOM; EXPOSURE;
D O I
10.1097/RLI.0000000000000133
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The objective of this study was to evaluate the radiation dose and image quality performance of thoracoabdominal examinations with an automated tube voltage selection (tube voltage adaptation), tube current modulation, and high pitch using a third-generation dual-source computed tomography (CT) compared intraindividually with 120-kV examinations with tube current modulation with special attention on clinically relevant lesions in the liver, the lungs, and extrahepatic soft tissues. Materials and Methods: This study was approved by the institutional review board. Computed tomography of the body was performed using a third-generation dual-source systemin 95 patients (mean body mass index, 25 kg/m(2); range, 18-35 kg/m(2)). For 49 of these patients, all calculated tube settings and resulting dose values were recorded for each of the 12 gradual contrast weightings of the tube voltage adaptation algorithm. Spiral CTwas performed for all patients with an intermediate weighting (grade 7) in a portal venous phase at 120 reference kV, 180 reference mAs, and pitch of 1.55. Objective image quality was assessed on the basis of contrast-to-noise ratio. Subjective image quality was assessed on the basis of clarity and sharpness of anatomical and pathological structures aswell as interfering beam hardening and spiral and motion artifacts (heart, lungs, diaphragm). Previous examinations on a 64-slice scanner served as reference. Results: All examinationswere rated good or excellent for clinical diagnosis. Automated tube voltage selection resulted in significantly lower effective radiation dose (9.5 mSv) compared with the reference (12.0 mSv; P < 0.01). Contrast-to-noise ratio and image quality of soft tissue lesions were significantly increased (P < 0.01). Motion artifacts were significantly reduced (P < 0.01). Conclusions: Automated tube voltage adaptation combined with high-pitch protocols allows for a substantial radiation dose reduction while substantially increasing the image quality, even at large-volume exposure.
引用
收藏
页码:352 / 360
页数:9
相关论文
共 50 条
  • [41] Performance of Iterative Reconstruction and Automated Tube Voltage Selection on the Image Quality and Radiation Dose in Abdominal CT Scans
    Desai, Gaurav S.
    Orrego, Jorge M. Fuentes
    Kambadakone, Avinash R.
    Sahani, Dushyant V.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (06) : 897 - 903
  • [42] Impact of heart rate on coronary computed tomographic angiography interpretability with a third-generation dual-source scanner
    Miller, Robert J. H.
    Eisenberg, Evann
    Friedman, John
    Cheng, Victor
    Hayes, Sean
    Tamarappoo, Balaji
    Thomson, Louise
    Berman, Daniel S.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2019, 295 : 42 - 47
  • [43] Full Field Image Reconstruction Is Suitable for High-Pitch Dual-Source Computed Tomography
    Mahnken, Andreas H.
    Allmendinger, Thomas
    Sedlmair, Martin
    Tamm, Miriam
    Reinartz, Sebastian D.
    Flohr, Thomas
    INVESTIGATIVE RADIOLOGY, 2012, 47 (11) : 642 - 648
  • [44] Image quality and radiation dose of ECG-triggered High-Pitch Dual-Source cardiac computed tomography angiography in children for the evaluation of central vascular stents
    Christian A. Barrera
    Hansel J. Otero
    Ammie M. White
    David Saul
    David M. Biko
    The International Journal of Cardiovascular Imaging, 2019, 35 : 367 - 374
  • [45] Image quality and radiation dose of ECG-triggered High-Pitch Dual-Source cardiac computed tomography angiography in children for the evaluation of central vascular stents
    Barrera, Christian A.
    Otero, Hansel J.
    White, Ammie M.
    Saul, David
    Biko, David M.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2019, 35 (02) : 367 - 374
  • [46] Strengths and limitations of coronary angiography with turbo high-pitch third-generation dual-source CT
    Ochs, M. M.
    Andre, F.
    Korosoglou, G.
    Fritz, T.
    Seitz, S.
    Bogomazov, Y.
    Schlett, C. L.
    Sokiranski, R.
    Sommer, A.
    Gueckel, F.
    Brado, M.
    Kauczor, H. -U.
    Goerich, J.
    Buss, S. J.
    CLINICAL RADIOLOGY, 2017, 72 (09) : 739 - 744
  • [47] Diagnostic Performance of Different Simulated Low-Dose Levels in Patients with Suspected Cervical Abscess Using a Third-Generation Dual-Source CT Scanner
    Winkelmann, Moritz T.
    Afat, Saif
    Walter, Sven S.
    Stock, Eva
    Schwarze, Vincent
    Brendlin, Andreas
    Kolb, Manuel
    Artzner, Christoph P.
    Othman, Ahmed E.
    DIAGNOSTICS, 2020, 10 (12)
  • [48] First Clinical Evaluation of High-Pitch Dual-Source Computed Tomographic Angiography Comparing Automated Tube Potential Selection With Automated Tube Current Modulation
    Beeres, Martin
    Williams, Kimberly
    Bauer, Ralf W.
    Scholtz, Jan
    Kaup, Moritz
    Gruber-Rouh, Tatjana
    Lee, Clara
    Wichmann, Julian L.
    Frellesen, Claudia
    Nour-Eldin, Nour-Eldin A.
    Vogl, Thomas J.
    Kerl, Josef Matthias
    Bodelle, Boris
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (04) : 624 - 628
  • [49] Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
    Linsen, Philip V. M.
    Coenen, Adriaan
    Lubbers, Marisa M.
    Dijkshoorn, Marcel L.
    Ouhlous, Mohamed
    Nieman, Koen
    JOURNAL OF CLINICAL IMAGING SCIENCE, 2016, 6
  • [50] Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study
    Felix G. Meinel
    Christian Canstein
    U. Joseph Schoepf
    Martin Sedlmaier
    Bernhard Schmidt
    Brett S. Harris
    Thomas G. Flohr
    Carlo N. De Cecco
    European Radiology, 2014, 24 : 1643 - 1650