Automated Tube Voltage Selection in Thoracoabdominal Computed Tomography at High Pitch Using a Third-Generation Dual-Source Scanner Image Quality and Radiation Dose Performance

被引:46
|
作者
Lurz, Markus [1 ]
Lell, Michael M. [1 ,2 ]
Wuest, Wolfgang [1 ]
Eller, Achim [1 ]
Scharf, Michael [1 ]
Uder, Michael [1 ,2 ]
May, Matthias Stefan [1 ,2 ]
机构
[1] Univ Erlangen Nurnberg, Dept Radiol, D-91054 Erlangen, Germany
[2] Imaging Sci Inst, Erlangen, Germany
关键词
tube voltage; dual-source CT; high pitch; chest; abdomen; image quality; CORONARY CT ANGIOGRAPHY; CURRENT MODULATION; ITERATIVE RECONSTRUCTIONS; KILOVOLT SELECTION; ABDOMINAL CT; REDUCTION; CHEST; STANDARD; PHANTOM; EXPOSURE;
D O I
10.1097/RLI.0000000000000133
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The objective of this study was to evaluate the radiation dose and image quality performance of thoracoabdominal examinations with an automated tube voltage selection (tube voltage adaptation), tube current modulation, and high pitch using a third-generation dual-source computed tomography (CT) compared intraindividually with 120-kV examinations with tube current modulation with special attention on clinically relevant lesions in the liver, the lungs, and extrahepatic soft tissues. Materials and Methods: This study was approved by the institutional review board. Computed tomography of the body was performed using a third-generation dual-source systemin 95 patients (mean body mass index, 25 kg/m(2); range, 18-35 kg/m(2)). For 49 of these patients, all calculated tube settings and resulting dose values were recorded for each of the 12 gradual contrast weightings of the tube voltage adaptation algorithm. Spiral CTwas performed for all patients with an intermediate weighting (grade 7) in a portal venous phase at 120 reference kV, 180 reference mAs, and pitch of 1.55. Objective image quality was assessed on the basis of contrast-to-noise ratio. Subjective image quality was assessed on the basis of clarity and sharpness of anatomical and pathological structures aswell as interfering beam hardening and spiral and motion artifacts (heart, lungs, diaphragm). Previous examinations on a 64-slice scanner served as reference. Results: All examinationswere rated good or excellent for clinical diagnosis. Automated tube voltage selection resulted in significantly lower effective radiation dose (9.5 mSv) compared with the reference (12.0 mSv; P < 0.01). Contrast-to-noise ratio and image quality of soft tissue lesions were significantly increased (P < 0.01). Motion artifacts were significantly reduced (P < 0.01). Conclusions: Automated tube voltage adaptation combined with high-pitch protocols allows for a substantial radiation dose reduction while substantially increasing the image quality, even at large-volume exposure.
引用
收藏
页码:352 / 360
页数:9
相关论文
共 50 条
  • [31] Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose
    May, Matthias S.
    Kramer, Manuel R.
    Eller, Achim
    Wuest, Wolfgang
    Scharf, Michael
    Brand, Michael
    Saake, Marc
    Schmidt, Bernhard
    Uder, Michael
    Lell, Michael M.
    NEURORADIOLOGY, 2014, 56 (09) : 797 - 803
  • [32] Cardiac Computed Tomography Angiography With Automatic Tube Potential Selection Effects on Radiation Dose and Image Quality
    Ghoshhajra, Brian B.
    Engel, Leif-Christopher
    Karolyi, Mihaly
    Sidhu, Manavjot Singh
    Wai, Bryan
    Barreto, Mitya
    Shanmugam, Uthamalingam
    Hoffmann, Udo
    Brady, Thomas J.
    Kalra, Manudeep
    Abbara, Suhny
    JOURNAL OF THORACIC IMAGING, 2013, 28 (01) : 40 - 48
  • [33] Estimated radiation dose and image quality comparison of the scan protocols in dual-source computed tomography coronary angiography
    Karaarslan, Ercan
    Ulus, Sila
    Bavbek, Cengiz
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2013, 57 (04) : 407 - 414
  • [34] Limits of the possible: diagnostic image quality in coronary angiography with third-generation dual-source CT
    Marco M. Ochs
    Fabian aus dem Siepen
    Thomas Fritz
    Florian Andre
    Gitsios Gitsioudis
    Grigorios Korosoglou
    Sebastian Seitz
    Yuriy Bogomazov
    Christopher L. Schlett
    Roman Sokiranski
    Andre Sommer
    Friedemann Gückel
    Matthias Brado
    Hans-Ulrich Kauczor
    Johannes Görich
    Matthias G. W. Friedrich
    Hugo A. Katus
    Sebastian J. Buss
    Clinical Research in Cardiology, 2017, 106 : 485 - 492
  • [35] Image quality and required radiation dose for coronary computed tomography angiography using an automatic tube potential selection technique
    Wang, Ying
    Wang, Xiaohua
    Zhang, Yan
    Lu, Yuliu
    Wu, Runze
    Yuan, Huishu
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2014, 30 : 89 - 94
  • [36] Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3rd generation dual-source CT
    Mangold, Stefanie
    Wichmann, Julian L.
    Schoepf, U. Joseph
    Poole, Zachary B.
    Canstein, Christian
    Varga-Szemes, Akos
    Caruso, Damiano
    Bamberg, Fabian
    Nikolaou, Konstantin
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2016, 26 (10) : 3608 - 3616
  • [37] Third-generation dual-source 70-kVp chest CT angiography with advanced iterative reconstruction in young children: image quality and radiation dose reduction
    Oliver Rompel
    Martin Glöckler
    Rolf Janka
    Sven Dittrich
    Robert Cesnjevar
    Michael M. Lell
    Michael Uder
    Matthias Hammon
    Pediatric Radiology, 2016, 46 : 462 - 472
  • [38] The ultrafast, high-pitch turbo FLASH mode of third-generation dual-source CT: Effect of different pitch and corresponding SFOV on image quality in a phantom study
    Zhou, Yang
    Hu, Lei
    Du, Silin
    Jin, Rui
    Li, Wangjia
    Lv, Fajin
    Zhang, Zhiwei
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (12): : 158 - 167
  • [39] Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3rd generation dual-source CT
    Stefanie Mangold
    Julian L. Wichmann
    U. Joseph Schoepf
    Zachary B. Poole
    Christian Canstein
    Akos Varga-Szemes
    Damiano Caruso
    Fabian Bamberg
    Konstantin Nikolaou
    Carlo N. De Cecco
    European Radiology, 2016, 26 : 3608 - 3616
  • [40] Tube Current Modulation Between Single- and Dual-Energy CT With a Second-Generation Dual-Source Scanner: Radiation Dose and Image Quality
    Matsubara, Kosuke
    Takata, Tadanori
    Kobayashi, Masanao
    Kobayashi, Satoshi
    Koshida, Kichiro
    Gabata, Toshifumi
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (02) : 354 - 361