Automated Tube Voltage Selection in Thoracoabdominal Computed Tomography at High Pitch Using a Third-Generation Dual-Source Scanner Image Quality and Radiation Dose Performance

被引:46
|
作者
Lurz, Markus [1 ]
Lell, Michael M. [1 ,2 ]
Wuest, Wolfgang [1 ]
Eller, Achim [1 ]
Scharf, Michael [1 ]
Uder, Michael [1 ,2 ]
May, Matthias Stefan [1 ,2 ]
机构
[1] Univ Erlangen Nurnberg, Dept Radiol, D-91054 Erlangen, Germany
[2] Imaging Sci Inst, Erlangen, Germany
关键词
tube voltage; dual-source CT; high pitch; chest; abdomen; image quality; CORONARY CT ANGIOGRAPHY; CURRENT MODULATION; ITERATIVE RECONSTRUCTIONS; KILOVOLT SELECTION; ABDOMINAL CT; REDUCTION; CHEST; STANDARD; PHANTOM; EXPOSURE;
D O I
10.1097/RLI.0000000000000133
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The objective of this study was to evaluate the radiation dose and image quality performance of thoracoabdominal examinations with an automated tube voltage selection (tube voltage adaptation), tube current modulation, and high pitch using a third-generation dual-source computed tomography (CT) compared intraindividually with 120-kV examinations with tube current modulation with special attention on clinically relevant lesions in the liver, the lungs, and extrahepatic soft tissues. Materials and Methods: This study was approved by the institutional review board. Computed tomography of the body was performed using a third-generation dual-source systemin 95 patients (mean body mass index, 25 kg/m(2); range, 18-35 kg/m(2)). For 49 of these patients, all calculated tube settings and resulting dose values were recorded for each of the 12 gradual contrast weightings of the tube voltage adaptation algorithm. Spiral CTwas performed for all patients with an intermediate weighting (grade 7) in a portal venous phase at 120 reference kV, 180 reference mAs, and pitch of 1.55. Objective image quality was assessed on the basis of contrast-to-noise ratio. Subjective image quality was assessed on the basis of clarity and sharpness of anatomical and pathological structures aswell as interfering beam hardening and spiral and motion artifacts (heart, lungs, diaphragm). Previous examinations on a 64-slice scanner served as reference. Results: All examinationswere rated good or excellent for clinical diagnosis. Automated tube voltage selection resulted in significantly lower effective radiation dose (9.5 mSv) compared with the reference (12.0 mSv; P < 0.01). Contrast-to-noise ratio and image quality of soft tissue lesions were significantly increased (P < 0.01). Motion artifacts were significantly reduced (P < 0.01). Conclusions: Automated tube voltage adaptation combined with high-pitch protocols allows for a substantial radiation dose reduction while substantially increasing the image quality, even at large-volume exposure.
引用
收藏
页码:352 / 360
页数:9
相关论文
共 50 条
  • [1] Automated Tube Voltage Adaptation in Combination with Advanced Modeled Iterative Reconstruction in Thoracoabdominal Third-Generation 192-Slice Dual-Source Computed Tomography: Effects on Image Quality and Radiation Dose
    Scholtz, Jan-Erik
    Wichmann, Julian L.
    Huesers, Kristina
    Beeres, Martin
    Nour-Eldin, Nour-Eldin Abdelrehim
    Frellesen, Claudia
    Vogl, Thomas J.
    Lehnert, Thomas
    ACADEMIC RADIOLOGY, 2015, 22 (09) : 1081 - 1087
  • [2] Impact on Image Quality and Radiation Dose of Third-Generation Dual-Source Computed Tomography of the Coronary Arteries
    Apfaltrer, Georg
    Szolar, Dieter H.
    Wurzinger, Eric
    Takx, Richard A. P.
    Nance, John W.
    Dutschke, Anja
    Tschauner, Sebastian
    Loewe, Christian
    Ringl, Helmut
    Sorantin, Erich
    Apfaltrer, Paul
    AMERICAN JOURNAL OF CARDIOLOGY, 2017, 119 (08) : 1156 - 1161
  • [3] Radiation Dose and Image Quality in Pediatric Cardiac Computed Tomography: A Comparison Between Sequential and Third-Generation Dual-Source High-Pitch Spiral Techniques
    Koh, Huiliang
    Ong, Ching Ching
    Choo, Yun Song
    Liang, Chong Ri
    Tan, Grace Hui Zhen
    Lim, Terence Chee Wen
    Quek, Swee Chye
    Sriram, Shankar
    San Teo, Lynette Li
    PEDIATRIC CARDIOLOGY, 2016, 37 (08) : 1397 - 1403
  • [4] Radiation Dose and Image Quality in Pediatric Cardiac Computed Tomography: A Comparison Between Sequential and Third-Generation Dual-Source High-Pitch Spiral Techniques
    Huiliang Koh
    Ching Ching Ong
    Yun Song Choo
    Chong Ri Liang
    Grace Hui Zhen Tan
    Terence Chee Wen Lim
    Swee Chye Quek
    Shankar Sriram
    Lynette Li San Teo
    Pediatric Cardiology, 2016, 37 : 1397 - 1403
  • [5] Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose
    Scholtz, Jan-Erik
    Wichmann, Julian L.
    Huesers, Kristina
    Albrecht, Moritz H.
    Beeres, Martin
    Bauer, Ralf W.
    Vogl, Thomas J.
    Bodelle, Boris
    EUROPEAN RADIOLOGY, 2016, 26 (08) : 2623 - 2631
  • [6] Radiation dose and image quality with abdominal computed tomography with automated dose-optimized tube voltage selection
    Hu, Lirong
    Wang, Yujun
    Hou, Hongtao
    Wei, Fuquan
    Yang, Guangzhao
    Chen, Yougeng
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2014, 42 (04) : 1011 - 1017
  • [7] Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT
    Agostini, Andrea
    Mari, Alberto
    Lanza, Cecilia
    Schicchi, Nicolo
    Borgheresi, Alessandra
    Maggi, Stefania
    Giovagnoni, Andrea
    RADIOLOGIA MEDICA, 2019, 124 (08): : 745 - 752
  • [8] Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: Feasibility, image quality, radiation dose, and effect of iterative reconstruction
    Hell, Michaela M.
    Bittner, Daniel
    Schuhbaeck, Annika
    Muschiol, Gerd
    Brand, Michael
    Lell, Michael
    Uder, Michael
    Achenbach, Stephan
    Marwan, Mohamed
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2014, 8 (06) : 418 - 425
  • [9] Radiation dose and image quality of high-pitch emergency abdominal CT in obese patients using third-generation dual-source CT (DSCT)
    Forbrig, Robert
    Ingrisch, Michael
    Stahl, Robert
    Winter, Katharina Stella
    Reiser, Maximilian
    Trumm, Christoph G.
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [10] Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography A Phantom Study on Radiation Dose and Image Quality
    Papadalds, Antonios E.
    Damilakis, John
    INVESTIGATIVE RADIOLOGY, 2019, 54 (05) : 265 - 272