Cardiopulmonary Resuscitation Interruptions With Use of a Load-Distributing Band Device During Emergency Department Cardiac Arrest

被引:43
作者
Ong, Marcus Eng Hock [1 ]
Annathurai, Annitha [1 ]
Shahidah, Ahmad [1 ]
Leong, Benjamin Sieu-Hon [3 ]
Ong, Victor Yeok Kein [3 ]
Tiah, Ling [4 ]
Ang, Shiang Hu [4 ]
Yong, Kok Leong [1 ]
Sultana, Papia [2 ]
机构
[1] Singapore Gen Hosp, Dept Emergency Med, Singapore 169608, Singapore
[2] Singapore Gen Hosp, Dept Clin Res, Singapore 169608, Singapore
[3] Natl Univ Singapore Hosp, Dept Emergency Med, Singapore 119074, Singapore
[4] Changi Gen Hosp, Singapore, Singapore
关键词
CHEST COMPRESSION DEVICE; AMERICAN-HEART-ASSOCIATION; AUTOMATIC MECHANICAL COMPRESSION; INTERNATIONAL LIAISON COMMITTEE; HEALTH-CARE PROFESSIONALS; STROKE-FOUNDATION; TASK-FORCE; RECOMMENDED GUIDELINES; UTSTEIN TEMPLATES; OUTCOME REPORTS;
D O I
10.1016/j.annemergmed.2010.01.004
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objective: Our primary aim is to measure no-flow time and no-flow ratio before and after an emergency department (ED) switched from manual to a load-distributing band mechanical cardiopulmonary resuscitation (CPR) device. Methods: This was a phased, before-after cohort evaluation at an urban tertiary hospital ED. We collected continuous video and chest compression data with the Physiocontrol CodeStat Suite 7.0 for resuscitations during the period just before and after adoption of load-distributing band CPR. All out-of-hospital, nontraumatic cardiac arrest, adult patients were eligible. From February 2007 to July 2008, there were 26 manual and 41 load-distributing band cases. Results: Patients in both phases were comparable in terms of demographics, medical history, witnessed arrest, arrest location, bystander CPR rates, out-of-hospital defibrillation, initial rhythm, and ED defibrillation. The median no-flow time, defined as the sum of all pauses between compressions longer than 1.5 seconds, during the first 5 minutes of resuscitation, was manual CPR 85 seconds (interquartile range [IQR] 45 to 112 seconds) versus load-distributing band 104 seconds (IQR 69 to 151 seconds). The mean no-flow ratio, defined as no-flow time divided by segment length, was manual 0.28 versus load-distributing band 0.40 (difference=-0.12; 95% confidence interval -0.22 to -0.02). However, from 5 to 10 minutes into the resuscitation, median no-flow time was manual 85 seconds (IQR 59 to 151 seconds) versus load-distributing band 52 seconds (IQR 34 to 82 seconds) and mean no-flow ratio manual 0.34 versus load-distributing band 0.21 (difference=0.13; 95% confidence interval 0.02 to 0.24). The average time to apply load-distributing band CPR during this period was 152 seconds. Conclusion: Application of a load-distributing band in the ED is associated with a higher no-flow ratio than manual CPR in the first 5 minutes of resuscitation. We suggest that attention to team training, rapid application of the device to minimize interruption, and feedback from defibrillator and video recordings may be useful to improve resuscitation team performance. [Ann Emerg Med. 2010;56:233-241.]
引用
收藏
页码:233 / 241
页数:9
相关论文
共 35 条
[1]   Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest [J].
Abella, BS ;
Alvarado, JP ;
Myklebust, H ;
Edelson, DP ;
Barry, A ;
O'Hearn, N ;
Vanden Hoek, TL ;
Becker, LB .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 293 (03) :305-310
[2]   Chest compression rates during cardiopulmonary resuscitation are suboptimal - A prospective study during in-hospital cardiac arrest [J].
Abella, BS ;
Sandbo, N ;
Vassilatos, P ;
Alvarado, JP ;
O'Hearn, N ;
Wigder, HN ;
Hoffman, P ;
Tynus, K ;
Vanden Hoek, TL ;
Becker, LB .
CIRCULATION, 2005, 111 (04) :428-434
[4]   OUTCOME OF CPR IN A LARGE METROPOLITAN-AREA - WHERE ARE THE SURVIVORS [J].
BECKER, LB ;
OSTRANDER, MP ;
BARRETT, J ;
KONDOS, GT .
ANNALS OF EMERGENCY MEDICINE, 1991, 20 (04) :355-361
[5]   INCIDENCE OF CARDIAC-ARREST - A NEGLECTED FACTOR IN EVALUATING SURVIVAL RATES [J].
BECKER, LB ;
SMITH, DW ;
RHODES, KV .
ANNALS OF EMERGENCY MEDICINE, 1993, 22 (01) :86-91
[6]   BYSTANDER CARDIOPULMONARY-RESUSCITATION - IS VENTILATION NECESSARY [J].
BERG, RA ;
KERN, KB ;
SANDERS, AB ;
OTTO, CW ;
HILWIG, RW ;
EWY, GA .
CIRCULATION, 1993, 88 (04) :1907-1915
[7]   Statistical inference for a linear function of medians: Confidence intervals, hypothesis testing, and sample size requirements [J].
Bonett, DG ;
Price, RM .
PSYCHOLOGICAL METHODS, 2002, 7 (03) :370-383
[8]  
Casner Michael, 2005, Prehosp Emerg Care, V9, P61, DOI 10.1080/10903120590891714
[9]   RECOMMENDED GUIDELINES FOR UNIFORM REPORTING OF DATA FROM OUT-OF-HOSPITAL CARDIAC-ARREST - THE UTSTEIN STYLE - A STATEMENT FOR HEALTH-PROFESSIONALS FROM A TASK-FORCE OF THE AMERICAN-HEART-ASSOCIATION, THE EUROPEAN-RESUSCITATION-COUNCIL, THE HEART-AND-STROKE-FOUNDATION-OF-CANADA, AND THE AUSTRALIAN-RESUSCITATION-COUNCIL [J].
CUMMINS, RO ;
CHAMBERLAIN, DA ;
ABRAMSON, NS ;
ALLEN, M ;
BASKETT, PJ ;
BECKER, L ;
BOSSAERT, L ;
DELOOZ, HH ;
DICK, WF ;
EISENBERG, MS ;
EVANS, TR ;
HOLMBERG, S ;
KERBER, R ;
MULLIE, A ;
ORNATO, JP ;
SANDOE, E ;
SKULBERG, A ;
TUNSTALLPEDOE, H ;
SWANSON, R ;
THIES, WH .
CIRCULATION, 1991, 84 (02) :960-975
[10]  
CUMMINS RO, 1991, ANN EMERG MED, V20, P861