Accuracy of a Portable Multisensor Body Monitor for Predicting Resting Energy Expenditure in Older People: A Comparison with Indirect Calorimetry

被引:18
|
作者
Heiermann, Stefanie [1 ]
Hedayati, Kerstin Khalaj [1 ]
Mueller, Manfred J. [2 ]
Dittmar, Manuela [1 ]
机构
[1] Univ Kiel, Inst Zool, Dept Human Biol, DE-24118 Kiel, Germany
[2] Univ Kiel, Inst Human Nutr & Food Sci, DE-24118 Kiel, Germany
关键词
Accelerometry; Motion sensor; Validity; Indirect calorimetry; Resting energy expenditure; SYMPATHETIC SKIN-RESPONSE; METABOLIC-RATE; PHYSICAL-ACTIVITY; VALIDATION; ARMBAND; VARIABILITY; OVERNIGHT; AGREEMENT; HUMANS; COST;
D O I
10.1159/000322109
中图分类号
R592 [老年病学]; C [社会科学总论];
学科分类号
03 ; 0303 ; 100203 ;
摘要
Background: Accurate and comfortable methods are needed to determine resting energy expenditure (REE) in older people who are characterized by a lowered metabolic rate. The portable SenseWear (R) armband (SWA) body monitor, worn on the right upper arm, can easily be used by this age group in an ambulatory manner. Objective: The purpose of this study was to evaluate the reliability and accuracy of the SWA armband in determining REE in healthy, normal-weight older people. Methods: Participants were 49 older, community-dwelling volunteers aged 60-87 years. Reliability was tested in all subjects. Accuracy was analyzed in 32 subjects by comparing REE estimated by SWA against REE measured by indirect calorimetry as a criterion method. Data were simultaneously and continuously recorded for 20 min in the morning. In the same subjects, REE determined by SWA from night recording was evaluated for accuracy against indirect calorimetry. The agreement between methods was assessed by the Bland-Altman procedure. Results: (1) REE measured by SWA was reliable when comparing 2 consecutive measurements (typical error: 1.9%). (2) REE calculated from morning recording (1,543 +/- 181 kcal/24 h) was higher than REE from night recording (1,564 +/- 192 kcal/24 h; p = 0.018). (3) Compared with REE by indirect calorimetry (1,377 +/- 228 kcal/24 h) for accuracy, the SWA overestimated mean REE in the older people by 12 and 14%, respectively, for morning and night recording (each p < 0.001). Conclusions: The SWA provides a reliable estimate of REE in healthy older subjects and has the advantage of easy handling. The 20-min recording time, which was recommended by the manufacturer, can be applied. However, the SWA overestimates REE in this group, possibly due to age-related changes in skin conductance and thermoregulation, both being measured by sensors of the SWA armband. This requires improving the SWA by developing better fitting algorithms for predicting REE in older people. Copyright (C) 2011 S. Karger AG, Basel
引用
收藏
页码:473 / 479
页数:7
相关论文
共 50 条
  • [1] Resting energy expenditure assessment in anorexia nervosa: comparison of indirect calorimetry, a multisensor monitor and the Muller equation
    El Ghoch, Marwan
    Alberti, Marta
    Capelli, Carlo
    Calugi, Simona
    Battistini, Nino Carlo
    Pellegrini, Massimo
    Subasic, Sandra
    Lanza, Massimo
    Dalle Grave, Riccardo
    INTERNATIONAL JOURNAL OF FOOD SCIENCES AND NUTRITION, 2012, 63 (07) : 796 - 801
  • [2] Validity of resting energy expenditure estimated by an activity monitor compared to indirect calorimetry
    Dellava, Jocilyn E.
    Hoffman, Daniel J.
    BRITISH JOURNAL OF NUTRITION, 2009, 102 (01) : 155 - 159
  • [3] Accuracy and reliability of a portable indirect calorimeter compared to whole-body indirect calorimetry for measuring resting energy expenditure
    Purcell, Sarah A.
    Johnson-Stoklossa, Carlene
    Tibaes, Jenneffer Rayane Braga
    Frankish, Alena
    Elliott, Sarah A.
    Padwal, Raj
    Prado, Carla M.
    CLINICAL NUTRITION ESPEN, 2020, 39 : 67 - 73
  • [4] Assessement of resting energy expenditure of obese patients: Comparison of indirect calorimetry with formulae
    Fabiano Alves, Valeria Girard
    Moreira da Rocha, Eduardo Eiras
    Gonzalez, Maria Cristina
    Vieira da Fonseca, Rosana Barcellos
    do Nascimento Silva, Monica Hissa
    Chiesa, Carlos Alberto
    CLINICAL NUTRITION, 2009, 28 (03) : 299 - 304
  • [5] Validation of the BIOPAC indirect calorimeter for determining resting energy expenditure in healthy free-living older people
    Khalaj-Hedayati, Kerstin
    Bosy-Westphal, Anja
    Mueller, Manfred J.
    Dittmar, Manuela
    NUTRITION RESEARCH, 2009, 29 (08) : 531 - 541
  • [6] Accuracy of a Portable Indirect Calorimeter for Measuring Resting Energy Expenditure in Individuals With Cancer
    Purcell, Sarah A.
    Elliott, Sarah A.
    Ryan, Aoife M.
    Sawyer, Michael B.
    Prado, Carla M.
    JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2019, 43 (01) : 145 - 151
  • [7] A Comparison of the Indirect Calorimetry and Different Energy Equations for the Determination of Resting Energy Expenditure of Patients With Renal Transplantation
    Tek, Nilufer Acar
    Yurtdas, Gamze
    Cemali, Ozge
    Bayazit, Ayse Derya
    Celik, Ozge Mengi
    Uyar, Gizem Ozata
    Gunes, Burcu Deniz
    Ozbas, Burak
    Erten, Yasemin
    JOURNAL OF RENAL NUTRITION, 2021, 31 (03) : 296 - 305
  • [8] Body circumferences are predictors of weight adjusted resting energy expenditure in older people
    Hedayati, K. Khalaj
    Dittmar, M.
    JOURNAL OF NUTRITION HEALTH & AGING, 2011, 15 (10) : 803 - 808
  • [9] Relative validity of an indirect calorimetry device for measuring resting energy expenditure and respiratory quotient
    Wang, Xi
    Wang, Yuan
    Ding, Zenghui
    Cao, Guangbei
    Hu, Fusong
    Sun, Yining
    Ma, Zuchang
    Zhou, Duoqi
    Su, Benyue
    ASIA PACIFIC JOURNAL OF CLINICAL NUTRITION, 2018, 27 (01) : 72 - 77
  • [10] Resting energy expenditure in cancer patients: Agreement between predictive equations and indirect calorimetry
    Barcellos, Priscila S.
    Borges, Nuno
    Torres, Duarte P. M.
    CLINICAL NUTRITION ESPEN, 2021, 42 : 286 - 291