An empirical model of diagnostic x-ray attenuation under narrow-beam geometry

被引:11
|
作者
Mathieu, Kelsey B. [1 ]
Kappadath, S. Cheenu [2 ]
White, R. Allen [3 ]
Atkinson, E. Neely [4 ]
Cody, Dianna D. [2 ]
机构
[1] Rice Univ, Dept Bioengn, Houston, TX 77005 USA
[2] Univ Texas MD Anderson Canc Ctr, Div Diagnost Imaging, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Div Bioinformat & Computat Biol, Houston, TX 77030 USA
[4] Rice Univ, Dept Stat, Houston, TX 77005 USA
关键词
CT; mammography; radiography; polyenergetic; half-value layer; HALF-VALUE LAYER; SPECTRA; CT;
D O I
10.1118/1.3592933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to develop and validate a mathematical model to describe narrow-beam attenuation of kilovoltage x-ray beams for the intended applications of half-value layer (HVL) and quarter-value layer (QVL) estimations, patient organ shielding, and computer modeling. Methods: An empirical model, which uses the Lambert W function and represents a generalized Lambert-Beer law, was developed. To validate this model, transmission of diagnostic energy x-ray beams was measured over a wide range of attenuator thicknesses [0.49-33.03 mm Al on a computed tomography (CT) scanner, 0.09-01.93 mm Al on two mammography systems, and 0.1-0.45 mm Cu and 0.49-14.87 mm Al using general radiography]. Exposure measurements were acquired under narrow-beam geometry using standard methods, including the appropriate ionization chamber, for each radiographic system. Nonlinear regression was used to find the best-fit curve of the proposed Lambert W model to each measured transmission versus attenuator thickness data set. In addition to validating the Lambert W model, we also assessed the performance of two-point Lambert W interpolation compared to traditional methods for estimating the HVL and QVL [i.e., semi-logarithmic (exponential) and linear interpolation]. Results: The Lambert W model was validated for modeling attenuation versus attenuator thickness with respect to the data collected in this study (R-2 > 0.99). Furthermore, Lambert W interpolation was more accurate and less sensitive to the choice of interpolation points used to estimate the HVL and/or QVL than the traditional methods of semilogarithmic and linear interpolation. Conclusions: The proposed Lambert W model accurately describes attenuation of both monoenergetic radiation and (kilovoltage) polyenergetic beams (under narrow-beam geometry). (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3592933]
引用
收藏
页码:4546 / 4555
页数:10
相关论文
共 50 条
  • [21] Semi-empirical model for fluorescence lines evaluation in diagnostic x-ray beams
    Bontempi, Marco
    Andreani, Lucia
    Labanti, Claudio
    Costa, Paulo Roberto
    Rossi, Pier Luca
    Baldazzi, Giuseppe
    APPLIED RADIATION AND ISOTOPES, 2016, 107 : 152 - 159
  • [22] Inverse Geometry X-Ray Fluoroscopy Using Scanning-Beam Digital X-Ray Technology
    Speidel, M.
    MEDICAL PHYSICS, 2011, 38 (06) : 3774 - +
  • [23] Biological tissue characterisation in narrow beam x-ray scattering techniques
    Tartari, A.
    De Felici, M.
    Gambaccini, M.
    Querzoli, P.
    Bonifazzi, C.
    Ferrero, C.
    Felici, R.
    X-RAY SPECTROMETRY, 2009, 38 (03) : 195 - 199
  • [24] BROAD-AND NARROW-BEAM ATTENUATION OF 500-KILOVOLT TO 1,400-KILOVOLT X-RAYS IN LEAD AND CONCRETE
    WYCKOFF, HO
    KENNEDY, RJ
    BRADFORD, WR
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1948, 41 (03): : 223 - 230
  • [25] Diagnostic x-ray spectra estimation from KVp and attenuation measurements
    Waggener, R
    Terry, J
    Wieckowska, B
    Lankipalli, B
    McDavid, W
    MEDICAL PHYSICS, 2003, 30 (06) : 1461 - 1461
  • [26] Broad beam attenuation measurements in lead in kilovoltage X-ray beams
    Akber, Syed F.
    Das, Indra J.
    Kehwar, Than S.
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2008, 18 (03): : 197 - 202
  • [27] PREVENTIVE X-RAY MEASUREMENTS AND THEIR INTERPRETATION IN X-RAY DIAGNOSTIC INCLUSIVE OF PEDIATRIC X-RAY DIAGNOSTIC
    TASCHNER, P
    MENZEL, B
    KONIG, W
    FRIEDRIC.H
    DEUTSCHE GESUNDHEITSWESEN-ZEITSCHRIFT FUR KLINISCHE MEDIZIN, 1974, 29 (31): : 1449 - 1452
  • [28] Correction for beam attenuation and indirect excitation in x-ray fluorescence holography
    Dul, D. T.
    Dabrowski, K. M.
    Korecki, P.
    EPL, 2013, 104 (06)
  • [29] X-ray attenuation models to account for beam hardening in computed tomography
    Yang, Q.
    Fullagar, W. K.
    Myers, G. R.
    Latham, S. J.
    Varslot, T.
    Sheppard, A. P.
    Kingston, A. M.
    APPLIED OPTICS, 2020, 59 (29) : 9126 - 9136
  • [30] Monitoring requirements and methods of measurement of total filtration of X-Ray beam in X-Ray diagnostic Apparatuses
    Petrushanskii M.G.
    Biomedical Engineering, 2009, 43 (1) : 20 - 23