Device interchangeability on anterior chamber depth and white -to -white measurements: a thorough literature review

被引:31
作者
Dominguez-Vicent, Alberto [1 ]
Perez-Vives, Cari [1 ]
Ferrer-Blasco, Teresa [1 ]
Garcia-Lazaro, Santiago [1 ]
Montes-Mico, Robert [1 ]
机构
[1] Univ Valencia, Dept Opt & Optometry & Visual Sci, Optometry Res Grp GIO, E-46100 Valencia, Spain
关键词
anterior camber depth measurement; white to white measurement; device interchangeability; anterior chamber eye; OPTICAL COHERENCE TOMOGRAPHY; CENTRAL CORNEAL THICKNESS; ORBSCAN-II; ULTRASOUND BIOMICROSCOPY; PENTACAM HR(R); PHAKIC EYES; IOL MASTER; IOLMASTER; REPEATABILITY; REPRODUCIBILITY;
D O I
10.18240/ijo.2016.07.20
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We have reviewed a set of recently published studies that compared the anterior chamber depth (ACD) and/or white-to-white (WTW) distance obtained by means of different measuring devices. Since some of those studies reached contradictory conclusions regarding device interchangeability, this review was carried out in attempting to clarify which clinical devices can or cannot be considered as interchangeable in clinical practice to measure ACD and/or WTW distance, among these devices: A -scan, ultrasound biomicroscopy, Orbscan and Orbscan II (Bausch&Lomb Surgical Inc., San Dimas, California, USA), Pentacam and Pentacam HR (Oculus, Wetzlar, Germany), Galilei (Ziemer, Switzerland), Visante optical coherence tomography (Visante OCT, Carl Zeiss Meditec Inc., Dublin, California, USA), IOLMaster (Carl Zeiss Meditec, Jena, Germany), and Lenstar LS 900/Biograph (Haag -Streit AG, Koeniz, Switzerland/Alcon Laboratories Inc., Ft Worth, Texas, USA).
引用
收藏
页码:1057 / 1065
页数:9
相关论文
共 32 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Orbscan computerized topography: Attributes, applications, and limitations [J].
Cairns, G ;
McGhee, CNJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :205-220
[3]   Assessment of Anterior Chamber Depth Using Visante Optical Coherence Tomography, Slitlamp Optical Coherence Tomography, IOL Master, Pentacam and Orbscan IIz [J].
Dinc, Umut Asli ;
Gorgun, Ebru ;
Oncel, Banu ;
Yenerel, Melda Nursal ;
Alimgil, Levent .
OPHTHALMOLOGICA, 2010, 224 (06) :341-346
[4]   Measurements of anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter using two Scheimpflug imaging devices [J].
Dominguez-Vicent, Alberto ;
Monsalvez-Romin, Daniel ;
Del Aguila-Carrasco, Antonio Jesus ;
Garcia-Lazaro, Santiago ;
Montes-Mico, Robert .
ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2014, 77 (04) :233-237
[5]   Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation [J].
Doors, Muriel ;
Cruysberg, Lars P. J. ;
Berendschot, Tos T. J. M. ;
de Brabander, John ;
Verbakel, Frenne ;
Webers, Carroll A. B. ;
Nuijts, Rudy M. M. A. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2009, 247 (08) :1139-1146
[6]   Comparison of different techniques of anterior chamber depth and keratometric measurements [J].
Elbaz, Uri ;
Barkana, Yaniv ;
Gerber, Yariv ;
Avni, Isaac ;
Zadok, David .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (01) :48-53
[7]   Comparison of anterior chamber depth measurements using Orbscan II and IOLMaster [J].
Frisch, I. B. ;
Rabsilber, T. M. ;
Becker, K. A. ;
Reuland, A. J. ;
Auffarth, G. U. .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2007, 17 (03) :327-331
[8]   Lenstar Versus Ultrasound for Ocular Biometry in a Pediatric Population [J].
Gursoy, Huseyin ;
Sahin, Afsun ;
Basmak, Hikmet ;
Ozer, Ahmet ;
Yildirim, Nilgun ;
Colak, Ertugrul .
OPTOMETRY AND VISION SCIENCE, 2011, 88 (08) :912-919
[9]  
Hashemi H, 2005, OPTOMETRY VISION SCI, V82, P900
[10]   Corneal changes after laser refractive surgery for myopia: Comparison of Orbscan II and Pentacam findings [J].
Hashemi, Hassan ;
Mehravaran, Shiva .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05) :841-847