Illuminated curved 25-gauge vitrectomy probe for removal of subsclerotomy vitreous in vitreoretinal surgery

被引:8
作者
Chalam, Kakarla V. [1 ]
Shah, Gaurav Y. [1 ]
Agarwal, Swati [1 ]
Gupta, Shailesh K. [1 ]
机构
[1] Univ Florida, Coll Med, Dept Ophthalmol, Jacksonville, FL 32209 USA
关键词
illuminated 25-gauge vitrectomy probe; pars plana vitrectomy; phakic; sclerotomy site;
D O I
10.4103/0301-4738.41420
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Incarceration of vitreous in sclerotomy sites during pars plana vitrectomy can lead to wound-related complications similar to vitreous incarceration in cataract surgery. We describe an illuminated curved 25-gauge vitrectomy probe for removing vitreous from sclerotomy sites. Polyester tubing is used to secure a fiber optic endoilluminator (0.5 mm) with the curved 25-gauge vitrector (0.5 mm). The resultant illuminated curved vitrector (20 G) has a diameter of 1.0 mm. It facilitates complete removal of vitreous around the internal sclerotomies under direct visualization in both phakic and pseudophakic eyes. The same was confirmed with ultrasound biomicroscopy of the sclerotomy sites. Curved vitrector reduces postoperative complications related to incarcerated vitreous in phakic and pseudophakic eyes and other sclerotomy-related wound complications.
引用
收藏
页码:331 / 334
页数:4
相关论文
共 10 条
[1]   Ultrasound biomicroscopy of sclerotomy sites after pars plana vitrectomy for diabetic vitreous hemorrhage [J].
Bhende, M ;
Agraharam, SG ;
Gopal, L ;
Sumasri, K ;
Sukumar, B ;
George, J ;
Sharma, T ;
Shanmugam, MP ;
Bhende, PS ;
Shetty, NS ;
Agrawal, RN ;
Deshpande, DA .
OPHTHALMOLOGY, 2000, 107 (09) :1729-1736
[2]   Illuminated curved vitrectomy probe for vitreoretinal surgery [J].
Chalam, K. V. ;
Gupta, Shailesh K. ;
Agarwal, Swati .
OPHTHALMIC SURGERY LASERS & IMAGING, 2007, 38 (06) :525-526
[3]   A curved vitrectomy probe [J].
Chalam, KV ;
Shah, VA ;
Tripathi, RC .
OPHTHALMIC SURGERY LASERS & IMAGING, 2004, 35 (03) :259-260
[4]   Evaluation and comparison of lens and peripheral retinal relationships with the use of endolaser probe and newly designed curved vitrectomy probe [J].
Chalam, KV ;
Shah, VA ;
Gupta, SK ;
Tripath, RC .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2003, 23 (06) :815-819
[5]  
CHARLES S, 2006, RETINA, V3, P2150
[6]   Inexpensive illuminated vitrectomy cutter [J].
Fisher, YL ;
Gross, NE ;
Wang, C .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2003, 23 (06) :891-891
[7]  
GROSS JG, 1990, ANN OPHTHALMOL, V22, P382
[8]  
PEYMAN GA, 1992, OPHTHALMIC SURG LAS, V23, P628
[9]   Ultrasound biomicroscopy of sclerotomy sites - The effect of vitreous shaving around sclerotomy sites during pars plana vitrectomy [J].
Sabti, K ;
Kapusta, M ;
Mansour, M ;
Overbury, O ;
Chow, D .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2001, 21 (05) :464-468
[10]  
Territo C, 1997, RETINA-J RET VIT DIS, V17, P430