On-line electrochemical pre-concentration of arsenic on a gold coated porous carbon electrode for graphite furnace atomic absorption spectrometry

被引:14
作者
Cacho, Frantisek [1 ]
Lauko, Lukas [1 ]
Manova, Alena [1 ]
Beinrohr, Ernest [1 ]
机构
[1] Slovak Tech Univ Bratislava, Fac Chem & Food Technol, Inst Analyt Chem, Bratislava 81237, Slovakia
关键词
RETICULATED VITREOUS CARBON; MICROCELL; MANGANESE; PALLADIUM; CADMIUM; SAMPLES; PROBE; LEAD; CELL;
D O I
10.1039/c2ja10332j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Trace arsenic was pre-concentrated from water samples in a two-electrode electrochemical flow cell coupled on-line to a graphite furnace atomic absorption spectrometer. The cell comprised a microporous electrode made of gold plated porous carbon material with a void volume of about 20 mu L. The electrodeposition was performed from hydrochloric acid media, the deposit was stripped into diluted nitric acid and the pre-concentrate was transported to a loop of about 60 mu L volume. The content of the loop was then transported by air flow to the graphite furnace and measured. Sample volumes up to 10 mu L were pre-concentrated enabling to achieve detection limits down to 0.004 mu g L-1. The procedure was validated by analyses of CRMs and various water samples. A good agreement with the hydride generation AAS was achieved for all samples.
引用
收藏
页码:695 / 699
页数:5
相关论文
共 14 条
[1]   DETERMINATION OF PLATINUM IN BIOTIC AND ENVIRONMENTAL-SAMPLES BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY AFTER ITS ELECTRODEPOSITION INTO A GRAPHITE TUBE PACKED WITH RETICULATED VITREOUS CARBON [J].
BEINROHR, E ;
LEE, ML ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (6-9) :689-692
[2]   ONLINE ELECTROCHEMICAL PRECONCENTRATION OF MANGANESE FOR GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY USING A FLOW-THROUGH ELECTROCHEMICAL-CELL [J].
BEINROHR, E ;
RAPTA, M ;
LEE, ML ;
TSCHOPEL, P ;
TOLG, G .
MIKROCHIMICA ACTA, 1993, 110 (1-3) :1-12
[3]   ELECTROLYTIC SAMPLE PRETREATMENT IN ATOMIC SPECTROMETRY - A REVIEW [J].
BEINROHR, E .
MIKROCHIMICA ACTA, 1995, 120 (1-4) :39-52
[4]   DETERMINATION OF LEAD BY ELECTROTHERMAL VAPORIZATION MICROWAVE-INDUCED PLASMA-ATOMIC EMISSION-SPECTROMETRY AFTER FLOW-THROUGH ELECTROLYTIC DEPOSITION IN A GRAPHITE TUBE PACKED WITH RETICULATED VITREOUS CARBON [J].
BEINROHR, E ;
BULSKA, E ;
TSCHOPEL, P ;
TOLG, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (07) :965-968
[5]   Analytical advantages of using electrochemistry for atomic spectrometry [J].
Bulska, E .
PURE AND APPLIED CHEMISTRY, 2001, 73 (01) :1-7
[6]   On-line preconcentration of lead and cadmium for flame atomic absorption spectrometry using a flow-through electrochemical microcell [J].
Bulska, E ;
Walcerz, M ;
Jedral, W ;
Hulanicki, A .
ANALYTICA CHIMICA ACTA, 1997, 357 (1-2) :133-140
[7]   On-line electrochemical preconcentration and flame atomic absorption spectrometric determination of manganese in urine samples [J].
Burguera, M ;
Burguera, JL ;
Rivas, D ;
Rondón, C ;
Carrero, P ;
Alarcón, OM ;
de Peña, YP ;
Brunetto, MR ;
Gallignani, M ;
Márquez, OP ;
Márquez, J .
TALANTA, 2005, 68 (02) :219-225
[8]   Preconcentration of palladium in a flow-through electrochemical cell for determination by graphite furnace atomic absorption spectrometry [J].
Godlewska-Zylkiewicz, B ;
Zaleska, M .
ANALYTICA CHIMICA ACTA, 2002, 462 (02) :305-312
[9]   Analytical methods for inorganic arsenic in water: a review [J].
Hung, DQ ;
Nekrassova, O ;
Compton, RG .
TALANTA, 2004, 64 (02) :269-277
[10]   Determination of cadmium by electrothermal atomic absorption spectrometry using electrochemical separation in a microcell [J].
Knápek, J ;
Komárek, J ;
Krásensky, P .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (03) :393-398