Comparison of radius of anterior lens surface curvature measurements in vivo using the anterior segment optical coherence tomography and Scheimpflug imaging

被引:14
作者
Liu, Zhenzhen [1 ]
Ruan, Xiaoting [1 ]
Wang, Wei [1 ]
Liu, Juzi [1 ]
Meng, Yijing [1 ]
Gu, Xiaoxun [1 ]
Fu, Jun [1 ]
Luo, Lixia [1 ]
Liu, Yizhi [1 ]
机构
[1] Sun Yat Sen Univ, Zhongshan Ophthalm Ctr, State Key Lab Ophthalmol, Guangzhou 510060, Peoples R China
基金
中国国家自然科学基金;
关键词
Crystalline lens; radius of anterior lens surface curvature (RAL); Scheimpflug Camera; anterior segment optical coherence tomography (AS-OCT); ACCOMMODATIVE CHANGES; REPEATABILITY; PACHYMETRY; PENTACAM; SYSTEM; EYE; AGE;
D O I
10.21037/atm.2020.01.100
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: To assess the radius of anterior lens surface curvature (RAL) measurements with anterior segment optical coherence tomography (AS-OCT) in comparison with Scheimpflug imaging. Methods: This prospective, cross-sectional study was carried out at Zhongshan Ophthalmic Center, Guangzhou, China. We enrolled 59 eyes, including 30 eyes from 30 cataractous volunteers (59 to 87 years) and 29 eyes from 29 young participants (19 to 49 years). After mydriasis, the RAL was measured automatically by the built-in software in the AS-OCT (CASIA 2). The Scheimpflug images were measured with the build-in caliper tool of the Scheimpflug camera (Pentacam), and RAL were further calculated with the principle of best-fitted circle. Intraobserver and interobserver reproducibility of RAL measurement using Scheimpflug camera were evaluated with limit of agreement (LoA) and intradass correlation coefficient (ICC). Consistency between RAL measurement of Scheimpflug camera and AS-OCT were assessed with LoA, correlation analysis and linear regression. Results: For all subjects, intraobserver (LoA: -0.25 to 0.23 mm, ICC: 0.996) and interobserver reproducibility (LoA: -0.85 to 0.92 mm, ICC: 0.947) of RAL were good using Scheimpflug imaging. Both AS-OCT and Scheimpflug imaging found that the age-related cataract participants had smaller RAL (P=0.010, P=0.001 respectively). LoA of RAL measurement between AS-OCT and Scheimpflug imaging was -3.83 to -0.79 mm, and the Pearson correlation efficient was 0.909 (P<0.001). The RAL values measured by AS-OCT were significantly greater than that by Scheimpflug camera with a mean difference of 2.31 mm for all participants (P4.001). The RAL measurement could be converted using the equation: Y-CASIA 2 =1.155 x X-Pentacam + 1.060. Conclusions: Both Scheimpflug camera system with internal caliper tool and the AS-OCT are fast and non-contact tools that could measure RAL successfully. The two measurement results are highly correlated and interchangeable through linear regression equation.
引用
收藏
页数:9
相关论文
共 32 条
[11]   Comparison of Central Corneal Thickness Measured by Standard Ultrasound Pachymetry, Corneal Topography, Tono-Pachymetry and Anterior Segment Optical Coherence Tomography [J].
Gonzalez-Perez, Javier ;
Queiruga Pineiro, Juan ;
Sanchez Garcia, Angelx ;
Gonzalez Meijome, Jose Manuel .
CURRENT EYE RESEARCH, 2018, 43 (07) :866-872
[12]   Constant Volume of the Human Lens and Decrease in Surface Area of the Capsular Bag during Accommodation: An MRI and Scheimpflug Study [J].
Hermans, Erik A. ;
Pouwels, Petra J. W. ;
Dubbelman, Michiel ;
Kuijer, Joost P. A. ;
van der Heijde, Rob G. L. ;
Heethaar, Rob M. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (01) :281-289
[13]   Scheimpflug and high-resolution magnetic resonance imaging of the anterior segment: a comparative study [J].
Koretz, JF ;
Strenk, SA ;
Strenk, LM ;
Semmlow, JL .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2004, 21 (03) :346-354
[15]   A Novel Congenital Cataract Category System Based on Lens Opacity Locations and Relevant Anterior Segment Characteristics [J].
Lin, Haotian ;
Lin, Duoru ;
Liu, Zhenzhen ;
Long, Erping ;
Wu, Xiaohang ;
Cao, Qianzhong ;
Chen, Jingjing ;
Lin, Zhuoling ;
Li, Xiaoyan ;
Zhang, Li ;
Chen, Hui ;
Zhang, Xiayin ;
Li, Jing ;
Chen, Weirong ;
Liu, Yizhi .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (14) :6389-6395
[16]   Radius of curvature and asphericity of the anterior and posterior surface of human cadaver crystalline lenses [J].
Manns, F ;
Fernandez, V ;
Zipper, S ;
Sandadi, S ;
Hamaoui, M ;
Ho, A ;
Parel, JM .
EXPERIMENTAL EYE RESEARCH, 2004, 78 (01) :39-51
[17]   Axial Ultrasound B-Scans of the Entire Eye With a 20-MHz Linear Array: Correction of Crystalline Lens Phase Aberration by Applying Fermat's Principle [J].
Mateo, Tony ;
Chang, Alexandre ;
Mofid, Yassine ;
Pisella, Pierre-Jean ;
Ossant, Frederic .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (11) :2149-2166
[18]  
Oatts JT, 2017, CORNEA, V36, P425, DOI 10.1097/ICO.0000000000001151
[19]   Optical distortion correction in Optical Coherence Tomography for quantitative ocular anterior segment by three-dimensional imaging [J].
Ortiz, Sergio ;
Siedlecki, Damian ;
Grulkowski, Ireneusz ;
Remon, Laura ;
Pascual, Daniel ;
Wojtkowski, Maciej ;
Marcos, Susana .
OPTICS EXPRESS, 2010, 18 (03) :2782-2796
[20]   Lens density tracking in mice by Scheimpflug imaging [J].
Puk, Oliver ;
de Angelis, Martin Hrabe ;
Graw, Jochen .
MAMMALIAN GENOME, 2013, 24 (7-8) :295-302