Test/Retest Reliability of Maximum Mouth Pressure Measurements With the MicroRPM in Healthy Volunteers

被引:89
作者
Dimitriadis, Zacharias [1 ,2 ]
Kapreli, Eleni [1 ]
Konstantinidou, Ioanna [1 ]
Oldham, Jacqueline [2 ]
Strimpakos, Nikolaos [1 ,2 ]
机构
[1] Technol Educ Inst Lamia, Physiotherapy Dept, Lamia, Greece
[2] Univ Manchester, Ctr Rehabil Sci, Manchester, Lancs, England
关键词
respiratory muscle; lung function; maximum expiratory pressure; maximum inspiratory pressure; manometer; reliability; RESPIRATORY MUSCLE FUNCTION; INSPIRATORY PRESSURE; NORMAL VALUES; REPRODUCIBILITY; DIAPHRAGM;
D O I
10.4187/respcare.00783
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Respiratory muscle strength is an important part of lung function. Assessment of the respiratory muscles' ability to generate force is important for recognizing respiratory muscle weakness in both sick and healthy people. OBJECTIVE: To assess the test/retest reliability of the MicroRPM portable manometer's measurements of maximum inspiratory pressure (MIP) and maximum expiratory pressure (MEP) in the sitting and standing positions; the number of expiratory maneuvers needed with the MicroRPM for reliability in MIP and MEP measurement; and the MicroRPM's test/retest reliability in other respiratory function indices, such as the maximum rate of pressure development (MRPD), the time constant of relaxation (tau), and the maximum relaxation rate (MRR). METHODS: We recruited 15 healthy volunteers (mean age 21.6 +/- 1.1 years). We assessed respiratory muscle strength on 3 separate occasions, each a week apart. We calculated reliability with the intraclass correlation coefficient (ICC), the standard error of measurement (SEM) and the smallest detectable difference (SDD). RESULTS: MicroRPM reliably measured MIP and MEP in both the sitting position (ICC 0.86-0.90, SEM 9-10, SDD 18-22) and standing position (ICC 0.78-0.83, SEM 12-14, SDD 23-26). After a 5-breath practice, 2 expiratory/inspiratory maneuvers on each testing occasion gave adequate MEP and MEP reliability (ICC > 0.90). MRR reliability was moderate to excellent (ICC 0.58-0.87), MRPD reliability was moderate (ICC 0.59-0.64), and tau reliability was insufficient (ICC 0.27-0.67). CONCLUSIONS: The MicroRPM reliably measures MIP and MEP, but its MRPD, MRR, and tau measurements should be considered with caution.
引用
收藏
页码:776 / 782
页数:7
相关论文
共 50 条
[41]   Test–retest reliability of Kinect’s measurements for the evaluation of upper body recovery of stroke patients [J].
A Mobini ;
S Behzadipour ;
M Saadat .
BioMedical Engineering OnLine, 14
[42]   Can Resting-State Functional MRI Serve as a Complement to Task-Based Mapping of Sensorimotor Function? A Test-Retest Reliability Study in Healthy Volunteers [J].
Mannfolk, Peter ;
Nilsson, Markus ;
Hansson, Henrik ;
Stahlberg, Freddy ;
Fransson, Peter ;
Weibull, Andreas ;
Svensson, Jonas ;
Wirestam, Ronnie ;
Olsrud, Johan .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 34 (03) :511-517
[43]   Test-Retest Reliability of the SERT Imaging Agent 11C-HOMADAM in Healthy Humans [J].
Adhikarla, Vikram ;
Dunlop, Boadie W. ;
Jarkas, Nashwa ;
Goodman, Mark M. ;
Mayberg, Helen ;
Owens, Michael J. ;
Nye, Jonathon A. .
JOURNAL OF NUCLEAR MEDICINE, 2018, 59 (02) :315-319
[44]   Test-retest reproducibility of 18F-MPPF PET in healthy humans:: A reliability study [J].
Costes, Nicolas ;
Zimmer, Luc ;
Reilhac, Anthonin ;
Lavenne, Franck ;
Ryvlin, Philippe ;
Le Bars, Didier .
JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (08) :1279-1288
[45]   Reliability of lumbar dynamometry measurements in patients with chronic low back pain with test-retest measurements on different days [J].
Hutten M.M.R. ;
Hermens H.J. .
European Spine Journal, 1997, 6 (1) :54-62
[46]   Test-retest reliability of single and paired pulse transcranial magnetic stimulation parameters in healthy subjects [J].
Hermsen, A. M. ;
Haag, A. ;
Duddek, C. ;
Balkenhol, K. ;
Bugiel, H. ;
Bauer, S. ;
Mylius, V. ;
Menzler, K. ;
Rosenow, F. .
JOURNAL OF THE NEUROLOGICAL SCIENCES, 2016, 362 :209-216
[47]   Impact of Potential Moderating Factors on Absolute Test-Retest Reliability of Grip Strength Measurements in Healthy Populations: A Systematic Review with Meta-Analysis [J].
Abe, Takashi ;
Song, Jun Seob ;
Dankel, Scott J. ;
Viana, Ricardo B. ;
Abe, Akemi ;
Loenneke, Jeremy P. .
JOURNAL OF SPORTS SCIENCE AND MEDICINE, 2025,
[48]   TEST-RETEST RELIABILITY OF TWO CLINICAL TESTS FOR THE ASSESSMENT OF HIP ABDUCTOR ENDURANCE IN HEALTHY FEMALES [J].
van Cant, Joachim ;
Dumont, Gregory ;
Pitance, Laurent ;
Demoulin, Christophe ;
Feipel, Veronique .
INTERNATIONAL JOURNAL OF SPORTS PHYSICAL THERAPY, 2016, 11 (01) :24-33
[49]   Test-retest reliability of muscular performance tests and compression garment interface pressure measurements: a comparison between consecutive and multiple day recovery [J].
Brown, Freddy ;
Hill, Mathew ;
Renshaw, Derek ;
Pedlar, Charles ;
Hill, Jessica ;
Tallis, Jason .
SPORTS ENGINEERING, 2023, 26 (01)
[50]   Test-retest reliability and standard error of measurement for the test of variables of attention (TOVA) with healthy school-age children [J].
Leark, RA ;
Wallace, DR ;
Fitzgerald, R .
ASSESSMENT, 2004, 11 (04) :285-289