Test/Retest Reliability of Maximum Mouth Pressure Measurements With the MicroRPM in Healthy Volunteers

被引:89
作者
Dimitriadis, Zacharias [1 ,2 ]
Kapreli, Eleni [1 ]
Konstantinidou, Ioanna [1 ]
Oldham, Jacqueline [2 ]
Strimpakos, Nikolaos [1 ,2 ]
机构
[1] Technol Educ Inst Lamia, Physiotherapy Dept, Lamia, Greece
[2] Univ Manchester, Ctr Rehabil Sci, Manchester, Lancs, England
关键词
respiratory muscle; lung function; maximum expiratory pressure; maximum inspiratory pressure; manometer; reliability; RESPIRATORY MUSCLE FUNCTION; INSPIRATORY PRESSURE; NORMAL VALUES; REPRODUCIBILITY; DIAPHRAGM;
D O I
10.4187/respcare.00783
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Respiratory muscle strength is an important part of lung function. Assessment of the respiratory muscles' ability to generate force is important for recognizing respiratory muscle weakness in both sick and healthy people. OBJECTIVE: To assess the test/retest reliability of the MicroRPM portable manometer's measurements of maximum inspiratory pressure (MIP) and maximum expiratory pressure (MEP) in the sitting and standing positions; the number of expiratory maneuvers needed with the MicroRPM for reliability in MIP and MEP measurement; and the MicroRPM's test/retest reliability in other respiratory function indices, such as the maximum rate of pressure development (MRPD), the time constant of relaxation (tau), and the maximum relaxation rate (MRR). METHODS: We recruited 15 healthy volunteers (mean age 21.6 +/- 1.1 years). We assessed respiratory muscle strength on 3 separate occasions, each a week apart. We calculated reliability with the intraclass correlation coefficient (ICC), the standard error of measurement (SEM) and the smallest detectable difference (SDD). RESULTS: MicroRPM reliably measured MIP and MEP in both the sitting position (ICC 0.86-0.90, SEM 9-10, SDD 18-22) and standing position (ICC 0.78-0.83, SEM 12-14, SDD 23-26). After a 5-breath practice, 2 expiratory/inspiratory maneuvers on each testing occasion gave adequate MEP and MEP reliability (ICC > 0.90). MRR reliability was moderate to excellent (ICC 0.58-0.87), MRPD reliability was moderate (ICC 0.59-0.64), and tau reliability was insufficient (ICC 0.27-0.67). CONCLUSIONS: The MicroRPM reliably measures MIP and MEP, but its MRPD, MRR, and tau measurements should be considered with caution.
引用
收藏
页码:776 / 782
页数:7
相关论文
共 50 条
[31]   Acoustic Stapedial Reflexes in Healthy Neonates: Normative Data and Test-Retest Reliability [J].
Kei, Joseph .
JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2012, 23 (01) :46-56
[32]   The test-retest reliability of centre of pressure measures in bipedal static task conditions - A systematic review of the literature [J].
Ruhe, Alexander ;
Fejer, Rene ;
Walker, Bruce .
GAIT & POSTURE, 2010, 32 (04) :436-445
[33]   Test-Retest Reliability of "High-Order" Functional Connectivity in Young Healthy Adults [J].
Zhang, Han ;
Chen, Xiaobo ;
Zhang, Yu ;
Shen, Dinggang .
FRONTIERS IN NEUROSCIENCE, 2017, 11
[34]   Intratest reliability and test-retest reproducibility of the oxygen uptake efficiency slope in healthy participants [J].
Van Laethem, Christophe ;
De Sutter, Johan ;
Peersman, Wim ;
Calders, Patrick .
EUROPEAN JOURNAL OF CARDIOVASCULAR PREVENTION & REHABILITATION, 2009, 16 (04) :493-498
[35]   Test-retest reliability of the isernhagen work systems functional capacity evaluation in healthy adults [J].
Reneman, MF ;
Brouwer, S ;
Meinema, A ;
Dijkstra, PU ;
Geertzen, JHB ;
Groothoff, JW .
JOURNAL OF OCCUPATIONAL REHABILITATION, 2004, 14 (04) :295-305
[36]   Test-retest reliability of a balance testing protocol with external perturbations in young healthy adults [J].
Robbins, Shawn M. ;
Caplan, Ryan M. ;
Aponte, Daniel I. ;
St-Onge, Nancy .
GAIT & POSTURE, 2017, 58 :433-439
[37]   Test–retest reliability of muscular performance tests and compression garment interface pressure measurements: a comparison between consecutive and multiple day recovery [J].
Freddy Brown ;
Mathew Hill ;
Derek Renshaw ;
Charles Pedlar ;
Jessica Hill ;
Jason Tallis .
Sports Engineering, 2023, 26
[38]   Test-retest reliability of EMG and peak torque during repetitive maximum concentric knee extensions [J].
Larsson, B ;
Karlsson, S ;
Eriksson, A ;
Gerdle, B .
JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2003, 13 (03) :281-287
[39]   Test-retest, intra- and inter-rater reliability of the reactive balance test in healthy recreational athletes [J].
Tassignon, Bruno ;
Verschueren, Jo ;
De Wachter, Jonas ;
Maricot, Alexandre ;
De Pauw, Kevin ;
Verhagen, Evert ;
Meeusen, Romain .
PHYSICAL THERAPY IN SPORT, 2020, 46 :47-53
[40]   Reliability of a standardized test in Swedish for evaluation of reading performance in healthy eyes. Interchart and test-retest analyses [J].
Thaung, Jorgen ;
Olseke, Kjell ;
Ahl, Johan ;
Sjostrand, Johan .
ACTA OPHTHALMOLOGICA, 2014, 92 (06) :557-562