Test/Retest Reliability of Maximum Mouth Pressure Measurements With the MicroRPM in Healthy Volunteers

被引:89
作者
Dimitriadis, Zacharias [1 ,2 ]
Kapreli, Eleni [1 ]
Konstantinidou, Ioanna [1 ]
Oldham, Jacqueline [2 ]
Strimpakos, Nikolaos [1 ,2 ]
机构
[1] Technol Educ Inst Lamia, Physiotherapy Dept, Lamia, Greece
[2] Univ Manchester, Ctr Rehabil Sci, Manchester, Lancs, England
关键词
respiratory muscle; lung function; maximum expiratory pressure; maximum inspiratory pressure; manometer; reliability; RESPIRATORY MUSCLE FUNCTION; INSPIRATORY PRESSURE; NORMAL VALUES; REPRODUCIBILITY; DIAPHRAGM;
D O I
10.4187/respcare.00783
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Respiratory muscle strength is an important part of lung function. Assessment of the respiratory muscles' ability to generate force is important for recognizing respiratory muscle weakness in both sick and healthy people. OBJECTIVE: To assess the test/retest reliability of the MicroRPM portable manometer's measurements of maximum inspiratory pressure (MIP) and maximum expiratory pressure (MEP) in the sitting and standing positions; the number of expiratory maneuvers needed with the MicroRPM for reliability in MIP and MEP measurement; and the MicroRPM's test/retest reliability in other respiratory function indices, such as the maximum rate of pressure development (MRPD), the time constant of relaxation (tau), and the maximum relaxation rate (MRR). METHODS: We recruited 15 healthy volunteers (mean age 21.6 +/- 1.1 years). We assessed respiratory muscle strength on 3 separate occasions, each a week apart. We calculated reliability with the intraclass correlation coefficient (ICC), the standard error of measurement (SEM) and the smallest detectable difference (SDD). RESULTS: MicroRPM reliably measured MIP and MEP in both the sitting position (ICC 0.86-0.90, SEM 9-10, SDD 18-22) and standing position (ICC 0.78-0.83, SEM 12-14, SDD 23-26). After a 5-breath practice, 2 expiratory/inspiratory maneuvers on each testing occasion gave adequate MEP and MEP reliability (ICC > 0.90). MRR reliability was moderate to excellent (ICC 0.58-0.87), MRPD reliability was moderate (ICC 0.59-0.64), and tau reliability was insufficient (ICC 0.27-0.67). CONCLUSIONS: The MicroRPM reliably measures MIP and MEP, but its MRPD, MRR, and tau measurements should be considered with caution.
引用
收藏
页码:776 / 782
页数:7
相关论文
共 50 条
[21]   Test-retest reliability of isometric mid-thigh pull maximum strength assessment: a systematic review [J].
Grgic, Jozo ;
Scapec, Bela ;
Mikulic, Pavle ;
Pedisic, Zeljko .
BIOLOGY OF SPORT, 2022, 39 (02) :407-414
[22]   Test-Retest Reliability of Isokinetic Strength Measurements in Lower Limbs in Elderly [J].
Parraca, Jose A. ;
Carmelo Adsuar, Jose ;
Javier Dominguez-Munoz, Francisco ;
Barrios-Fernandez, Sabina ;
Tomas-Carus, Pablo .
BIOLOGY-BASEL, 2022, 11 (06)
[23]   The inter-rater and test-retest reliability of skin thickness and skin elasticity measurements by the DermaLab Combo in healthy participants [J].
Peperkamp, Kirsten ;
Verhulst, Arico C. ;
Tielemans, Hanneke J. P. ;
Winters, Harm ;
van Dalen, Demi ;
Ulrich, Dietmar J. O. .
SKIN RESEARCH AND TECHNOLOGY, 2019, 25 (06) :787-792
[24]   Test-retest reliability of clinical measurements of lower extremity joint motion and alignment in the pediatric population [J].
Saabye, Thea ;
Colding-Rasmussen, Thomas ;
Balslev-Clausen, Andreas ;
Bodtker, Soren ;
Wong, Christian ;
Harsted, Steen .
JOURNAL OF CHILDRENS ORTHOPAEDICS, 2025, 19 (03) :189-198
[25]   INTERRATER AND TEST-RETEST RELIABILITY OF THE Y BALANCE TEST IN HEALTHY, EARLY ADOLESCENT FEMALE ATHLETES [J].
Greenberg, Eric T. ;
Bark, Matthew ;
Glassmann, Erica ;
Jung, Min-Kyung .
INTERNATIONAL JOURNAL OF SPORTS PHYSICAL THERAPY, 2019, 14 (02) :204-213
[26]   Test-retest reliability of the lateral step-up test in young adult healthy subjects [J].
Ross, CM .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 1997, 25 (02) :128-132
[27]   Test-retest reliability of Quantitative Sensory Testing in knee osteoarthritis and healthy participants [J].
Wylde, V. ;
Palmer, S. ;
Learmonth, I. D. ;
Dieppe, P. .
OSTEOARTHRITIS AND CARTILAGE, 2011, 19 (06) :655-658
[28]   Test-retest reliability of joint position sense in the elbow among healthy adults [J].
V. Rider, John ;
Valdes, Kristin A. .
JOURNAL OF HAND THERAPY, 2024, 37 (02) :243-249
[29]   Test-retest reliability of a finger-tapping fMRI task in a healthy population [J].
Wuethrich, Florian ;
Lefebvre, Stephanie ;
Nadesalingam, Niluja ;
Bernard, Jessica A. ;
Mittal, Vijay A. ;
Shankman, Stewart A. ;
Walther, Sebastian .
EUROPEAN JOURNAL OF NEUROSCIENCE, 2023, 57 (01) :78-90
[30]   Test–Retest Reliability of the Isernhagen Work Systems Functional Capacity Evaluation in Healthy Adults [J].
M.F. Reneman ;
S. Brouwer ;
A. Meinema ;
P.U. Dijkstra ;
J.H.B. Geertzen ;
J.W. Groothoff .
Journal of Occupational Rehabilitation, 2004, 14 :295-305