CT number definition

被引:32
作者
Bryant, J. A. [1 ]
Drage, N. A. [1 ]
Richmond, S. [1 ]
机构
[1] Cardiff Univ, Sch Dent, Cardiff CF14 4XY, S Glam, Wales
关键词
CT; X-ray physics; 3D imaging; CONE-BEAM CT; RECONSTRUCTION;
D O I
10.1016/j.radphyschem.2011.12.026
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The accuracy of CT number plots has been found lacking in several medical applications. This is of concern since the ability to compare and evaluate results on a reproducible and standard basis is essential to long term development. Apart from the technical limitations arising from the CT scanner and the data treatment, there are fundamental issues with the definition of the Hounsfield number, namely the absence of a standard photon energy and the need to specify the attenuation mechanism for standard measurements. This paper presents calculations to demonstrate the shortcomings of the present definition with a brief discussion. The remedy is straightforward, but probably of long duration as it would require an international agreement. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:358 / 361
页数:4
相关论文
共 13 条
[1]   BEAM HARDENING IN X-RAY RECONSTRUCTIVE TOMOGRAPHY [J].
BROOKS, RA ;
DICHIRO, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (03) :390-398
[2]   Study of the scan uniformity from an i-CAT cone beam computed tomography dental imaging system [J].
Bryant, J. A. ;
Drage, N. A. ;
Richmond, S. .
DENTOMAXILLOFACIAL RADIOLOGY, 2008, 37 (07) :365-374
[3]   Reconstruction of multi-energy x-ray computed tomography images of laboratory mice [J].
Gleason, SS ;
Sari-Sarraf, H ;
Paulus, MJ ;
Johnson, DK ;
Norton, SJ ;
Abidi, MA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (04) :1081-1086
[4]   CT number variations due to different image acquisition and reconstruction parameters: a thorax phantom study [J].
Groell, R ;
Rienmueller, R ;
Schaffler, GJ ;
Portugaller, HR ;
Graif, E ;
Willfurth, P .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2000, 24 (02) :53-58
[5]  
Kachelriess M, 2005, IEEE NUCL SCI CONF R, P1905
[6]   Density conversion factor determined using a cone-beam computed tomography unit NewTom QR-DVT 9000 [J].
Lagravere, M. O. ;
Fang, Y. ;
Carey, J. ;
Toogood, R. W. ;
Packota, G. V. ;
Major, P. W. .
DENTOMAXILLOFACIAL RADIOLOGY, 2006, 35 (06) :407-409
[7]   THE UNRELIABILITY OF CT-NUMBERS AS ABSOLUTE VALUES [J].
LEVI, C ;
GRAY, JE ;
MCCULLOUGH, EC ;
HATTERY, RR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 139 (03) :443-447
[8]   Assessment of bone segmentation quality of cone-beam CT versus multislice spiral CT: a pilot study [J].
Loubele, Miet ;
Maes, Frederik ;
Schutyser, Filip ;
Marchal, Guy ;
Jacobs, Reinhilde ;
Suetens, Paul .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2006, 102 (02) :225-234
[9]   Development of segmented semiconductor arrays for quantum imaging [J].
Mikulec, B .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 510 (1-2) :1-23
[10]   A design of mammography units using a quasimonochromatic x-ray source [J].
Piestrup, MA ;
Wu, XZ ;
Kaplan, VV ;
Uglov, SR ;
Cremer, JT ;
Rule, DW ;
Fiorito, RB .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (04) :2159-2170