Experimental Stress Characterization of a Biomedical Ultrasound Probe Soldered With Innovative Silver Isotropically Conductive Adhesive

被引:6
作者
Catelani, Marcantonio [1 ]
Scarano, Valeria L. [1 ]
Bertocci, Francesco [1 ]
机构
[1] Univ Florence, Dept Elect & Telecommun, I-50139 Florence, Italy
关键词
Accelerated qualification testing (AQT); electroacoustic characterization; phased-array probe; pulse stress test; random vibration test; silver conductive adhesive; ultrasonic instrumentation; PHASED-ARRAY PROBE; MHZ; TRANSDUCERS; COMPOSITE; FILLERS;
D O I
10.1109/TIM.2011.2170370
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In the coming years, biomedical instrument manufacturers will have to inform the Notified Body of RoHS conformity assessment of the new medical instruments. In particular, in the case of lead-free solders, the Notified Body will require evidence that such solders are equally reliable as tin-lead, for the proposed use and lifetime of the product. So, biomedical instrument manufacturers are strongly interested in evaluating the performance of new soldering lead-free material used for the implementation of electronic parts. In order to support them and fill the technology knowledge gap, this paper presents a novel 128-channel phased-array probe for echocardiograph applications realized by new soldering technique where isotropic silver electrically conductive adhesive is used. In particular, the new soldering material is taken into consideration in the design phase of ultrasound array transducer, which is the core of a linear phased-array probe that is able to monitor the cardiac muscle. Being such material innovative in the application, a detailed study of the failure condition regarding the phased-array probe by means of accelerated qualification testing is proposed in order to assure its performances under stress and over time. A comparison of the electroacoustic measurement results with respect to the conventional probe, soldered with Sn-Pb, is also presented in this paper.
引用
收藏
页码:719 / 728
页数:10
相关论文
共 40 条
[31]   Measurement issues in quantitative ultrasonic imaging [J].
Siegel, M .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1998, 47 (06) :1435-1438
[32]   An industry vision on the implementation of WEEE and RoHS [J].
Stevels, A ;
Huisman, J .
2003 3RD INTERNATIONAL SYMPOSIUM ON ENVIRONMENTALLY CONSCIOUS DESIGN AND INVERSE MANUFACTURING - ECODESIGN '03, 2003, :9-16
[33]  
Suhir E., 2011, P PRINT CIRC DES FAB, P26
[34]  
White F.M., 2006, Viscous Fluid Flow, V3rd
[35]   Adhesive joining process and joint property with low melting point filler [J].
Yasuda, K ;
Kim, JM ;
Fujimoto, K .
JOURNAL OF ELECTRONIC PACKAGING, 2005, 127 (01) :12-17
[36]  
Yasuda K., 2003, P INT C EL PACK, P149
[37]   A 120-MHz ultrasound probe for tissue imaging [J].
Yokosawa, K ;
Shinomura, R ;
Sano, S ;
Ito, Y ;
Ishikawa, S ;
Sato, Y .
ULTRASONIC IMAGING, 1996, 18 (04) :231-239
[38]   High frequency broadband PZT thick film ultrasonic transducers for medical imaging applications [J].
Zhang, Q. Q. ;
Djuth, F. T. ;
Zhou, Q. F. ;
Hu, C. H. ;
Cha, J. H. ;
Shung, K. K. .
ULTRASONICS, 2006, 44 :E711-E715
[39]   Piezoceramics for high-frequency (20 to 100 MHz) single-element imaging transducers [J].
Zipparo, MJ ;
Shung, KK ;
Shrout, TR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1997, 44 (05) :1038-1048
[40]   Wideband high-frequency line-focus PVDF transducer for materials characterization [J].
Zou, W ;
Holland, S ;
Kim, KY ;
Sachse, W .
ULTRASONICS, 2003, 41 (03) :157-161