Experimental Stress Characterization of a Biomedical Ultrasound Probe Soldered With Innovative Silver Isotropically Conductive Adhesive

被引:6
作者
Catelani, Marcantonio [1 ]
Scarano, Valeria L. [1 ]
Bertocci, Francesco [1 ]
机构
[1] Univ Florence, Dept Elect & Telecommun, I-50139 Florence, Italy
关键词
Accelerated qualification testing (AQT); electroacoustic characterization; phased-array probe; pulse stress test; random vibration test; silver conductive adhesive; ultrasonic instrumentation; PHASED-ARRAY PROBE; MHZ; TRANSDUCERS; COMPOSITE; FILLERS;
D O I
10.1109/TIM.2011.2170370
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In the coming years, biomedical instrument manufacturers will have to inform the Notified Body of RoHS conformity assessment of the new medical instruments. In particular, in the case of lead-free solders, the Notified Body will require evidence that such solders are equally reliable as tin-lead, for the proposed use and lifetime of the product. So, biomedical instrument manufacturers are strongly interested in evaluating the performance of new soldering lead-free material used for the implementation of electronic parts. In order to support them and fill the technology knowledge gap, this paper presents a novel 128-channel phased-array probe for echocardiograph applications realized by new soldering technique where isotropic silver electrically conductive adhesive is used. In particular, the new soldering material is taken into consideration in the design phase of ultrasound array transducer, which is the core of a linear phased-array probe that is able to monitor the cardiac muscle. Being such material innovative in the application, a detailed study of the failure condition regarding the phased-array probe by means of accelerated qualification testing is proposed in order to assure its performances under stress and over time. A comparison of the electroacoustic measurement results with respect to the conventional probe, soldered with Sn-Pb, is also presented in this paper.
引用
收藏
页码:719 / 728
页数:10
相关论文
共 40 条
[1]  
[Anonymous], 2003, OFFICIAL J EUR UNION, V46, pL37/24
[2]  
[Anonymous], 2003, OFFICIAL J EUR UNION, V46, pL37/19
[3]  
[Anonymous], 2010, 6060119 IEC
[4]  
Brown LF, 1997, 1997 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 & 2, P1725, DOI 10.1109/ULTSYM.1997.663344
[5]   Design of efficient, broadband single-element (20-80 MHz) ultrasonic transducers for medical imaging applications [J].
Cannata, JM ;
Ritter, TA ;
Chen, WH ;
Silverman, RH ;
Shung, KK .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (11) :1548-1557
[6]  
Catelani M., 2011, IEEE T COMPON PACKAG, V1
[7]  
Catelani M., 2010, P INT WORKSH MEMEA, P81
[8]   Implementation and Characterization of a Medical Ultrasound Phased Array Probe With New Pb-Free Soldering Materials [J].
Catelani, Marcantonio ;
Scarano, Valeria L. ;
Bertocci, Francesco .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2010, 59 (10) :2522-2529
[9]   On the optimization of the transmitted beam in contrast-enhanced ultrasound medical imaging [J].
Curletto, Simone ;
Palmese, Maria ;
Trucco, Andrea .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2007, 56 (04) :1239-1248
[10]  
Das RN, 2007, ELEC COMP C, P74