A deformable phantom for 4D radiotherapy verification: Design and image registration evaluation

被引:86
作者
Serban, Monica [1 ]
Heath, Emily [1 ]
Stroian, Gabriela [1 ]
Collins, D. Louis [1 ]
Seuntjens, Jan [1 ]
机构
[1] McGill Univ, Ctr Hlth, Dept Med Phys, Montreal, PQ H3G 1A4, Canada
基金
加拿大健康研究院;
关键词
4D radiotherapy; deformable lung phantom; deformable image registration;
D O I
10.1118/1.2836417
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Motion of thoracic tumors with respiration presents a challenge for three-dimensional (3D) conformal radiation therapy treatment. Validation of techniques aimed at measuring and minimizing the effects of respiratory motion requires a realistic deformable phantom for use as a gold standard. The purpose of this study was to develop and study the characteristics of a reproducible, tissue equivalent, deformable lung phantom. The phantom consists of a Lucite cylinder filled with water containing a latex balloon stuffed with dampened natural sponges. The balloon is attached to a piston that mimics the human diaphragm. Nylon wires and Lucite beads, emulating vascular and bronchial bifurcations, were uniformly glued at various locations throughout the sponges. The phantom is capable of simulating programmed irregular breathing patterns with varying periods and amplitudes. A tissue equivalent tumor, suitable for holding radiochromic film for dose measurements was embedded in the sponge. To assess phantom motion, eight 3D computed tomography data sets of the static phantom were acquired for eight equally spaced positions of the piston. The 3D trajectories of 12 manually chosen point landmarks and the tumor center-of-mass were studied. Motion reproducibility tests of the deformed phantom were established on seven repeat scans of three different states of compression. Deformable image registration (DIR) of the extreme breathing phases was performed. The accuracy of the DIR was evaluated by visual inspection of image overlays and quantified by the distance-to-agreement (DTA) of manually chosen point landmarks and triangulated surfaces obtained from 3D contoured structures. In initial tests of the phantom, a 20-mm excursion of the piston resulted in deformations of the balloon of 20 mm superior-inferior, 4 mm anterior-posterior, and 5 mm left-right. The change in the phantom mean lung density ranged from 0.24 (0.12 SD) g/cm(3) at peak exhale to 0.19 (0.12 SD) g/cm(3) at peak inhale. The SI displacement of the landmarks varied between 94% and 3% of the piston excursion for positions closer and farther away from the piston, respectively. The reproducibility of the phantom deformation was within the image resolution (0.7 x 0.7 x 1.25 mm(3)). Vector average registration accuracy based on point landmarks was found to be 0.5 (0.4 SD) mm. The tumor and lung mean 3D DTA obtained from triangulated surfaces were 0.4 (0.1 SD) mm and 1.0 (0.8 SD) mm, respectively. This phantom is capable of reproducibly emulating the physically realistic lung features and deformations and has a wide range of potential applications, including four-dimensional (4D) imaging, evaluation of deformable registration accuracy, 4D planning and dose delivery. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:1094 / 1102
页数:9
相关论文
共 34 条
  • [1] A novel four-dimensional radiotherapy method for lung cancer: imaging, treatment planning and delivery
    Alasti, H.
    Cho, Y. B.
    Vandermeer, A. D.
    Abbas, A.
    Norrlinger, B.
    Shubbar, S.
    Bezjak, A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (12) : 3251 - 3267
  • [2] [Anonymous], 2004, Cancer Facts and Figures
  • [3] Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing
    Balter, JM
    TenHaken, RK
    Lawrence, TS
    Lam, KL
    Robertson, JM
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (01): : 167 - 174
  • [4] Intensity-modulated radiotherapy: Current status and issues of interest
    Boyer, AL
    Butler, EB
    DiPetrillo, TA
    Engler, MJ
    Fraass, B
    Grant, W
    Ling, CC
    Low, DA
    Mackie, TR
    Mohan, R
    Purdy, JA
    Roach, M
    Rosenman, JG
    Verhey, LJ
    Wong, JW
    Cumberlin, RL
    Stone, H
    Palta, JR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04): : 880 - 914
  • [5] Accuracy of finite element model-based multi-organ deformable image registration
    Brock, KK
    Sharpe, MB
    Dawson, LA
    Kim, SM
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1647 - 1659
  • [6] Animal: Validation and applications of nonlinear registration-based segmentation
    Collins, DL
    Evans, AC
    [J]. INTERNATIONAL JOURNAL OF PATTERN RECOGNITION AND ARTIFICIAL INTELLIGENCE, 1997, 11 (08) : 1271 - 1294
  • [7] Gated CT imaging using a free-breathing respiration signal from flow-volume spirometry
    D'Souza, WD
    Kwok, Y
    Deyoung, C
    Zacharapoulos, N
    Pepelea, M
    Klahr, P
    Yu, CX
    [J]. MEDICAL PHYSICS, 2005, 32 (12) : 3641 - 3649
  • [8] Respiration-correlated spiral CT: A method of measuring respiratory-induced anatomic motion for radiation treatment planning
    Ford, EC
    Mageras, GS
    Yorke, E
    Ling, CC
    [J]. MEDICAL PHYSICS, 2003, 30 (01) : 88 - 97
  • [9] Intrathoracic tumour motion estimation from CT imaging using the 3D optical flow method
    Guerrero, T
    Zhang, G
    Huang, TC
    Lin, KP
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17) : 4147 - 4161
  • [10] Quantification of accuracy of the automated nonlinear image matching and anatomical labeling (ANIMAL) nonlinear registration algorithm for 4D CT images of lung
    Heath, E.
    Collins, D. L.
    Keall, P. J.
    Dong, L.
    Seuntjens, J.
    [J]. MEDICAL PHYSICS, 2007, 34 (11) : 4409 - 4421