Comparison on the properties of bovine pericardium and porcine pericardium used as leaflet materials of transcatheter heart valve

被引:8
作者
Jin, Chang [1 ,2 ]
Zhao, Li [1 ,2 ]
Wu, Zebin [3 ,4 ]
Li, Bin [1 ,2 ]
Liu, Ronghui [1 ,2 ]
He, Hongping [1 ,2 ]
Wang, Lizhen [3 ,4 ]
Wang, Weidong [1 ,2 ]
机构
[1] Chinese Peoples Liberat Army Gen Hosp, Res Ctr Biomed Engn, Med Innovat & Res Div, South Ward Bldg,Room 105,28 Fuxing Rd, Beijing 100853, Peoples R China
[2] Chinese Peoples Liberat Army Gen Hosp, Key Lab Biomed Engn & Translat Med, Minist Ind & Informat Thchnol, Beijing, Peoples R China
[3] Beihang Univ, Beijing Adv Innovat Ctr Biomed Engn, Sch Biol Sci & Med Engn, Beijing, Peoples R China
[4] Beihang Univ, Key Lab Biomech & Mechanobiol, Minist Educ, Beijing, Peoples R China
关键词
bovine pericardium; crimp; dehydration; porcine pericardium; transcatheter heart valve; AORTIC-STENOSIS; IMPLANTATION; CALCIFICATION; REPLACEMENT; MANAGEMENT;
D O I
10.1111/aor.14074
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background In order to obtain the smaller delivery diameter, porcine pericardium had been used as a substitute material of bovine pericardium for the leaflet materials of transcatheter heart valve (THV). However, the differences between them had not been fully studied. Therefore, this study compared the microstructure, biochemical and mechanical properties of two materials and hydrodynamics of THV made by the two materials in detail. Methods In this study, firstly, the microstructure of pericardium was analyzed by staining and scanning electron microscope; secondly, the biochemical properties of pericardium after different processes were compared by heat shrinkage temperature test, free amino and carboxyl concentration test, enzyme degradation test, subcutaneous implantation calcification analysis in rats; finally, the mechanical properties were evaluated by uniaxial tensile test before and after the pericardium being crimped, and then, the hydrodynamics of THV was studied according to the ISO5840 standard. Results Compared with bovine pericardium, after the same process, porcine pericardium showed a looser and tinier fiber bundle, a similar free carboxyl concentration, a lower resistance to enzyme degradation, a significantly lower calcification, bearing capacity and damage after being crimped, a better hydrodynamic and adaption with lower cardiac output and deformation of implantation position. Meanwhile the dehydration process of pericardium almost had preserved all the biochemical advantages of two materials. Conclusion In this study, porcine and bovine pericardium showed some significant differences in biochemical, mechanical properties and hydrodynamics. According to the results, it was presumed that the thinner porcine pericardium might be more suitable for THV of right heart system. Meanwhile, more attention should be taken for the calcification of THV made by the bovine pericardium.
引用
收藏
页码:427 / 438
页数:12
相关论文
共 55 条
  • [1] Quantification and Analysis of Leaflet Flutter on Biological Prosthetic Cardiac Valves
    Avelar, Artur H. de F.
    Canestri, Jean A.
    Bim, Camila
    Silva, Maira G. M.
    Huebner, Rudolf
    Pinotti, Marcos
    [J]. ARTIFICIAL ORGANS, 2017, 41 (09) : 835 - 844
  • [2] Native Bovine and Porcine Pericardia Respond to Load With Additive Recruitment of Collagen Fibers
    Bagno, Andrea
    Aguiari, Paola
    Fiorese, Michele
    Iop, Laura
    Spina, Michele
    Gerosa, Gino
    [J]. ARTIFICIAL ORGANS, 2018, 42 (05) : 540 - 548
  • [3] Bailey MT., 2010, J BIOMED MATER RES A, V66, P93
  • [4] Evaluation of transcatheter heart valve biomaterials: Biomechanical characterization of bovine and porcine pericardium
    Caballeros, Andres
    Sulejmanil, Fatiesa
    Martin, Caitlin
    Thuy Pham
    Sun, Wei
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 75 : 486 - 494
  • [5] Biochemical properties of heat-treated valvular bioprostheses
    Carpentier, SM
    Shen, M
    Chen, L
    Cunanan, CM
    Martinet, B
    Carpentier, A
    [J]. ANNALS OF THORACIC SURGERY, 2001, 71 (05) : S410 - S412
  • [6] Heat treatment mitigates calcification of valvular bioprostheses
    Carpentier, SM
    Chen, L
    Shen, M
    Fornes, P
    Martinet, B
    Quintero, LJ
    Witzel, TH
    Carpentier, AF
    [J]. ANNALS OF THORACIC SURGERY, 1998, 66 (06) : S264 - S266
  • [7] Chang J., 2019, J BIOMED ENG, V36, P129
  • [8] Chang J., 2018, INT J COMP METH-SING, V16
  • [9] Chen Z., 1984, ACTA U MED 2 SHANGHA, V5, P385
  • [10] Traumalic leaflet injury: comparison of porcine leaflet self-expandable and bovine leaflet balloon-expandable prostheses
    Convelbo, Channing
    El Hafci, Hanane
    Petite, Herve
    Zegdi, Rachid
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2018, 53 (05) : 1062 - 1067