Proton-transfer reaction mass spectrometry (PTRMS) in combination with thermal desorption (TD) for sensitive off-line analysis of volatiles

被引:11
作者
Crespo, Elena [1 ]
Devasena, Samudrala [1 ]
Sikkens, Cor [1 ]
Centeno, Raymund [1 ]
Cristescu, Simona M. [1 ]
Harren, Frans J. M. [1 ]
机构
[1] Radboud Univ Nijmegen, Life Sci Trace Gas Facil, Inst Mol & Mat, NL-6525 AJ Nijmegen, Netherlands
基金
比尔及梅琳达.盖茨基金会;
关键词
TRACE GAS-DETECTION; ORGANIC-COMPOUNDS; TENAX TA; BREATH; TUBE; TERPENES; DECOMPOSITION; PERFORMANCE; ADSORPTION; EMISSIONS;
D O I
10.1002/rcm.6191
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
RATIONALE: When performing trace gas analysis, it is not always possible to bring the source of volatiles and the gas analyzer together. In these cases, volatile storage containers, such as thermal desorption (TD) tubes, can be used for off-line measurement. TD is routinely combined with gas chromatography/mass spectrometry (GC/MS), but so far not with proton-transfer reaction mass spectrometry (PTRMS), which has a faster response. METHODS: A PTR-quadrupole-MS instrument and a PTR-ion-trap-MS instrument were separately coupled to a TD unit for off-line analysis of trace volatiles in air. Carbograph 1TD/Carbopack X sorbent tubes were filled with different concentrations of a trace gas mixture containing low molecular weight volatiles (32 g/mol up to 136 g/mol) and measured with the above-mentioned combinations. The carrier gas in the TD unit was changed from helium to nitrogen to be able to combine this instrument with the mass spectrometer. RESULTS: Good linearity and reproducibility with the amount of gas stored were obtained. The storage capacity over time (up to 14 days) showed larger variability (<11% for all compounds, except for acetone 27%). Several tubes were filled with breath of different persons, and the breath of a smoker showed increased levels of acetonitrile and benzene. The combination of the PTR ion-trap instrument with the TD unit was also investigated. Due to its higher sampling rate, the ion-trap system showed higher throughput capabilities than the quadrupole system. CONCLUSIONS: The combination of TD with PTRMS using both a quadrupole and an ion trap for off-line volatile analysis has been validated. TD tubes can be a robust and compact volatile storage method when the mass spectrometry and the sampling cannot be performed in the same place, for example in large screening studies. In addition, a higher measurement throughput than with GC/MS could be obtained. Copyright (C) 2012 John Wiley & Sons, Ltd.
引用
收藏
页码:990 / 996
页数:7
相关论文
共 42 条
[1]   Geographical differentiation of saffron by GC-MS/FID and chemometrics [J].
Anastasaki, E. ;
Kanakis, C. ;
Pappas, C. ;
Maggi, L. ;
del Campo, C. P. ;
Carmona, M. ;
Alonso, G. L. ;
Polissiou, M. G. .
EUROPEAN FOOD RESEARCH AND TECHNOLOGY, 2009, 229 (06) :899-905
[2]   Analysis and stability of aldehydes and terpenes in electropolished canisters [J].
Batterman, SA ;
Zhang, GZ ;
Baumann, M .
ATMOSPHERIC ENVIRONMENT, 1998, 32 (10) :1647-1655
[3]   Trace gas detection from fermentation processes in apples; an intercomparison study between proton-transfer-reaction mass spectrometry and laser photoacoustics [J].
Boamfa, EI ;
Steeghs, MML ;
Cristescu, SM ;
Harren, FJM .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2004, 239 (2-3) :193-201
[4]   Fruit ripeness monitoring using an Electronic Nose [J].
Brezmes, J ;
Llobet, E ;
Vilanova, X ;
Saiz, G ;
Correig, X .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 69 (03) :223-229
[5]   Human exhaled air analytics: Biomarkers of diseases [J].
Buszewski, Boguslaw ;
Kesy, Martyna ;
Ligor, Tomasz ;
Amann, Anton .
BIOMEDICAL CHROMATOGRAPHY, 2007, 21 (06) :553-566
[6]   Decomposition of terpenes by ozone during sampling on Tenax [J].
Calogirou, A ;
Larsen, BR ;
Brussol, C ;
Duane, M ;
Kotzias, D .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1499-1506
[7]   BUILDUP OF ARTIFACTS ON ADSORBENTS DURING STORAGE AND ITS EFFECT ON PASSIVE SAMPLING AND GAS CHROMATOGRAPHY-FLAME IONIZATION DETECTION OF LOW CONCENTRATIONS OF VOLATILE ORGANIC-COMPOUNDS IN AIR [J].
CAO, XL ;
HEWITT, CN .
JOURNAL OF CHROMATOGRAPHY A, 1994, 688 (1-2) :368-374
[8]   Decomposition of alpha-pinene and sabinene on solid sorbents, Tenax TA and Carboxen [J].
Coeur, C ;
Jacob, V ;
Denis, I ;
Foster, P .
JOURNAL OF CHROMATOGRAPHY A, 1997, 786 (01) :185-187
[9]   Proton Transfer Reaction Mass Spectrometry detects rapid changes in volatile metabolite emission by Mycobacterium smegmatis after the addition of specific antimicrobial agents [J].
Crespo, E. ;
Cristescu, S. M. ;
de Ronde, H. ;
Kuijper, S. ;
Kolk, A. H. J. ;
Anthony, R. M. ;
Harren, F. J. M. .
JOURNAL OF MICROBIOLOGICAL METHODS, 2011, 86 (01) :8-15
[10]   Classification of Tempranillo wines according to geographic origin: Combination of mass spectrometry based electronic nose and chemometrics [J].
Cynkar, Wies ;
Dambergs, Robert ;
Smith, Paul ;
Cozzolino, Daniel .
ANALYTICA CHIMICA ACTA, 2010, 660 (1-2) :227-231