Right Ventricular Septal Pacing Using a Thin Lumenless Pacing Lead and Delivery System with a Deflectable Catheter Long-Term Outcomes

被引:5
作者
Suzuki, Takeya [1 ]
Fujino, Tadashi [1 ]
Shinohara, Masaya [1 ]
Koike, Hideki [1 ]
Kinoshita, Toshio [1 ]
Yuzawa, Hitomi [1 ]
Ikeda, Takanori [1 ]
机构
[1] Toho Univ, Fac Med, Dept Cardiovasc Med, Tokyo, Japan
关键词
Pacemaker; SelectSecure; GRADE ATRIOVENTRICULAR-BLOCK; HEART-FAILURE; DUAL-CHAMBER; CONGENITAL HEART; SITE; METAANALYSIS; PERFORMANCE; RISK; THERAPY; ATRIAL;
D O I
10.1536/ihj.17-526
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The SelectSecure (TM) lead system (SSLS), which is composed of a thin lumenless, active-fixation lead and a deflectable catheter, is approved for use in Japan. This study aimed to evaluate the long-term clinical outcomes of right ventricular (RV) septal pacing with the SSLS along with the system's safety and electrical performance. A total of 129 patients were divided into the following 3 groups: the RV septal pacing with the SSLS group (SSP, n = 21); the RV septal pacing with the conventional lead group (Septal, n = 77); and the RV apical pacing with the conventional lead group (Apical. n = 31). All lead-related complications and pacing parameters during follow-up were compared among the groups. The clinical outcome was heart failure-associated hospitalization. The SSP and Septal groups showed significantly shorter paced QRS duration than the Apical group. During the follow-up for a mean of 49.5 +/- 13.1 months, no lead-related complications occurred in any of the groups. A case of pericardial effusion occurred in the SSP group, but cardiac tamponade did not occur, and it spontaneously resolved. The ventricular pacing threshold after the follow-up period was higher in the SSP group than in the other 2 groups. There was no difference in the primary heart failure hospitalization among the 3 groups. The SSLS could be effective in producing a narrow QRS width with RV septal pacing, but its pacing threshold was higher than conventional leads in the chronic phase.
引用
收藏
页码:1253 / 1260
页数:8
相关论文
共 32 条
[21]  
Shepherd E, 2015, HEART RHYTHM, V12, P1227, DOI 10.1016/j.hrthm.2015.03.004
[22]   Beneficial effects of right ventricular non-apical vs. apical pacing: a systematic review and meta-analysis of randomized-controlled trials [J].
Shimony, Avi ;
Eisenberg, Mark J. ;
Filion, Kristian B. ;
Amit, Guy .
EUROPACE, 2012, 14 (01) :81-91
[23]   Regional alterations in protein expression in the dyssynchronous failing heart [J].
Spragg, DD ;
Leclercq, C ;
Loghmani, M ;
Faris, OP ;
Tunin, RS ;
DiSilvestre, D ;
McVeigh, ER ;
Tomaselli, GF ;
Kass, DA .
CIRCULATION, 2003, 108 (08) :929-932
[24]   Heart failure during cardiac pacing [J].
Sweeney, MO ;
Hellkamp, AS .
CIRCULATION, 2006, 113 (17) :2082-2088
[25]   Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction [J].
Sweeney, MO ;
Hellkamp, AS ;
Ellenbogen, KA ;
Greenspon, AJ ;
Freedman, RA ;
Lee, KL ;
Lamas, GA .
CIRCULATION, 2003, 107 (23) :2932-2937
[26]   A randomized comparison of permanent septal versus apical right ventricular pacing: Short-term results [J].
Victor, F ;
Mabo, P ;
Mansour, H ;
Pavin, D ;
Kabalu, G ;
De Place, C ;
Leclercq, C ;
Daubert, JC .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (03) :238-242
[27]   Effects of Right Ventricular Nonapical Pacing on Cardiac Function: A Meta-analysis of Randomized Controlled Trials [J].
Wang Weizong ;
Wang Zhongsu ;
Zhang Yujiao ;
Gao Mei ;
Wang Jiangrong ;
Zhang Yong ;
Xie Xinxing ;
Hou Yinglong .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2013, 36 (08) :1032-1051
[28]  
Wilkoff BL, 2002, JAMA-J AM MED ASSOC, V288, P3115
[29]   Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications, and Patient Management [J].
Wilkoff, Bruce L. ;
Love, Charles J. ;
Byrd, Charles L. ;
Bongiorni, Maria Grazia ;
Carrillo, Roger G. ;
Crossley, George H., III ;
Epstein, Laurence M. ;
Friedman, Richard A. ;
Kennergren, Charles E. H. ;
Mitkowski, Przemyslaw ;
Schaerf, Raymond H. M. ;
Wazni, Oussama M. .
HEART RHYTHM, 2009, 6 (07) :1085-1104
[30]   Selective Site Pacing From the Right Ventricular Mid-Septum Follow-up of Lead Performance and Procedure Technique [J].
Yusu, Satoru ;
Mera, Hisaaki ;
Hoshida, Kyoko ;
Miyakoshi, Mutsumi ;
Miwa, Yosuke ;
Tsukada, Takehiro ;
Yoshino, Hideaki ;
Ikeda, Takanori .
INTERNATIONAL HEART JOURNAL, 2012, 53 (02) :113-116