Beam modeling and commissioning for Monte Carlo photon beam on an Elekta Versa HD LINAC

被引:2
|
作者
Can, Sumeyra [1 ]
Karacetin, Didem [2 ]
Meric, Niyazi [1 ]
机构
[1] Ankara Univ, Inst Nucl Sci, TR-06100 Ankara, Turkey
[2] Istanbul Basaksehir Cam & Sakura City Hosp, Istanbul, Turkey
关键词
Beam data measurements; Beam matching; Elekta VersaHD LINAC; Monte Carlo simulation; Monaco Treatment planning system; Virtual source model; TREATMENT PLANNING SYSTEM; DOSE CALCULATION TOOL; VIRTUAL SOURCE MODEL; CLINICAL IMPLEMENTATION; VALIDATION; IMRT; THERAPY;
D O I
10.1016/j.apradiso.2021.110054
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: This study aims at analyzing beam data commissioning along with verifying Monte Carlo-based treatment planning system on the basis of the manufacturer guidelines for Elekta Versa HD Linear Accelerator. Moreover, evaluating the beam match process in terms of quality index, photon profile and multi leaf collimator (MLC) offset is aimed as well. Materials and methods: The process of collecting beam data for Monaco 5.51 Treatment Planning System commissioning was done based on the instructions provided by the manufacturer as well as AAPM TG-106. Monte Carlo analysis was done for output factors in water, percent depth dose and beam profiles. A set of eight static and intensity modulated radiation therapy fields were used to verify the MLC parameters. Results: The difference between the measured and modeled penetrative quality (D-10) was achieved to be 0.54%. The output factors for 6 MV photon energy were measured and the difference between the measured and Monte Carlo output results was smaller than 1% for all the fields. The average percentage of passing the gamma criteria for commissioning test fields was (95+-4)%, however, the minimum agreement was 87.5% belonging to "7SEGA". Additionally, the agreement between both LINAC is 96%, however, the second LINAC reveals a positive offset in the point of approximately -4 cm on the x-axis at the crossplane. Conclusion: Test commissioning was successfully verified using a homogeneous phantom for point dose measurements, post modelling MLC parameters and patient QA plans. All plan parameters pass the gamma criteria. 6 MV photon beam was successfully commissioned for Elekta VersaHD LINAC and is ready for clinical implementation.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Monte Carlo framework for commissioning a synchrotron-based discrete spot scanning proton beam system and treatment plan verification
    Moskvin, Vadim P.
    Faught, Austin
    Pirlepesov, Fakhriddin
    Zhao, Li
    Hua, Chia-Ho
    Merchant, Thomas E.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (04)
  • [32] First Monte Carlo beam model for ultra-high dose rate radiotherapy with a compact electron LINAC
    Dai, Tianyuan
    Sloop, Austin M.
    Rahman, Mahbubur R.
    Sunnerberg, Jacob P.
    Clark, Megan A.
    Young, Ralph
    Adamczyk, Sebastian
    Von Voigts-Rhetz, Philip
    Patane, Chris
    Turk, Michael
    Jarvis, Lesley
    Pogue, Brian W.
    Gladstone, David J.
    Bruza, Petr
    Zhang, Rongxiao
    MEDICAL PHYSICS, 2024, 51 (07) : 5109 - 5118
  • [33] Monte Carlo calculated beam quality correction factors for high energy photon and electron fields
    Czarnecki, Damian
    Zink, Klemens
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2025, 131
  • [34] Validation of PRIMO Monte Carlo Model of Clinac®iX 6MV Photon Beam
    Sarin, B.
    Bindhu, B.
    Saju, B.
    Nair, Raguram K.
    JOURNAL OF MEDICAL PHYSICS, 2020, 45 (01) : 24 - 35
  • [35] Inference of the optimal pretarget electron beam parameters in a Monte Carlo virtual linac model through simulated annealing
    Bush, Karl
    Zavgorodni, Sergei
    Beckham, Wayne
    MEDICAL PHYSICS, 2009, 36 (06) : 2309 - 2319
  • [36] Benchmarking of Monte Carlo model of Siemens Oncor® linear accelerator for 18 MV photon beam: Determination of initial electron beam parameters
    Najafzadeh, Milad
    Hoseini-Ghafarokhi, Mojtaba
    Bolagh, Rezgar Shahi Mayn
    Haghparast, Mohammad
    Zarifi, Shiva
    Nickfarjam, Abolfazl
    Farhood, Bagher
    Chow, James C. L.
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2019, 27 (06) : 1047 - 1070
  • [37] Using PRIMO to determine the initial beam parameters of Elekta Synergy linac for electron beam energies of 6, 9, 12, and 15 MeV
    Rezzoug, Mohammed
    Zerfaoui, Mustapha
    Oulhouq, Yassine
    Rrhioua, Abdeslem
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2023, 28 (05) : 592 - 600
  • [38] Modeling the process of electron-beam-induced deposition by dynamic Monte Carlo simulation
    Liu, ZQ
    Mitsuishi, K
    Furuya, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2005, 44 (7B): : 5659 - 5663
  • [39] Monte Carlo modeling for dose assessment in cone beam CT for oral and maxillofacial applications
    Zhang, G.
    Marshall, N.
    Bogaerts, R.
    Jacobs, R.
    Bosmans, H.
    MEDICAL PHYSICS, 2013, 40 (07)
  • [40] Modeling the biophysical effects in a carbon beam delivery line by using Monte Carlo simulations
    Cho, Ilsung
    Yoo, SeungHoon
    Cho, Sungho
    Kim, Eun Ho
    Song, Yongkeun
    Shin, Jae-ik
    Jung, Won-Gyun
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 69 (05) : 868 - 874