Beam modeling and commissioning for Monte Carlo photon beam on an Elekta Versa HD LINAC

被引:2
|
作者
Can, Sumeyra [1 ]
Karacetin, Didem [2 ]
Meric, Niyazi [1 ]
机构
[1] Ankara Univ, Inst Nucl Sci, TR-06100 Ankara, Turkey
[2] Istanbul Basaksehir Cam & Sakura City Hosp, Istanbul, Turkey
关键词
Beam data measurements; Beam matching; Elekta VersaHD LINAC; Monte Carlo simulation; Monaco Treatment planning system; Virtual source model; TREATMENT PLANNING SYSTEM; DOSE CALCULATION TOOL; VIRTUAL SOURCE MODEL; CLINICAL IMPLEMENTATION; VALIDATION; IMRT; THERAPY;
D O I
10.1016/j.apradiso.2021.110054
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: This study aims at analyzing beam data commissioning along with verifying Monte Carlo-based treatment planning system on the basis of the manufacturer guidelines for Elekta Versa HD Linear Accelerator. Moreover, evaluating the beam match process in terms of quality index, photon profile and multi leaf collimator (MLC) offset is aimed as well. Materials and methods: The process of collecting beam data for Monaco 5.51 Treatment Planning System commissioning was done based on the instructions provided by the manufacturer as well as AAPM TG-106. Monte Carlo analysis was done for output factors in water, percent depth dose and beam profiles. A set of eight static and intensity modulated radiation therapy fields were used to verify the MLC parameters. Results: The difference between the measured and modeled penetrative quality (D-10) was achieved to be 0.54%. The output factors for 6 MV photon energy were measured and the difference between the measured and Monte Carlo output results was smaller than 1% for all the fields. The average percentage of passing the gamma criteria for commissioning test fields was (95+-4)%, however, the minimum agreement was 87.5% belonging to "7SEGA". Additionally, the agreement between both LINAC is 96%, however, the second LINAC reveals a positive offset in the point of approximately -4 cm on the x-axis at the crossplane. Conclusion: Test commissioning was successfully verified using a homogeneous phantom for point dose measurements, post modelling MLC parameters and patient QA plans. All plan parameters pass the gamma criteria. 6 MV photon beam was successfully commissioned for Elekta VersaHD LINAC and is ready for clinical implementation.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Developing a new Monte Carlo algorithm as an alternative tool to simulate virtual source model on an Elekta Versa HD Linac
    Can, Sumeyra
    Sahiner, Eren
    Karacetin, Didem
    Meric, Niyazi
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2022, 15 (03) : 299 - 308
  • [2] Electron beam modeling and commissioning for Monte Carlo treatment planning
    Jiang, SB
    Kapur, A
    Ma, CM
    MEDICAL PHYSICS, 2000, 27 (01) : 180 - 191
  • [3] Auto-commissioning of a Monte Carlo electron beam model with application to photon MLC shaped electron fields
    Fix, M. K.
    Frei, D.
    Mueller, S.
    Guyer, G.
    Loebner, H. A.
    Volken, W.
    Manser, P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (04)
  • [4] Monte carlo model and output factors of elekta Infinity™ 6 and 10 MV photon beam
    Yani, Sitti
    Budiansah, Indra
    Rhani, Mohamad Fahdillah
    Haryanto, Freddy
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (04) : 470 - 478
  • [5] Monte Carlo Simulation-Based BEAMnrc Code of a 6 MV Photon Beam Produced by a Linear Accelerator (LINAC)
    Sapundani, R.
    Ekawati, R.
    Wibowo, K. M.
    ATOM INDONESIA, 2021, 47 (03) : 199 - 204
  • [6] Simulation of a 6 MV Elekta Precise Linac photon beam using GATE/GEANT4
    Grevillot, L.
    Frisson, T.
    Maneval, D.
    Zahra, N.
    Badel, J-N
    Sarrut, D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (04) : 903 - 918
  • [7] Simple beam models for Monte Carlo photon beam dose calculations in radiotherapy
    Fix, MK
    Keller, H
    Rüegsegger, P
    Born, EJ
    MEDICAL PHYSICS, 2000, 27 (12) : 2739 - 2747
  • [8] Monte Carlo based beam model using a photon MLC for modulated electron radiotherapy
    Henzen, D.
    Manser, P.
    Frei, D.
    Volken, W.
    Neuenschwander, H.
    Born, E. J.
    Vetterli, D.
    Chatelain, C.
    Stampanoni, M. F. M.
    Fix, M. K.
    MEDICAL PHYSICS, 2014, 41 (02)
  • [9] Optimisation of variance reduction techniques in EGSnrc Monte Carlo for a 6 MV photon beam of an Elekta Synergy linear accelerator
    Almatani, Turki
    JOURNAL OF KING SAUD UNIVERSITY SCIENCE, 2021, 33 (04)
  • [10] Representation and illustration of the initial parameters in GATE 8.1 monte carlo simulation of an Elekta Versa-HD linear accelerator
    Abolaban, Fouad A.
    Taha, Eslam M.
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2020, 13 (01) : 642 - 647