Real-time arrhythmia detection with supplementary ECG quality and pulse wave monitoring for the reduction of false alarms in ICUs

被引:25
作者
Krasteva, Vessela [1 ]
Jekova, Irena [1 ]
Leber, Remo [2 ]
Schmid, Ramun [2 ]
Abacherli, Roger [2 ,3 ]
机构
[1] Bulgarian Acad Sci, Inst Biophys & Biomed Engn, Acad G Bonchev Str B1 105, BU-1113 Sofia, Bulgaria
[2] Schiller AG, Res & Signal Proc, Altgasse 68, CH-6341 Baar, Switzerland
[3] Univ Basel Hosp, CRIB, Spitalstr 2, CH-4056 Basel, Switzerland
关键词
ECG monitoring; QRS detection; heartbeat classification; ECG quality; pulse wave analysis; life-threatening arrhythmia; false ICU alarms; ARTERIAL-BLOOD PRESSURE; CARE; SYSTEM; NOISE; STATEMENT; STRESS;
D O I
10.1088/0967-3334/37/8/1273
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
False intensive care unit (ICU) alarms induce stress in both patients and clinical staff and decrease the quality of care, thus significantly increasing both the hospital recovery time and rehospitalization rates. In the PhysioNet/CinC Challenge 2015 for reducing false arrhythmia alarms in ICU bedside monitor data, this paper validates the application of a real-time arrhythmia detection library (ADLib, Schiller AG) for the robust detection of five types of life-threatening arrhythmia alarms. The strength of the application is to give immediate feedback on the arrhythmia event within a scan interval of 3 s-7.5 s, and to increase the noise immunity of electrocardiogram (ECG) arrhythmia analysis by fusing its decision with supplementary ECG quality interpretation and real-time pulse wave monitoring (quality and hemodynamics) using arterial blood pressure or photoplethysmographic signals. We achieved the third-ranked real- time score (79.41) in the challenge (Event 1), however, the rank was not officially recognized due to the 'closed-source' entry. This study shows the optimization of the alarm decision module, using tunable parameters such as the scan interval, lead quality threshold, and pulse wave features, with a follow-up improvement of the real- time score (80.07). The performance (true positive rate, true negative rate) is reported in the blinded challenge test set for different arrhythmias: asystole (83%, 96%), extreme bradycardia (100%, 90%), extreme tachycardia (98%, 80%), ventricular tachycardia (84%, 82%), and ventricular fibrillation (78%, 84%). Another part of this study considers the validation of ADLib with four reference ECG databases (AHA, EDB, SVDB, MIT-BIH) according to the international recommendations for performance reports in ECG monitors (ANSI/AAMI EC57). The sensitivity (Se) and positive predictivity (+P) are: QRS detector QRS (Se, +P) > 99.7%, ventricular ectopic beat (VEB) classifier VEB (Se, +P) = 95%, and ventricular fibrillation detector VFIB (P + = 94.8%) > VFIB (Se = 86.4%), adjusted to the clinical setting requirements, giving preference to low false positive alarms.
引用
收藏
页码:1273 / 1297
页数:25
相关论文
共 48 条
[1]   Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveform. [J].
Aboukhalil, Anton ;
Nielsen, Larry ;
Saeed, Mohammed ;
Mark, Roger G. ;
Clifford, Gari D. .
JOURNAL OF BIOMEDICAL INFORMATICS, 2008, 41 (03) :442-451
[2]  
AHA, 1985, AM HEART ASS VENTR A
[3]   Detecting ventricular fibrillation by time-delay methods [J].
Amann, Anton ;
Tratnig, Robert ;
Unterkofler, Karl .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (01) :174-177
[4]  
[Anonymous], 2011, 60601227 IEC
[5]  
[Anonymous], 2007, EC132002 ANSIAAMI
[6]  
[Anonymous], 2003, EC571998 ANSIAAMI
[7]  
Ansari S, 2015, COMPUT CARDIOL CONF, V42, P1181, DOI 10.1109/CIC.2015.7411127
[8]  
Antink CH, 2015, COMPUT CARDIOL CONF, V42, P285, DOI 10.1109/CIC.2015.7408642
[9]  
BAKER C F, 1992, Critical Care Nursing Quarterly, V15, P75
[10]   ECG Signal Quality During Arrhythmia and Its Application to False Alarm Reduction [J].
Behar, Joachim ;
Oster, Julien ;
Li, Qiao ;
Clifford, Gari D. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (06) :1660-1666