Experience with the Vibrant Soundbridge RW-Coupler for round window Vibroplasty with tympanosclerosis

被引:21
作者
Iwasaki, Satoshi [2 ]
Suzuki, Hiroaki [1 ]
Moteki, Hideaki [1 ,2 ]
Miyagawa, Maiko [1 ]
Takumi, Yutaka [1 ,2 ]
Usami, Shin-ichi [1 ]
机构
[1] Shinshu Univ, Sch Med, Dept Otolaryngol, Matsumoto, Nagano 3908621, Japan
[2] Shinshu Univ, Sch Med, Dept Hearing Implant Sci, Matsumoto, Nagano 3908621, Japan
关键词
Floating mass transducer; sclerosis; round window membrane; perilymphatic fistula; FLOATING MASS TRANSDUCER;
D O I
10.3109/00016489.2011.649492
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Usage of the Vibrant Soundbridge (VSB) with round window (RW)-Coupler placement at the RW has been shown to successfully treat mixed hearing loss. Coupling between the VSB's floating mass transducer (FMT) and the RW membrane is difficult in the case of sclerosis in the RW and drilling down the bony lip until the RW membrane can be seen completely can possibly induce a perilymphatic fistula. A 68-year-old woman who had bilateral mixed hearing loss with sclerosis in the RW due to tympanosclerosis underwent a RW-Vibroplasty with a RW-Coupler. Speech discrimination scores in quiet and noise and functional gain with the VSB with RW-Coupler were better than those using a conventional hearing aid. The results of the present case have shown the feasibility of implanting a VSB with RW-Coupler in patients with mixed hearing loss due to tympanosclerosis.
引用
收藏
页码:676 / 682
页数:7
相关论文
共 14 条
[1]   The floating mass transducer at the round window: Direct transmission or bone conduction? [J].
Arnold, Andreas ;
Kompis, Martin ;
Candreia, Claudia ;
Pfiffner, Flurin ;
Haeusler, Rudolf ;
Stieger, Christof .
HEARING RESEARCH, 2010, 263 (1-2) :120-127
[2]   Tympanosclerosis and our surgical results [J].
Aslan, Hale ;
Katilmis, Hueseyin ;
Oeztuerkcan, Sedat ;
Ilknur, Ali Ekber ;
Basoglu, Sinan .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2010, 267 (05) :673-677
[3]   Coupling the Vibrant Soundbridge to Cochlea Round Window: Auditory Results in Patients With Mixed Hearing Loss [J].
Beltrame, Achille M. ;
Martini, Alessandro ;
Prosser, Silvano ;
Giarbini, Nadia ;
Streitberger, Christian .
OTOLOGY & NEUROTOLOGY, 2009, 30 (02) :194-201
[4]   Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window [J].
Colletti, Vittorio ;
Soli, Sigfrid D. ;
Carner, Marco ;
Colletti, L. .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2006, 45 (10) :600-608
[5]   Phase III clinical trial results with the Vibrant Soundbridge implantable middle ear hearing device: A prospective controlled multicenter study [J].
Luetje, CM ;
Brackman, D ;
Balkany, TJ ;
Maw, J ;
Baker, RS ;
Kelsall, D ;
Backous, D ;
Miyamoto, R ;
Parisier, S ;
Arts, A .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2002, 126 (02) :97-107
[6]   Round Window Otosclerosis: Radiologic Classification and Clinical Correlations [J].
Mansour, Salah ;
Nicolas, Karen ;
Ahmad, Hassan Haidar .
OTOLOGY & NEUROTOLOGY, 2011, 32 (03) :384-392
[7]  
Mlynski R., 2010, OPER TECH OTOLARYNGO, V21, P272
[8]   Evaluation of Round Window Stimulation Using the Floating Mass Transducer by Intracochlear Sound Pressure Measurements in Human Temporal Bones [J].
Nakajima, Hideko Heidi ;
Dong, Wei ;
Olson, Elizabeth S. ;
Rosowski, John J. ;
Ravicz, Michael E. ;
Merchant, Saumil N. .
OTOLOGY & NEUROTOLOGY, 2010, 31 (03) :506-511
[9]   Impact of Floating Mass Transducer Coupling and Positioning in Round Window Vibroplasty [J].
Rajan, Gunesh P. ;
Lampacher, Peter ;
Ambett, Ranjeeta ;
Dittrich, Gregor ;
Kuthubutheen, Jafri ;
Wood, Bradley ;
McArthur, Anne ;
Marino, Roberta .
OTOLOGY & NEUROTOLOGY, 2011, 32 (02) :271-277
[10]  
Service GJ, 2010, OPER TECH OTOLARYNGO, V21, P194