Pectoral cardioverter defibrillators: Comparison of prepectoral and submuscular implantation techniques

被引:18
作者
Manolis, AS
Chiladakis, J
Vassilikos, V
Maounis, T
Cokkinos, DV
机构
[1] Univ Patras, Patras, Greece
[2] Onassis Cardiac Surg Ctr, Athens, Greece
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1999年 / 22卷 / 03期
关键词
cardioverter defibrillators; ventricular tachyarrhythmias; sudden death; pectoral implantation techniques;
D O I
10.1111/j.1540-8159.1999.tb00475.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to compare the two techniques of pectoral ICD implantation, prepectoral and submuscular, performed by an electrophysiologist in the catheterization laboratory with use of general or local anesthesia in 45 consecutive patients. Over a period of 30 months, we implanted pectoral transvenous ICDs in 43 men and 2 women, aged 59 +/- 12 years, with use of general (n = 20) or local (n = 25) anesthesia in the catheterization laboratory. Patients had coronary (n = 30) or valvular (n = 4) disease, cardiomyopathy (n = 20) or no organic disease (n = 1), a mean left ventricular election fraction of 31%, and presented with ventricular tachycardia (n = 40) or fibrillation (n = 5). One-lead ICD systems (18 Endotak, 20 Transvene/8 Sprint, 2 EnGuard) were used in 38 patients, 2-lead (5 Transvene, 1 EnGuard) systems in 6 patients, and 2 atrioventricular lead ICD system in 1 patient. The prepectoral technique was employed in 29 patients with adequate subcutaneous tissue, while the sub-muscular technique was used in 2 6 patients who had a thin layer of subcutaneous tissue. The defibrillation threshold averaged 9-10 J in both groups and there were no differences in pace/sense thresholds. All implants were entirely transvenous with no subcutaneous patch. Biphasic ICD devices were employed in all patients. Active or hot can devices were used in 39 patients. There were no complications, operative deaths, or infections; Patients were discharged at a mean of 3 days. All devices functioned well at predischarge testing. Over 14 +/- 8 months, 20 patients received appropriate device therapy (antitachycardia pacing or shocks). No late complications occurred. One patient died at 3 months of pump failure; there were no sudden deaths. In conclusion, for exclusive pectoral implantation of transvenous ICDs, electrophysiologists should master both prepectoral and submuscular techniques. One can thus avoid potential skin erosion or need for abdominal implantation in patients with a thin layer of subcutaneous tissue. Finally, there are no differences in pacing or defibrillation thresholds between the two techniques.
引用
收藏
页码:469 / 478
页数:10
相关论文
共 24 条
[1]   Comparison of three cardioverter defibrillator implantation techniques: Initial results with transvenous pectoral implantation [J].
Anvari, A ;
Stix, G ;
Grabenwoger, M ;
Schneider, B ;
Turel, Z ;
Schmidinger, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (07) :1061-1069
[2]   A SIMPLIFIED, SINGLE-LEAD UNIPOLAR TRANSVENOUS CARDIOVERSION-DEFIBRILLATION SYSTEM [J].
BARDY, GH ;
JOHNSON, G ;
POOLE, JE ;
DOLACK, GL ;
KUDENCHUK, PJ ;
KELSO, D ;
MITCHELL, R ;
MEHRA, R ;
HOFER, B .
CIRCULATION, 1993, 88 (02) :543-547
[3]  
Estes NAM, 1994, AUTOMATIC IMPLANTABL
[4]   ELECTROPHYSIOLOGICAL LABORATORY, ELECTROPHYSIOLOGIST-IMPLANTED, NONTHORACOTOMY-IMPLANTABLE CARDIOVERTER/DEFIBRILLATORS [J].
FITZPATRICK, AP ;
LESH, MD ;
EPSTEIN, LM ;
LEE, RJ ;
SIU, A ;
MERRICK, S ;
GRIFFIN, JC ;
SCHEINMAN, MM .
CIRCULATION, 1994, 89 (06) :2503-2508
[5]   TECHNIQUE FOR IMPLANTATION OF CARDIOVERTER-DEFIBRILLATORS IN THE SUBPECTORAL POSITION [J].
FOSTER, AH .
ANNALS OF THORACIC SURGERY, 1995, 59 (03) :764-767
[6]   Complications associated with pectoral cardioverter-defibrillator implantation: Comparison of subcutaneous and submuscular approaches [J].
Gold, MR ;
Peters, RW ;
Johnson, JW ;
Shorofsky, SR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 28 (05) :1278-1282
[7]  
GROGLER FM, 1975, J THORAC CARDIOV SUR, V69, P895
[8]   SINGLE-INCISION IMPLANTATION OF CARDIOVERTER-DEFIBRILLATORS USING NONTHORACOTOMY LEAD SYSTEMS [J].
HAMMEL, D ;
BLOCK, M ;
GEIGER, A ;
BOCKER, D ;
STADLBAUER, T ;
BREITHARDT, G ;
SCHELD, HH .
ANNALS OF THORACIC SURGERY, 1994, 58 (06) :1614-1616
[9]   EVALUATING AND IMPROVING THE COST-EFFECTIVENESS OF THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
KUPERSMITH, J ;
HOGAN, A ;
GUERRERO, P ;
GARDINER, J ;
MELLITS, ED ;
BAUMGARDNER, R ;
ROVNER, D ;
HOLMESROVNER, M ;
MCLANE, A ;
LEVINE, J ;
SAKSENA, S ;
GRIFFITH, L .
AMERICAN HEART JOURNAL, 1995, 130 (03) :507-515
[10]   COST-EFFECTIVENESS OF THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFECT OF IMPROVED BATTERY LIFE AND COMPARISON WITH AMIODARONE THERAPY [J].
LARSEN, GC ;
MANOLIS, AS ;
SONNENBERG, FA ;
BESHANSKY, JR ;
ESTES, NAM ;
PAUKER, SG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (06) :1323-1334