Controlling flow in microfluidic channels with a manually actuated pin valve

被引:8
作者
Brett, Marie-Elena [1 ]
Zhao, Shuping [1 ]
Stoia, Jonathan L. [1 ]
Eddington, David T. [1 ]
机构
[1] Univ Illinois, Dept Bioengn, Chicago, IL 60607 USA
关键词
Microfluidics; Valve; On-chip flow control; Manual actuation; Polydimethylsiloxane; Islet;
D O I
10.1007/s10544-011-9533-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is a need for a simple method to control fluid flow within microfluidic channels. To meet this need, a simple push pin with a polydimethylsiloxane (PDMS) tip has been integrated into microfluidic networks to be placed within the microchannel to obstruct flow. This new valve design can attain on/off control of fluid flow without an external power source using readily-available, low-cost materials. The valve consists of a 14 gauge (1.6 mm) one inch piece of metal tubing with a PDMS pad at the tip to achieve a fluidic seal when pressed against a microfluidic channel's substrate. The metal tubing or pin is then either manually pushed down to block or pulled up to allow fluid flow. The valve was validated using a pressure transducer and fluorescent dye to determine the breakthrough pressure the valve can withstand over multiple cycles. In the first cycle, the median value for pressure withstood by the valve was 8.8 psi with a range of 17.5-2.7 psi. The pressure the valves were able to withstand during each successive trial was lower suggesting they may be most valuable as a method to control the initial introduction of fluids into a microfluidic device. These valves can achieve flow regulation within microfluidic devices, have a small dead volume, and are simple to fabricate and use, making this technique widely suitable for a range of applications.
引用
收藏
页码:633 / 639
页数:7
相关论文
共 12 条
[1]   Disposable Smart lab on a chip for point-of-care clinical diagnostics [J].
Ahn, CH ;
Choi, JW ;
Beaucage, G ;
Nevin, JH ;
Lee, JB ;
Puntambekar, A ;
Lee, JY .
PROCEEDINGS OF THE IEEE, 2004, 92 (01) :154-173
[2]   Functional hydrogel structures for autonomous flow control inside microfluidic channels [J].
Beebe, DJ ;
Moore, JS ;
Bauer, JM ;
Yu, Q ;
Liu, RH ;
Devadoss, C ;
Jo, BH .
NATURE, 2000, 404 (6778) :588-+
[3]   A Microfabricated Rotary Pump [J].
Chou, Hou-Pu ;
Unger, Marc A. ;
Quake, Stephen R. .
BIOMEDICAL MICRODEVICES, 2001, 3 (04) :323-330
[4]   Integration of a zero dead-volume PDMS rotary switch valve in a miniaturised (bio)electroanalytical system [J].
Godino, Neus ;
Javier del Campo, Francisco ;
Xavier Munoz, Francesc ;
Hansen, Mikkel Fougt ;
Kutter, Jorg P. ;
Snakenborg, Detlef .
LAB ON A CHIP, 2010, 10 (14) :1841-1847
[5]   Computerized microfluidic cell culture using elastomeric channels and Braille displays [J].
Gu, W ;
Zhu, XY ;
Futai, N ;
Cho, BS ;
Takayama, S .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2004, 101 (45) :15861-15866
[6]   Microfluidics in structural biology: smaller, faster... better [J].
Hansen, C ;
Quake, SR .
CURRENT OPINION IN STRUCTURAL BIOLOGY, 2003, 13 (05) :538-544
[7]   Incorporation of prefabricated screw, pneumatic, and solenoid valves into microfluidic devices [J].
Hulme, S. Elizabeth ;
Shevkoplyas, Sergey S. ;
Whitesides, George M. .
LAB ON A CHIP, 2009, 9 (01) :79-86
[8]   Tape underlayment rotary-node (TURN) valves for simple on-chip microfluidic flow control [J].
Markov, Dmitry A. ;
Manuel, Steven ;
Shor, Leslie M. ;
Opalenik, Susan R. ;
Wikswo, John P. ;
Samson, Philip C. .
BIOMEDICAL MICRODEVICES, 2010, 12 (01) :135-144
[9]   A review of microvalves [J].
Oh, Kwang W. ;
Ahn, Chong H. .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2006, 16 (05) :R13-R39
[10]   Independent optical control of microfluidic valves formed from optomechanically responsive nanocomposite hydrogels [J].
Sershen, SR ;
Mensing, GA ;
Ng, M ;
Halas, NJ ;
Beebe, DJ ;
West, JL .
ADVANCED MATERIALS, 2005, 17 (11) :1366-+