Viscosity-adjusted estimation of pressure head and pump flow with quasi-pulsatile modulation of rotary blood pump for a total artificial heart

被引:7
作者
Yurimoto, Terumi [1 ]
Hara, Shintaro [1 ]
Isoyama, Takashi [1 ]
Saito, Itsuro [1 ]
Ono, Toshiya [1 ]
Abe, Yusuke [1 ]
机构
[1] Univ Tokyo, Grad Sch Med, Dept Biomed Engn, Bunkyo Ku, 7-3-1 Hongo, Tokyo 1130033, Japan
关键词
Table estimation method; Rotary blood pump; Pulsatile; Real time; Viscosity; ASSIST; SYSTEM;
D O I
10.1007/s10047-016-0898-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Estimation of pressure and flow has been an important subject for developing implantable artificial hearts. To realize real-time viscosity-adjusted estimation of pressure head and pump flow for a total artificial heart, we propose the table estimation method with quasi-pulsatile modulation of rotary blood pump in which systolic high flow and diastolic low flow phased are generated. The table estimation method utilizes three kinds of tables: viscosity, pressure and flow tables. Viscosity is estimated from the characteristic that differential value in motor speed between systolic and diastolic phases varies depending on viscosity. Potential of this estimation method was investigated using mock circulation system. Glycerin solution diluted with salty water was used to adjust viscosity of fluid. In verification of this method using continuous flow data, fairly good estimation could be possible when differential pulse width modulation (PWM) value of the motor between systolic and diastolic phases was high. In estimation under quasi-pulsatile condition, inertia correction was provided and fairly good estimation was possible when the differential PWM value was high, which was not different from the verification results using continuous flow data. In the experiment of real-time estimation applying moving average method to the estimated viscosity, fair estimation could be possible when the differential PWM value was high, showing that real-time viscosity-adjusted estimation of pressure head and pump flow would be possible with this novel estimation method when the differential PWM value would be set high.
引用
收藏
页码:219 / 225
页数:7
相关论文
共 11 条
[1]   Animal Experiments of the Helical Flow Total Artificial Heart [J].
Abe, Yusuke ;
Isoyama, Takashi ;
Saito, Itsuro ;
Inoue, Yusuke ;
Ishii, Kohei ;
Sato, Masami ;
Hara, Shintaro ;
Yurimoto, Terumi ;
Li, Xinyang ;
Murakami, Haruka ;
Ariyoshi, Koki ;
Kawase, Yukino ;
Ono, Toshiya ;
Fukazawa, Kyoko ;
Ishihara, Kazuhiko .
ARTIFICIAL ORGANS, 2015, 39 (08) :670-680
[2]   The helical flow pump with a hydrodynamic levitation impeller [J].
Abe, Yusuke ;
Ishii, Kohei ;
Isoyama, Takashi ;
Saito, Itsuro ;
Inoue, Yusuke ;
Ono, Toshiya ;
Nakagawa, Hidemoto ;
Nakano, Emiko ;
Fukazawa, Kyoko ;
Ishihara, Kazuhiko ;
Fukunaga, Kazuyoshi ;
Ono, Minoru ;
Imachi, Kou .
JOURNAL OF ARTIFICIAL ORGANS, 2012, 15 (04) :331-340
[3]   Sensorless flow and head estimation in the VentrAssist rotary blood pump [J].
Ayre, PJ ;
Vidakovic, SS ;
Tansley, GD ;
Watterson, PA ;
Lovell, NH .
ARTIFICIAL ORGANS, 2000, 24 (08) :585-588
[4]   Flow rate and pressure head estimation in a centrifugal blood pump [J].
Funakubo, A ;
Ahmed, S ;
Sakuma, I ;
Fukui, Y .
ARTIFICIAL ORGANS, 2002, 26 (11) :985-990
[5]   Physical model-based indirect measurements of blood pressure and flow using a centrifugal pump [J].
Kitamura, T ;
Matsushima, Y ;
Tokuyama, T ;
Kono, F ;
Nishimura, F ;
Komeda, M ;
Yanai, M ;
Kijima, T ;
Nojiri, C .
ARTIFICIAL ORGANS, 2000, 24 (08) :589-593
[6]   Noninvasive average flow estimation for an implantable rotary blood pump: A new algorithm incorporating the role of blood viscosity [J].
Malagutti, Nicolo ;
Karantonis, Dean M. ;
Cloherty, Shaun L. ;
Ayre, Peter J. ;
Mason, David G. ;
Salamonsen, Robert F. ;
Lovell, Nigel H. .
ARTIFICIAL ORGANS, 2007, 31 (01) :45-52
[7]  
Morshuis M, 2010, EXPERT REV MED DEVIC, V7, P173, DOI [10.1586/erd.09.68, 10.1586/ERD.09.68]
[8]   Intraoperative Evaluation of the HeartMate II Flow Estimator [J].
Slaughter, Mark S. ;
Bartoli, Carlo R. ;
Sobieski, Mike A. ;
Pantalos, George M. ;
Giridharan, Guruprasad A. ;
Dowling, Robert D. ;
Prabhu, Sumanth D. ;
Farrar, David J. ;
Koenig, Steven C. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (01) :39-43
[9]   In vivo test of pressure head and flow rate estimation in a continuous-flow artificial heart [J].
Tanaka, A ;
Yoshizawa, M ;
Abe, K ;
Takeda, H ;
Yambe, T ;
Nitta, S .
ARTIFICIAL ORGANS, 2003, 27 (01) :99-103
[10]   Development of a flow estimation and control system of an implantable centrifugal blood pump for circulatory assist [J].
Wakisaka, Y ;
Okuzono, Y ;
Taenaka, Y ;
Chikanari, K ;
Endo, S ;
Masuzawa, T ;
Takano, H .
ARTIFICIAL ORGANS, 1998, 22 (06) :488-492