Dose Reduction in Pediatric Computed Tomography with Automated Exposure Control

被引:30
作者
Alibek, Sedat [1 ]
Brand, Martin [2 ]
Suess, Christoph [3 ]
Wuest, Wolfgang [1 ]
Uder, Michael [1 ]
Greess, Holger [4 ]
机构
[1] Univ Erlangen Nurnberg, Inst Radiol, Erlangen, Germany
[2] Klinikum Nordoberpfalz AG, Dept Neurol, Weiden, Germany
[3] Siemens Healthcare, Forchheim, Germany
[4] Kreiskliniken Esslingen, Clin Diagnost Radiol, Ruit, Germany
关键词
Computed tomography; pediatric; dose reduction; image quality; MULTISLICE SPIRAL CT; TUBE CURRENT;
D O I
10.1016/j.acra.2011.01.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Since the introduction of computed tomographic (CT) imaging in the 1970s, the number of examinations has increased steadily. CT imaging is an essential part of routine workup in diagnostic radiology. The great advantage of multidetector computed tomography is the acquisition of a large amount of data in a short time period, thus speeding up diagnostic procedures. To protect patients from unnecessary radiation exposure, different approaches have been developed. In this study, the efficacy of automated exposure control (AEC) software in multidetector CT imaging with a focus on dose reduction in pediatric examinations was assessed. Materials and Methods: Between August 2004 and September 2005, a total of 71 children (40 male, 31 female; age range, 2-13 years; mean age, 7.2 years) were examined using a multisource CT scanner. Three different regions (chest, upper abdomen, and pelvis) were examined. Overall image quality was assessed with a subjective scale (1 = excellent, 2 = diagnostic, 3 = nondiagnostic). For all examinations, AEC was used. From the scanner's patient protocol, dose-length product, volume CT dose index, and tube current-time product were calculated for each examination. Results: With AEC, a mean dose reduction of 30.6% was calculated. Images were rated as excellent (n = 39) or diagnostic (n = 32). Nondiagnostic image quality was not seen. Dose-length product and volume CT dose index were reduced by 30.4% and 29.5%, respectively. Overall, a mean dose reduction of 30.1% of the effective dose (5.8 +/- 3.1 vs 8.4 +/- 4.6 mSv) was achieved (P < .001). Conclusions: With AEC software, a mean dose reduction of 30% without any loss in diagnostic image quality is possible.
引用
收藏
页码:690 / 693
页数:4
相关论文
共 10 条
  • [1] Estimated risks of radiation-induced fatal cancer from pediatric CT
    Brenner, DJ
    Elliston, CD
    Hall, EJ
    Berdon, WE
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) : 289 - 296
  • [2] Individually adapted examination protocols for reduction of radiation exposure for 16-MDCT chest examinations
    Das, M
    Mahnken, AH
    Mühlenbruch, G
    Stargardt, A
    Weiss, C
    Sennst, DA
    Flohr, TG
    Günther, RW
    Wildberger, JE
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (05) : 1437 - 1443
  • [3] Radiosensitivity of children: potential for overexposure in CR and DR and magnitude of doses in ordinary radiographic examinations
    Don, S
    [J]. PEDIATRIC RADIOLOGY, 2004, 34 (Suppl 3) : S167 - S172
  • [4] Automatic exposure control to reduce the dose in subsecond multislice spiral CT: Phantom measurements and clinical results
    Greess, H
    Wolf, H
    Suess, C
    Kalender, WA
    Bautz, W
    Baum, U
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (06): : 862 - 869
  • [5] Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation
    Greess, H
    Lutze, J
    Nömayr, A
    Wolf, H
    Hothorn, T
    Kalender, WA
    Bautz, W
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (06) : 995 - 999
  • [6] Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions
    Greess, H
    Wolf, H
    Baum, U
    Lell, M
    Pirkl, M
    Kalender, W
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 2000, 10 (02) : 391 - 394
  • [7] Dose reduction in spiral CT:: Detection of pulmonary nodules with and without anatomic adaptation of tube current.
    Greess, H
    Baum, U
    Wolf, H
    Lell, M
    Nömayr, A
    Schmidt, B
    Kalender, WA
    Bautz, W
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2001, 173 (05): : 466 - 470
  • [8] Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers
    Kalender, WA
    Wolf, H
    Suess, C
    Gies, M
    Greess, H
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (02) : 323 - 328
  • [9] KALENDER WA, 2006, COMPUTERTOMOGRAPHIE, P216
  • [10] CT-Expo - a novel program for dose evaluation in CT
    Stamm, G
    Nagel, HD
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (12): : 1570 - 1576