Report of adverse event with electroacupuncture

被引:4
作者
Cummings, Mike [1 ]
机构
[1] Royal London Hosp Integrated Med, British Med Acupuncture Soc, London WC1N 3HR, England
关键词
ACUPUNCTURE;
D O I
10.1136/acupmed-2011-010020
中图分类号
R [医药、卫生];
学科分类号
10 ;
摘要
Electroacupuncture (EA) is becoming more common as a treatment for chronic musculoskeletal pain. It can be associated with adverse events related to the small electric currents used, in addition to the adverse events related to needle penetration of tissues. This paper reports a case of minor tissue damage following high intensity EA for 30 min with a device delivering a waveform that does not appear to be completely charge-balanced. This case highlights a rare but preventable adverse event. Manufacturers should be encouraged to develop EA stimulators that use charge-balanced waveforms.