Long Segment Blocking Screws Increase the Stability of Retrograde Nail Fixation in Geriatric Supracondylar Femur Fractures: Eliminating the "Bell-Clapper Effect"

被引:20
作者
Auston, Darryl [1 ]
Donohue, David [2 ,3 ]
Stoops, Kyle [4 ]
Cox, Jacob [2 ,3 ]
Diaz, Miguel [4 ]
Santoni, Brandon [4 ]
Mir, Hassan [2 ,3 ]
机构
[1] Orange Pk Med Ctr, Hughston Orthopaed Trauma, Orange Pk, FL USA
[2] Univ S Florida, Dept Orthopaed Surg, Tampa, FL USA
[3] Florida Orthopaed Inst, Orthopaed Trauma Serv, Tampa, FL USA
[4] Fdn Orthopaed Res & Educ, Tampa, FL USA
关键词
DISTAL FEMORAL FRACTURES; BIOMECHANICAL EVALUATION; INTERNAL-FIXATION; TIBIAL FRACTURES; STANDARD;
D O I
10.1097/BOT.0000000000001284
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To determine the change in stiffness and horizontal translations of a geriatric supracondylar femur fracture model with the addition of distal segment blocking screws versus proximal (long) segment blocking screws to the standard retrograde intramedullary nail construct. Methods: Unstable supracondylar femur fractures (OTA/AO 33-A3) were created; all specimens were instrumented with a retrograde intramedullary nail. Specimens were divided into 2 groups (6 matched pairs per group). Group 1 compared the standard construct (1 proximal screw and 3 distal screws) to a distally augmented construct, with blocking screws placed in the distal metaphyseal segment. Group 2 compared the distally augmented construct to one in which blocking screws were placed just proximal to the fracture (long segment blocking screws). Specimens were then axially loaded and cycled to failure or run-out. Results: There was no significant difference in baseline stiffness, survival through cyclic loading, stiffness after cyclic loading, or cycles to failure between femurs treated with distal blocking screws and femurs treated without blocking screws (group 1). Femurs with blocking screws in the long proximal segment had significantly greater baseline stiffness, stiffness after cyclic loading, and less horizontal translation at the fracture site (group 2). There was no difference in survival through cyclic loading or cycles to failure. Conclusion: Long segment blocking screws are biomechanically superior to blocking screws in the distal segment or no blocking screws initially and after cyclic loading in an unstable geriatric supracondylar femur fracture model treated with intramedullary nail.
引用
收藏
页码:559 / 564
页数:6
相关论文
共 19 条
  • [1] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [2] Chan DS, 2015, J ORTHOP TRAUMA, V29, P98, DOI 10.1097/BOT.0000000000000195
  • [3] Biomechanical Evaluation of Interlocking Lag Screw Design in Intramedullary Nailing of Unstable Pertrochanteric Fractures
    Hoffmann, Stefanie
    Paetzold, Robert
    Stephan, Daniel
    Pueschel, Klaus
    Buehren, Volker
    Augat, Peter
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2013, 27 (09) : 483 - 490
  • [4] Unreamed intramedullary nailing of femoral shaft fractures: Operative technique and early clinical experience with the standard locking option
    Krettek, C
    Rudolf, J
    Schandelmaier, P
    Guy, P
    Konemann, B
    Tscherne, H
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1996, 27 (04): : 233 - 254
  • [5] The use of Poller screws as blocking screws in stabilising tibial fractures treated with small diameter intramedullary nails
    Krettek, C
    Stephan, C
    Schandelmaier, P
    Richter, M
    Papa, HC
    Miclau, T
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (06): : 963 - 968
  • [6] Nüchtern JV, 2014, J ORTHOP TRAUMA, V28, P276, DOI 10.1097/BOT.0000000000000008
  • [7] Distal Third Femur Fractures Treated With Retrograde Femoral Nailing and Blocking Screws
    Ostrum, Robert F.
    Maurer, James P.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (09) : 681 - 684
  • [8] The role and efficacy of retrograding nailing for the treatment of diaphyseal and distal femoral fractures: a systematic review of the literature
    Papadokostakis, G
    Papakostidis, C
    Dimitriou, R
    Giannoudis, PV
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2005, 36 (07): : 813 - 822
  • [9] Retrograde intramedullary nails with distal screws locked to the nail have higher fatigue strength than locking plates in the treatment of supracondylar femoral fractures A CADAVER-BASED LABORATORY INVESTIGATION
    Pekmezci, M.
    McDonald, E.
    Buckley, J.
    Kandemir, U.
    [J]. BONE & JOINT JOURNAL, 2014, 96B (01) : 114 - 121
  • [10] Biomechanical evaluation of dual-energy x-ray absorptiometry for predicting fracture loads of the infant femur for injury investigation: An in vitro porcine model
    Pierce, MC
    Valdevit, A
    Anderson, L
    Inoue, N
    Hauser, DL