Calcitonin levels in normal individuals with new highly sensitive chemiluminescent enzyme immunoassay

被引:0
作者
Suzuki, H [1 ]
机构
[1] Fujirebio Inc, Diagnost Strateg Planning Dept, Chuo Ku, Tokyo 1030007, Japan
关键词
human calcitonin; human serum; enzyme immunoassay; chemiluminescence; age;
D O I
10.1002/(SICI)1098-2825(1998)12:4<218::AID-JCLA5>3.0.CO;2-3
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Human serum calcitonin concentration in normal individuals was measured with a new assay based on the chemiluminescent enzyme immunoassay (CLEIA) method. The CLEIA assay was highly sensitive and was able to determine a calcitonin concentration of 0.04 pg/ml as sensitivity limit at a condition of 0+3SD. With this CLEIA assay, the mean value of calcitonin in males and females was 2.26 and 1.33 pg/ml, respectively, highlighting a significant difference between genders. The mean value and range of human serum calcitonin in this assay were approximately 1/10 those reported previously in competitive radioimmunoassay (RIA) methods. Since RIAs for calcitonin showed much variability at a low concentration range due to the competitive format, they seemed to lack the necessary sensitivity to cover the normal range and appeared only useful for hyper-calcitonin phenomenon in diseases such as medullary thyroid carcinoma. The CLEIA for calcitonin provided a lower detection limit than normal range, and it can therefore be assumed that it could be applied for the measurement of hypo-calcitonin phenomena typically found in some disorders such as osteoporosis. (C) 1998 Wiley-Liss, Inc.
引用
收藏
页码:218 / 222
页数:5
相关论文
共 8 条