A quality assurance procedure to evaluate cone-beam CT image center congruence with the radiation isocenter of a linear accelerator

被引:14
作者
Du, Weiliang [1 ]
Yang, James N.
Chang, Eric L. [2 ]
Luo, Dershan
McAleer, Mary Frances [2 ]
Shiu, Almon
Martel, Mary K.
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 94, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2010年 / 11卷 / 04期
关键词
CBCT; isocenter; IGRT; SBRT; COMPUTED-TOMOGRAPHY; THERAPY; RADIOSURGERY; SYSTEM; QA;
D O I
10.1120/jacmp.v11i4.3297
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A quality assurance (QA) procedure was developed to evaluate the congruence between the cone-beam computed tomography (CBCT) image center and the radiation isocenter on a Varian Trilogy linac. In contrast to the published QA procedures, this method did not require a ball bearing (BB) phantom to be placed exactly at the radiation isocenter through precalibrated room lasers or light field crosshairs. The only requirement was that the BB phantom be in a stationary position near the radiation isocenter during the image acquisition process. The radiation isocenter was determined with respect to the center of the BB using a Winston-Lutz test. The CBCT image center was found to have excellent short-term positional reproducibility (i.e., less than 0.1 mm of wobble in each of the x (lateral), y (vertical), and z (longitudinal) directions) in 10 consecutive acquisitions. Measured over a seven-month period, the CBCT image center deviated from the radiation isocenter by 0.40 +/- 0.12 mm (x), 0.43 +/- 0.04 mm (y), and 0.34 +/- 0.14 mm (z). The z displacement of the 3D CBCT image center was highly correlated (p = 0.997) with that of the 2D kV portal image center. The correlation coefficients in the x and y directions were poor (p = 0.66 and -0.35, respectively). Systematic discrepancies were found between the CBCT image center and the 2D MV, kV portal image centers. For the linear accelerator studied, we detected a 0.8 mm discrepancy between the CBCT image center and the MV EPID image center in the anterior-posterior direction. This discrepancy was demonstrated in a clinical case study where the patient was positioned with CBCT followed by MV portal verification. The results from the new QA procedure are useful for guiding high-precision patient positioning in stereotactic body radiation therapy.
引用
收藏
页码:15 / 26
页数:12
相关论文
共 18 条
  • [1] Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone-beam CT from kV on-board imagers
    Ali, Imad
    Ahmad, Salahuddin
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (03): : 180 - 194
  • [2] A simple method to quantify the coincidence between portal image graticules and radiation field centers or radiation isocenter
    Du, Weiliang
    Yang, James
    Luo, Dershan
    Martel, Mary
    [J]. MEDICAL PHYSICS, 2010, 37 (05) : 2256 - 2263
  • [3] A robust Hough transform algorithm for determining the radiation centers of circular and rectangular fields with subpixel accuracy
    Du, Weiliang
    Yang, James
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (03) : 555 - 567
  • [4] Guan H, 2009, J APPL CLIN MED PHYS, V10, P2954, DOI DOI 10.1120/jacmp.v10i4.2954
  • [5] Clinical use of electronic portal imaging: Report of AAPM radiation therapy committee Task Group 58
    Herman, MG
    Balter, JM
    Jaffray, DA
    McGee, KP
    Munro, P
    Shalev, S
    Van Herk, M
    Wong, JW
    [J]. MEDICAL PHYSICS, 2001, 28 (05) : 712 - 737
  • [6] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [7] Task Group 142 report: Quality assurance of medical accelerators
    Klein, Eric E.
    Hanley, Joseph
    Bayouth, John
    Yin, Fang-Fang
    Simon, William
    Dresser, Sean
    Serago, Christopher
    Aguirre, Francisco
    Ma, Lijun
    Arjomandy, Bijan
    Liu, Chihray
    Sandin, Carlos
    Holmes, Todd
    [J]. MEDICAL PHYSICS, 2009, 36 (09) : 4197 - 4212
  • [8] Comparison of kilovoltage cone-beam computed tomography with megavoltage projection pairs for paraspinal radiosurgery patient alignment and position verification
    Kriminski, Sergey A.
    Lovelock, D. Michael
    Seshan, Venkatraman E.
    Ali, Imad
    Munro, Peter
    Amols, Howard I.
    Fuks, Zvi
    Bilsky, Mark
    Yamada, Yoshiya
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (05): : 1572 - 1580
  • [9] COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40
    KUTCHER, GJ
    COIA, L
    GILLIN, M
    HANSON, WF
    LEIBEL, S
    MORTON, RJ
    PALTA, JR
    PURDY, JA
    REINSTEIN, LE
    SVENSSON, GK
    WELLER, M
    WINGFIELD, L
    [J]. MEDICAL PHYSICS, 1994, 21 (04) : 581 - 618
  • [10] Cone-beam-CT guided radiation therapy:: technical implementation
    Létourneau, D
    Wong, JW
    Oldham, M
    Gulam, M
    Watt, L
    Jaffray, DA
    Siewerdsen, JH
    Martinez, AA
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 75 (03) : 279 - 286