Detection performance of a portable ion mobility spectrometer with 63Ni radioactive ionization for chemical warfare agents

被引:40
作者
Yamaguchi, Shintaro [1 ]
Asada, Ryuji [1 ]
Kishi, Shintaro [1 ]
Sekioka, Ryoji [1 ]
Kitagawa, Nobuyoshi [2 ]
Tokita, Kenichi [2 ]
Yamamoto, Soichiro [2 ]
Seto, Yasuo [1 ]
机构
[1] Natl Res Inst Police Sci, Dept Forens Sci 3, Kashiwa, Chiba 2770882, Japan
[2] ST Japan Inc, Tokyo, Japan
关键词
Chemical warfare agents; Ion mobility spectrometer; SABRE; 4000; Ni-63; ionization; Portable detector; Sarin; SITE DETECTION METHOD;
D O I
10.1007/s11419-010-0092-z
中图分类号
R99 [毒物学(毒理学)];
学科分类号
100405 ;
摘要
The detection performance of a portable ion mobility spectrometer (IMS) (SABRE 4000, Smiths Detection) with Ni-63 ionization, air purification, and reduced ion mobility measurements using calibrants was investigated for vapors of chemical warfare agents. In a matter of several seconds, the SABRE 4000 enabled tentative identification of sarin, soman, cyclohexylsarin, tabun, and nitrogen mustard 3, each with a limit of alarm (LOA) of 0.005-0.5 mg m(-3) in positive ion collection mode. Hydrogen cyanide could be identified with a LOA of 0.2 mg m(-3) in the negative mode. Mustard gas, nitrogen mustards 1, 2, and 3, phosgene, and chloropicrin showed a positive alarm of "HD/Phos" with a LOA of 0.2-2 mg m(-3) in negative mode. Lewisite 1, cyanogen chloride, and chlorine vapors did not show any alarm, although the characteristic ion peaks appeared in negative mode. The peak areas of the respective detected ions on the second differentiation spectra were positively correlated with the corresponding vapor concentrations in low concentration ranges. Several chemical agent simulants and organic solvents were also examined for detection by the machine. The SABRE 4000 is superior to other portable IMS machines at this time, although some improvements of the system are necessary.
引用
收藏
页码:84 / 95
页数:12
相关论文
共 32 条
[1]  
CARRICO JP, 1986, AM LAB, V18, P152
[2]  
Cottingham K, 2003, ANAL CHEM, V75, p202A
[3]   Ion mobility spectrometers in national defense. [J].
Eiceman, GA ;
Stone, JA .
ANALYTICAL CHEMISTRY, 2004, 76 (21) :390A-397A
[4]   ENHANCED SELECTIVITY IN ION MOBILITY SPECTROMETRY ANALYSIS OF COMPLEX-MIXTURES BY ALTERNATE REAGENT GAS CHEMISTRY [J].
EICEMAN, GA ;
WANG, YF ;
GARCIAGONZALEZ, L ;
HARDEN, CS ;
SHOFF, DB .
ANALYTICA CHIMICA ACTA, 1995, 306 (01) :21-33
[5]   Technology challenges in responding to biological or chemical attacks in the civilian sector [J].
Fitch, JP ;
Raber, E ;
Imbro, DR .
SCIENCE, 2003, 302 (5649) :1350-1354
[7]   ION MOBILITY SPECTROMETRY [J].
HILL, HH ;
SIEMS, WF ;
STLOUIS, RH ;
MCMINN, DG .
ANALYTICAL CHEMISTRY, 1990, 62 (23) :A1201-A1209
[8]   Rapid preseparation of interferences for ion mobility spectrometry [J].
Kanu, Abu B. ;
Wu, Ching ;
Hill, Herbert H., Jr. .
ANALYTICA CHIMICA ACTA, 2008, 610 (01) :125-134
[9]   ION MOBILITY SPECTROMETRY MASS-SPECTROMETRY OF 2 STRUCTURALLY DIFFERENT IONS HAVING IDENTICAL ION MASS [J].
KIM, SH ;
SPANGLER, GE .
ANALYTICAL CHEMISTRY, 1985, 57 (02) :567-569
[10]   Evaluation of Detection Performance of Portable Aspiration-Type Ion Mobility Spectrometer with Seven Detection Cells for Chemical Warfare Agents [J].
Kishi, Shintaro ;
Sekioka, Ryoji ;
Sodeyama, Manabu ;
Shiga, Masao ;
Seto, Yasuo .
BUNSEKI KAGAKU, 2010, 59 (01) :65-76