Evaluation of a Novel Non-Diffractive Extended Depth of Focus Intraocular Lens - First Results from a Prospective Study

被引:12
作者
Tahmaz, Volkan [1 ]
Siebelmann, Sebastian [1 ,2 ]
Koch, Konrad R. [2 ]
Cursiefen, Claus [1 ]
Langenbucher, Achim [3 ]
Hoerster, Robert [1 ,2 ]
机构
[1] Univ Cologne, Ctr Ophthalmol, Cologne, Germany
[2] MVZ ADTC Moenchengladbach Erkelenz, Ostpromenade 41, D-41812 Erkelenz, Germany
[3] Univ Saarland, Dept Expt Ophthalmol, Homburg, Germany
关键词
Cataract surgery; extended depth of focus; non-diffractive intraocular lens; presbyopia; VISUAL PERFORMANCE; OF-FOCUS; PATIENT SATISFACTION; CLINICAL-OUTCOMES; SUBJECTIVE DEPTH; DISTANCE; VISION;
D O I
10.1080/02713683.2022.2074046
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose :To evaluate a novel hydrophobic, non-diffractive, extended depth of focus (EDOF) intraocular lens (IOL) design in comparison to two monofocal aspheric lenses. Methods : Inclusion criteria for this prospective, monocentric cohort study were opacification of the crystalline lens and patients' wishes for surgery. In the case of the EDOF IOL, patients asked for a presbyopia correction. All patients received surgery on both eyes. Corrected and uncorrected distance visual acuity (CDVA, UCDVA), uncorrected and distance corrected intermediate visual acuity (UIVA, DCIVA) and defocus curves (all monocular and binocular) were compared three months postoperatively. Results : Fifty-six eyes were implanted with an EDOF IOL (LuxSmart (TM), Bausch & Lomb GmbH, Berlin, Germany), 50 eyes with a monofocal aspheric IOL: 32 eyes with a clear IOL (Polylens (R) AS 61, Polytech Domilens, Rossdorf, Germany), 16 eyes with a yellow IOL (iSert (R) 251, Hoya Surgical Optics GmbH, Frankfurt, Germany). Three months postoperatively, UCDVA was comparable with the EDOF IOL, versus the monofocal IOL (P > 0.9). Binocular DCIVA in the EDOF IOL was significantly higher than in the monofocal IOL (P = 0.001). Monocular DCIVA better than 20/23 Snellen was achieved in 10% with the monofocal IOL and in 68% (P < 0.0001) with the EDOF IOL. Defocus curves showed a depth of focus at 20/23 Snellen of 1.6 vs. 0.83 diopters (D) in the EDOF IOL, vs. the monofocal IOL. No patient reported halos or starbursts in non-standardized questioning. Conclusion :This non-diffractive EDOF IOL provided comparably high UCDVA and significantly higher DCIVA than the mono-focal lenses, causing only mild optical phenomena.
引用
收藏
页码:1149 / 1155
页数:7
相关论文
共 20 条
[1]   A Comparative Evaluation of Visual, Refractive, and Patient-Reported Outcomes of Three Extended Depth of Focus (EDOF) Intraocular Lenses [J].
Ang, Robert Edward ;
Picache, Gian Carlo S. ;
Rivera, Mark Christian R. ;
Lopez, Lord Ryan L. ;
Cruz, Emerson M. .
CLINICAL OPHTHALMOLOGY, 2020, 14 :2339-2351
[2]   Optimizing the subjective depth-of-focus with combinations of fourth- and sixth-order spherical aberration [J].
Benard, Yohann ;
Lopez-Gil, Norberto ;
Legras, Richard .
VISION RESEARCH, 2011, 51 (23-24) :2471-2477
[3]   Subjective depth of field in presence of 4th-order and 6th-order Zernike spherical aberration using adaptive optics technology [J].
Benard, Yohann ;
Lopez-Gil, Norberto ;
Legras, Richard .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (12) :2129-2138
[4]   Preliminary Clinical Outcomes of a New Enhanced Depth of Focus Intraocular Lens [J].
Campos, Nuno ;
Loureiro, Tomas ;
Rodrigues-Barros, Sandra ;
Carreira, Ana Rita ;
Gouveia-Moraes, Filipe ;
Carreira, Pedro ;
Machado, Ines .
CLINICAL OPHTHALMOLOGY, 2021, 15 :4801-4807
[5]   Clinical outcomes of a new extended range of vision intraocular lens: International Multicenter Concerto Study [J].
Cochener, Beatrice .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (09) :1268-1275
[6]   Multifocal versus monofocal intraocular lenses after cataract extraction [J].
de Silva, Samantha R. ;
Evans, Jennifer R. ;
Kirthi, Varo ;
Ziaei, Mohammed ;
Leyland, Martin .
COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2016, (12)
[7]   Visual and Refractive Outcomes following Bilateral Implantation of Extended Range of Vision Intraocular Lens with Micromonovision [J].
Ganesh, Sri ;
Brar, Sheetal ;
Pawar, Archana ;
Relekar, Kirti J. .
JOURNAL OF OPHTHALMOLOGY, 2018, 2018
[8]   Extended depth-of-focus technology in intraocular lenses [J].
Kohnen, Thomas ;
Suryakumar, Rajaraman .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (02) :298-304
[9]   Visual performance of an extended depth of focus intraocular lens for treatment selection [J].
Kohnen, Thomas ;
Boehm, Myriam ;
Hemkeppler, Eva ;
Schoenbrunn, Sabrina ;
DeLorenzo, Nina ;
Petermann, Kerstin ;
Herzog, Michael .
EYE, 2019, 33 (10) :1556-1563
[10]   The Influence of Induced Astigmatism on the Depth of Focus [J].
Leube, Alexander ;
Ohlendorf, Arne ;
Wahl, Siegfried .
OPTOMETRY AND VISION SCIENCE, 2016, 93 (10) :1228-1234