Pulsating potentiometric titration technique for assay of dissolved oxygen in water at trace level

被引:24
作者
Sahoo, P. [1 ]
Ananthanarayanan, R. [1 ]
Malathi, N. [1 ]
Rajiniganth, M. P. [1 ]
Murali, N. [1 ]
Swaminathan, P. [1 ]
机构
[1] Indira Gandhi Ctr Atom Res, EIG, Innovat Instrumentat Sect, Real Time Syst Div, Kalpakkam 603102, Tamil Nadu, India
关键词
Dissolved oxygen; Potentiometric titration; Pulsating sensor; Winkler titration; Voltage to frequency converter; Dissolved oxygen sampler; INJECTION-ANALYSIS; SENSOR;
D O I
10.1016/j.aca.2010.04.040
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple but high performance potentiometric titration technique using pulsating sensors has been developed for assay of dissolved oxygen (DO) in water samples clown to 10.0 mu g L-1 levels. The technique involves Winkler titration chemistry, commonly used for determination of dissolved oxygen in water at mg L-1 levels, with modification in methodology for accurate detection of end point even at 10.0 mu g L-1 levels DO present in the sample. An indigenously built sampling cum pretreatment vessel has been deployed for collection and chemical fixing of dissolved oxygen in water samples from flowing water line without exposure to air. A potentiometric titration facility using pulsating sensors developed in-house is used to carry out titration. The power of the titration technique has been realised in estimation of very dilute solution of iodine equivalent to 10 mu g L-1 O-2. Finally, several water samples containing dissolved oxygen from mg L-1 to mu g L-1 levels were successfully analysed with excellent reproducibility using this new technique. The precision in measurement of DO in water at 10 mu g L-1 O-2 level is 0.14 (n = 5), RSD: 1.4%. Probably for the first time a potentiometric titration technique has been successfully deployed for assay of dissolved oxygen in water samples at 10 mu g L-1 levels. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:17 / 24
页数:8
相关论文
共 18 条
[1]  
[Anonymous], 2009, ASTM Standard C29, DOI [DOI 10.1520/D6951, 10.1520/C0029C0029M-17A]
[2]  
[Anonymous], 1976, STAND METH EX WAT WA, P440
[3]   Nondestructive monitoring of oxygen profiles in packaged foods using phase-fluorimetric oxygen sensor [J].
Fitzgerald, M ;
Papkovsky, DB ;
Smiddy, M ;
Kerry, JP ;
O'Sullivan, CK ;
Buckley, DJ ;
Guilbault, GG .
JOURNAL OF FOOD SCIENCE, 2001, 66 (01) :105-110
[4]   A novel sensor based on the porous plastic probe for determination of dissolved oxygen in seawater [J].
Guo, Liangqia ;
Ni, Qinyan ;
Li, Jiaqian ;
Zhang, Lan ;
Lin, Xucong ;
Xie, Zenghong ;
Chen, Guonan .
TALANTA, 2008, 74 (04) :1032-1037
[5]  
*HATCH, 10360 HATCH
[6]   Micro-Winkler titration method for dissolved oxygen concentration measurement [J].
Helm, Irja ;
Jalukse, Lauri ;
Vilbaste, Martin ;
Leito, Ivo .
ANALYTICA CHIMICA ACTA, 2009, 648 (02) :167-173
[7]  
HITCHMAN ML, 1978, MEASUREMENT DISSOLVE, V49, P195
[8]   Sequential injection analysis for automation of the Winkler methodology, with real-time SIMPLEX optimization and shipboard application [J].
Horstkotte, Burkhard ;
Tovar-Sanchez, Antonio ;
Duarte, Carlos M. ;
Cerda, Victor .
ANALYTICA CHIMICA ACTA, 2010, 658 (02) :147-155
[9]  
LANGE R, 1969, CHEM OCEANOGRAPHY IN, P123
[10]   A simple fluorescent probe for the determination of dissolved oxygen based on the catalytic activation of oxygen by iron(II) chelates [J].
Luo, Wei ;
Abbas, M. E. ;
Zhu, Lihua ;
Zhou, Wenyi ;
Li, Kejing ;
Tang, Heqing ;
Liu, Shushen ;
Li, Weiying .
ANALYTICA CHIMICA ACTA, 2009, 640 (1-2) :63-67