BREATH-HOLD TARGET LOCALIZATION WITH SIMULTANEOUS KILOVOLTAGE/MEGAVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY AND FAST RECONSTRUCTION

被引:19
作者
Blessing, Manuel [1 ]
Stsepankou, Dzmitry [1 ]
Wertz, Hansjoerg [1 ]
Arns, Anna [1 ]
Lohr, Frank [1 ]
Hesser, Juergen [1 ]
Wenz, Frederik [1 ]
机构
[1] Heidelberg Univ, Med Fak Mannheim, Univ Med Ctr Mannheim, Dept Radiat Oncol, D-68167 Mannheim, Germany
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2010年 / 78卷 / 04期
关键词
Image guidance; treatment verification; cone-beam computed tomography; CBCT; kilovoltage/megavoltage imaging; fast tomographic reconstruction; CT; RADIOTHERAPY; KILOVOLTAGE; THERAPY;
D O I
10.1016/j.ijrobp.2010.01.030
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Hypofractionated high-dose radiotherapy for small lung tumors has typically been based on stereotaxy. Cone-beam computed tomography and breath-hold techniques have provided a noninvasive basis for precise cranial and extracranial patient positioning. The cone-beam computed tomography acquisition time of 60 s, however, is beyond the breath-hold capacity of patients, resulting in respiratory motion artifacts. By combining megavoltage (MV) and kilovoltage (kV) photon sources (mounted perpendicularly on the linear accelerator) and accelerating the gantry rotation to the allowed limit, the data acquisition time could be reduced to 15 s. Methods and Materials: An Elekta Synergy 6-MV linear accelerator, with iViewGT as the MV- and XVI as the kV-imaging device, was used with a Catphan phantom and an anthropomorphic thorax phantom. Both image sources performed continuous image acquisition, passing an angle interval of 90 within 15 s. For reconstruction, filtered back projection on a graphics processor unit was used. It reconstructed 100 projections acquired to a 512 x 512 x 512 volume within 6 s. Results: The resolution in the Catphan phantom (CTP528 high-resolution module) was 3 lines/cm. The spatial accuracy was within 2-3 mm. The diameters of different tumor shapes in the thorax phantom were determined within an accuracy of 1.6 mm. The signal-to-noise ratio was 68% less than that with a 180 degrees-kV scan. The dose generated to acquire the MV frames accumulated to 82.5 mGy, and the kV contribution was <6 mGy. Conclusion: The present results have shown that fast breath-hold, on-line volume imaging with a linear accelerator using simultaneous kV MV cone-beam computed tomography is promising and can potentially be used for image-guided radiotherapy for lung cancer patients in the near future. (C) 2010 Elsevier Inc.
引用
收藏
页码:1219 / 1226
页数:8
相关论文
共 50 条
  • [1] Breath-hold Target Localization with simultaneous Kilovoltage/Megavoltage Cone-Beam CT and fast Reconstruction
    Blessing, M.
    Stsepankou, D.
    Wertz, H.
    Arns, A.
    Lohr, F.
    Hesser, J.
    Wenz, F.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 201 - 204
  • [2] Essentials of Algebraic Reconstruction in Cone-Beam Computed Tomography
    Chernukha, A. E.
    Shestopalov, A. I.
    Adarova, A. I.
    Shershnev, R. V.
    Kizilova, Ya. V.
    Koryakin, S. N.
    Ivanov, S. A.
    Solovev, A. N.
    BULLETIN OF THE LEBEDEV PHYSICS INSTITUTE, 2023, 50 (10) : 438 - 444
  • [3] Kilovoltage cone-beam computed tomography imaging dose estimation and optimization: Need of daily cone-beam computed tomography
    Trivedi, Gaurav
    Dixit, Chandra K.
    Oinam, Arun S.
    Kapoor, Rakesh
    Bahl, Amit
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (03) : 470 - 474
  • [4] A study on image quality provided by a kilovoltage cone-beam computed tomography
    Garayoa, Julia
    Castro, Pablo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (01): : 239 - 257
  • [5] Methods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space
    Zhang, J.
    Bao, Z. R.
    Huang, X. T.
    Jiang, D. Z.
    Xie, C. H.
    Zhou, Y. F.
    Liu, H.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2019, 17 (02): : 189 - 202
  • [6] Essentials of Algebraic Reconstruction in Cone-Beam Computed Tomography
    A. E. Chernukha
    A. I. Shestopalov
    A. I. Adarova
    R. V. Shershnev
    Ya. V. Kizilova
    S. N. Koryakin
    S. A. Ivanov
    A. N. Solovev
    Bulletin of the Lebedev Physics Institute, 2023, 50 : 438 - 444
  • [7] Comparison of kilovoltage cone-beam computed tomography with megavoltage projection pairs for paraspinal radiosurgery patient alignment and position verification
    Kriminski, Sergey A.
    Lovelock, D. Michael
    Seshan, Venkatraman E.
    Ali, Imad
    Munro, Peter
    Amols, Howard I.
    Fuks, Zvi
    Bilsky, Mark
    Yamada, Yoshiya
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (05): : 1572 - 1580
  • [8] A Phantom Study on Target Localization Accuracy Using Cone-Beam Computed Tomography
    Yan, Hui
    Zhang, Liwei
    Yin, Fang-Fang
    CLINICAL MEDICINE INSIGHTS-ONCOLOGY, 2008, 2 : 501 - 510
  • [9] A moving blocker-based strategy for simultaneous megavoltage and kilovoltage scatter correction in cone-beam computed tomography image acquired during volumetric modulated arc therapy
    Ouyang, Luo
    Lee, Huichen Pam
    Wang, Jing
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 (03) : 425 - 430
  • [10] Evaluation of radiation dose to organs during kilovoltage cone-beam computed tomography using Monte Carlo simulation
    Son, Kihong
    Cho, Seungryong
    Kim, Jin Sung
    Han, Youngyih
    Ju, Sang Gyu
    Choi, Doo Ho
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 295 - 302