A magnetic resonance image-guided breast needle intervention robot system: overview and design considerations

被引:16
作者
Park, Samuel Byeongjun [1 ]
Kim, Jung-Gun [1 ]
Lim, Ki-Woong [1 ]
Yoon, Chae-Hyun [1 ]
Kim, Dong-Jun [1 ]
Kang, Han-Sung [2 ]
Jo, Yung-Ho [1 ]
机构
[1] Natl Canc Ctr, Dept Biomed Engn, 323 Ilsanro, Goyang Si 10408, Gyeonggi Do, South Korea
[2] Natl Canc Ctr, Ctr Breast Canc, 323 Ilsanro, Goyang Si 10408, Gyeonggi Do, South Korea
关键词
Breast cancer; Medical robotics; Magnetic resonance imaging; Needle intervention; Bendable needle; MRI; NEUROARM; BIOPSY;
D O I
10.1007/s11548-017-1528-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose We developed an image-guided intervention robot system that can be operated in a magnetic resonance (MR) imaging gantry. The system incorporates a bendable needle intervention robot for breast cancer patients that overcomes the space limitations of the MR gantry. Methods Most breast coil designs for breast MR imaging have side openings to allow manual localization. However, for many intervention procedures, the patient must be removed from the gantry. A robotic manipulation system with integrated image guidance softwarewas developed. Our robotic manipulator was designed to be slim, so as to fit between the patient's side and the MR gantry wall. Only non-magnetic materials were used, and an electromagnetic shieldwas employed for cables and circuits. The image guidance software was built using open source libraries. In situ feasibility tests were performed in a 3-T MR system. One target point in the breast phantom was chosen by the clinician for each experiment, and our robot moved the needle close to the target point. Results Without image-guided feedback control, the needle end could not hit the target point (distance = 5 mm) in the first experiment. Using our robotic system, the needle hits the target lesion of the breast phantom at a distance of 2.3mm from the same target point using image-guided feedback. The second experiment was performed using other target points, and the distance between the final needle end point and the target point was 0.8 mm. Conclusions We successfully developed an MR-guided needle intervention robot for breast cancer patients. Further research will allow the expansion of these interventions.
引用
收藏
页码:1319 / 1331
页数:13
相关论文
共 31 条
[1]  
[Anonymous], 2012, WHAT AR KEY STAT BRE
[2]  
Chinzei K, 2001, Med Sci Monit, V7, P153
[3]  
Cole G.A., 2014, EXPT ROBOTICS, P785, DOI DOI 10.1007/978-3-642-28572-1_
[4]  
Eliahou R, 2009, ISR MED ASSOC J, V11, P275
[5]   Cancer incidence and mortality worldwide: Sources, methods and major patterns in GLOBOCAN 2012 [J].
Ferlay, Jacques ;
Soerjomataram, Isabelle ;
Dikshit, Rajesh ;
Eser, Sultan ;
Mathers, Colin ;
Rebelo, Marise ;
Parkin, Donald Maxwell ;
Forman, David ;
Bray, Freddie .
INTERNATIONAL JOURNAL OF CANCER, 2015, 136 (05) :E359-E386
[6]  
Gonzalez R.C., 2007, Digital Image Processing, V3rd
[7]  
Kim J-G, 2014, 10 ANN AS C COMP AID, P70
[8]   MRI-compatible micromanipulator - Design and implementation and MRI-compatibility tests [J].
Koseki, Yoshihiko ;
Tanikawa, Tamio ;
Chinzei, Kiyoyuki .
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, :465-+
[9]   An MRI-Compatible Robotic System With Hybrid Tracking for MRI-Guided Prostate Intervention [J].
Krieger, Axel ;
Iordachita, Iulian I. ;
Guion, Peter ;
Singh, Anurag K. ;
Kaushal, Aradhana ;
Menard, Cynthia ;
Pinto, Peter A. ;
Camphausen, Kevin ;
Fichtinger, Gabor ;
Whitcomb, Louis L. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (11) :3049-3060
[10]   Robotic System for MRI-Guided Stereotactic Neurosurgery [J].
Li, Gang ;
Su, Hao ;
Cole, Gregory A. ;
Shang, Weijian ;
Harrington, Kevin ;
Camilo, Alex ;
Pilitsis, Julie G. ;
Fischer, Gregory S. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (04) :1077-1088